    Device Name :

    NEOMED ENTERAL ONLY EXTENSION SET

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NeoMed Enteral Only Extension Set is intended for use as an extension set for nasogastric/oralgastric or gastric tube enteral feeding tubes, incorporating safety connectors which help mitigate the risk of accidental misconnection with an I.V. system to the enteral system or the enteral system to an I.V. system.

    Device Description

    The NeoMed Enteral Only Extension Set is a sterile disposable for single patient use only device designed to help minimize the potential for inadvertent delivery of enteral feedings through the intravenous route. The device consists of flexible PVC tubing designed to connect existing feeding tubes (nasogastric, oralgastric, etc) to various delivery systems in including pumps and syringes. The set consists of tubing with an enteral connector (catheter tip) and an oral syringe connector not compatible with IV tubing or stopcocks. The NeoMed Enteral Only Extension Set provides an orange stripe for easy quick identification of enteral feeding lines as well as an "Enteral Only" tag and slide clamp to provide the additional safety assurance for connection errors.

    AI/ML Overview

    The provided text describes the NeoMed Enteral Only Extension Set, a Class II medical device. However, it does not contain the information needed to fill out a table of acceptance criteria, detailed device performance, or a comprehensive study description as requested.

    The document is a 510(k) summary, which focuses on establishing substantial equivalence to a predicate device rather than presenting detailed clinical study results or specific performance criteria with numerical targets.

    Here's a breakdown of what can be extracted and what is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryAcceptance Criteria (Target)Reported Device Performance (Result)
    Mechanical TestingNot specified quantitativelyPassed
    TensileNot specified quantitativelyPassed
    StrainNot specified quantitativelyPassed
    ElongationNot specified quantitativelyPassed
    Physical PropertiesNot specified quantitativelyPassed
    Dimensional VerificationNot specified quantitativelyPassed
    Flow RateNot specified quantitativelyPassed
    Hub CompatibilityNot specified quantitativelyPassed
    Biocompatibility TestingPassed per ISO 10993-1Passed
    CytotoxicityNot specified quantitativelyPassed
    SensitizationNot specified quantitativelyPassed
    IrritationNot specified quantitativelyPassed
    Packaging ChallengesNot specified quantitativelyPassed
    Package IntegrityNot specified quantitativelyPassed
    Product StabilityNot specified quantitativelyPassed

    Explanation of Table: The document states that "Design verification functional test results demonstrate that the NeoMed Enteral Only Extension Set performs its intended use." It then lists the types of tests conducted: "mechanical: tensile, strain, elongation and physical/properties testing comprised of dimensional verification, flow rate, and hub compatibility. Packaging challenges for package integrity and product stability were conducted with aging studies. All materials have been evaluated in accordance with ISO 10993-1: Biological Evaluation of Medical Devices – part 1: Evaluation and Testing and the three tests conducted - cytotoxicity, sensitization, and irritation pass."

    *Crucially, the specific quantitative acceptance criteria (e.g., "tensile strength > X N") and the numerical results (e.g., "tensile strength = Y N") for these tests are NOT provided in this document. The document only states that the tests were conducted and "pass" or "demonstrate that the device performs its intended use."


    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not specified. The document refers to "functional tests conducted" but does not give the number of units tested for each type of performance test.
    • Data Provenance: Not specified (e.g., country of origin of data). The testing appears to be internal validation by the manufacturer, NeoMed, Inc., or its contracted labs.
    • Retrospective or Prospective: N/A for this type of bench/functional testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This document describes bench and biocompatibility testing for a medical device (an extension set), not a diagnostic or prognostic device that requires expert-established ground truth from patient data. The "ground truth" for these tests would be the established engineering standards or regulatory requirements.

    4. Adjudication method for the test set

    • Not Applicable. As above, this is laboratory and bench testing, not a clinical study involving human readers or subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No such study was done. This device is an enteral extension set, a physical medical device, not an AI-powered diagnostic or decision support system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Engineering Standards and Regulatory Guidelines. For the functional and biocompatibility tests, the "ground truth" is adherence to established industry standards (e.g., ISO 10993-1 for biocompatibility) and the device's design specifications.

    8. The sample size for the training set

    • Not Applicable. There is no "training set" as this device does not involve machine learning or AI.

    9. How the ground truth for the training set was established

    • Not Applicable. As above, no training set.