Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K192021
    Manufacturer
    Date Cleared
    2019-08-22

    (24 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NCB Polyaxial Locking Plate System is indicated for temporary internal fixation of fractures and osteotomies of long bones.

    Device Description

    The NCB Polyaxial Locking Plate System is an extramedullary internal fixation plate system to be used for proximal humeral fractures. It is intended to be implanted either percutaneously or by a traditional open method. The Tivanium 3.5 mm locking screws are used to engage the proximal locking screw holes within the NCB Polyaxial Locking Plate System, Proximal Humeral Plates.

    AI/ML Overview

    I am sorry, but the provided text is a 510(k) summary for a medical device (NCB Polyaxial Locking Plate System). It describes the device, its intended use, comparison to predicate devices, and regulatory information.

    However, it does not contain the specific information required to answer your request about acceptance criteria for a study and its proving evidence.

    The "Performance Data (Nonclinical and/or Clinical)" section explicitly states: "Clinical data and conclusions were not needed for this device." This indicates that a clinical study with acceptance criteria for device performance, as you've requested, was not performed or at least not submitted as part of this 510(k). Instead, the performance data cited is: "Class II: Amendment of Design Controls with verification of mechanical integrity and resistance." This likely refers to non-clinical, benchtop testing to ensure the device meets engineering specifications, not a clinical study on patient outcomes or human reader performance.

    Therefore, I cannot provide:

    1. A table of acceptance criteria and reported device performance.
    2. Sample size for test set or data provenance.
    3. Number of experts or their qualifications for ground truth.
    4. Adjudication method for the test set.
    5. MRMC comparative effectiveness study results or effect size.
    6. Standalone algorithm performance.
    7. Type of ground truth used.
    8. Sample size for the training set.
    9. How ground truth for the training set was established.

    This 510(k) summary is for a bone fixation system, not an AI/software device that would typically involve the type of performance evaluation you are asking about (e.g., assessing accuracy, sensitivity, specificity, and human reader improvement with AI).

    Ask a Question

    Ask a specific question about this device

    K Number
    K081759
    Manufacturer
    Date Cleared
    2008-10-16

    (118 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NCB Polyaxial Locking Plate System is indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones.

    The Tivanium 3.5mm Locking Screws are intended to be used with the NCB Polyaxial Locking Plate System.

    Device Description

    The NCB Polyaxial Locking Plate System is an extramedullary internal fixation plate system to be used for proximal humeral fractures. It is intended to be implanted either percutaneously or by a traditional open method.

    The Tivanium 3.5mm locking screws are used to engage the proximal locking screw holes within the NCB Polyaxial Locking Plate System, Proximal Humeral Plates.

    AI/ML Overview

    This document describes the Zimmer NCB® Polyaxial Locking Plate System and Zimmer® Universal Locking System with 3.5mm Tivanium® Ti-6Al-4V Alloy Locking Screws. It's a 510(k) premarket notification for a Class II medical device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of the device (bone fixation plates and screws), the "acceptance criteria" are primarily related to safety, effectiveness, and substantial equivalence to predicate devices, rather than specific performance metrics like accuracy, sensitivity, or specificity commonly seen in diagnostic AI/software devices. The performance is assessed through non-clinical (laboratory/performance) testing.

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Intended UseSame as predicate deviceNCB Polyaxial Locking Plate System: Indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones. Tivanium 3.5mm Locking Screws: Intended to be used with the NCB Polyaxial Locking Plate System. (Matches predicate's intended use)
    Performance CharacteristicsSimilar to predicate deviceNon-clinical (Laboratory / performance) testing demonstrated that the device is safe and effective. (Implies performance is comparable to predicate, meeting safety and effectiveness standards)
    DesignSimilar to predicate deviceNCB plates are similar in design to the predicate device. Universal Locking System screws have similar design characteristics.
    MaterialsSimilar to predicate device (or justified difference)NCB plates are similar in materials to the predicate device. The proposed Universal Locking System screws are made from Tivanium (Ti-6Al-4V) alloy versus the Stainless Steel of the predicate device. (This difference was likely justified through non-clinical testing showing equivalence or superiority in relevant properties).
    Safety and EffectivenessDevice is safe and effectiveNon-clinical (Laboratory / performance) testing demonstrated that the device is safe and effective.

    2. Sample Size Used for the Test Set and Data Provenance

    This device is a hardware implant (plates and screws for bone fixation). The performance testing described is non-clinical (laboratory/performance) testing, not clinical trials involving human subjects or data sets in the way AI/software devices are typically evaluated.

    • Sample size for the test set: Not applicable in the context of clinical data. The document does not specify sample sizes for mechanical or laboratory tests, which would typically involve testing a certain number of devices or components under simulated conditions.
    • Data provenance: Not applicable. The "data" comes from laboratory and performance testing, not human-derived data like images or patient records.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. Ground truth for clinical data is not relevant here as the evaluation relies on non-clinical, laboratory performance testing of the hardware. The "ground truth" for such devices is established through engineering standards, biomechanical principles, and documented performance of predicate devices.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods are typically used for clinical data interpretation (e.g., in diagnostic studies) where human experts might disagree. This device was evaluated through non-clinical laboratory testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic or AI-assisted interpretation devices to assess the impact of AI on human reader performance. This document pertains to a bone fixation hardware device, for which MRMC studies are not applicable.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done. This device is a physical implant, not an algorithm or software.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's evaluation is based on engineering specifications, biomechanical testing standards, and performance characteristics established for bone fixation devices. It relies on laboratory test results demonstrating mechanical strength, fatigue resistance, biocompatibility (often implicitly covered by material commonality), and other physical properties under simulated conditions.

    8. The Sample Size for the Training Set

    Not applicable. This device is a hardware product, not a machine learning model. There is no "training set" in this context.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1