Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150234
    Date Cleared
    2015-03-31

    (57 days)

    Product Code
    Regulation Number
    892.5720
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Myriad Prostate Caddy¿ Immobilization Rectal Balloon Item #9901,3301

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Myriad Prostate Caddy™ Immobilization Rectal Balloon is a single-use disposable, inflatable, non-powered positioning device intended for use in the temporary positioning of the rectal wall and adjacent structure in the male human anatomies. The purpose of the device is to stabilize the prostate during Computed Tomography (CT) exam and Xray, when these imaging techniques are used for Radiation Therapy (RT) planning. The placement of the balloon requires a Physician or a Physician directed healthcare professional, and is performed as a separate procedure apart from the standard CT exam and RT treatment.

    Device Description

    The Myriad Prostate Caddy™ Immobilization Rectal Balloon is designed as an immobilizer to assist in positioning the prostate in a more predictable and reproducible location during Computed Tomography (CT) exam and X-ray, when these imaging techniques are used for Radiation Therapy (RT) planning. The Prostate Immobilization Rectal Balloon is inserted into the rectum and inflated prior to the start of a CT scan or RT therapy procedure. The device stabilizes the prostate once the device is inflated. The Prostate Immobilizer Rectal Balloon is deflated and removed after each individual scan or therapy procedure is complete, and a new balloon is used in the next therapy session. Myriad's device is designed for single use, is provided non-sterile to the end user, is not intended to be sterilized by the end user, and is packaged in a kit configuration. Each kit contains the following items: Rectal Balloon (PVC); Syringe (optional accessory to the device); Locking Stopper (accessory to the device) and the Instructions for Use. The balloon consists of a single lumen shaft with a blow molded balloon attached to the distal end of the shaft. The balloon is bonded to the shaft at the proximal end of the balloon. The shaft extends into the balloon but it is not attached at the end with a rounded atraumatic tip. Several eye holes are provided to inflate/deflation of the balloon. The main section of the device consists of a balloon and single lumen tubing which are made of PVC. The remainder of the device is made of other medical grade materials consisting of the following components shown below in Table 1. The device can be inflated with either air or water, and this process is carried out with the single use disposable syringe that is supplied (or offered as an option) with the device in the accessory kit. The locking stopper supplied with the Prostate Immobilizer Rectal Balloon controls the depth at which the balloon is inserted into the rectum. The device can therefore be locked into place once inserted into place.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Myriad Prostate Caddy™ Immobilization Rectal Balloon, which aims to demonstrate substantial equivalence to a predicate device, the RadiaDyne Prostate Immobilizer Rectal Balloon RB Classic Series (K132194). This document focuses on non-clinical/bench studies to assess the device's performance, rather than a clinical study involving human subjects or AI algorithms. As such, information typically found in studies for AI/CADe devices (like sample size for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth types) is not applicable or available in this document.

    However, the document does detail the acceptance criteria and performance of the device based on bench testing.

    1. A table of acceptance criteria and the reported device performance:

    Test Method (Acceptance Criteria)Reported Device Performance
    Biocompatibility:
    Cytotoxicity (non-cytotoxic)Not considered to be warranted as the safety of the device has been supported by the data presented. (Implies non-cytotoxic)
    Rectal Irritation (non-irritant, irritation index = 0)For the test samples, the irritation index was 0 or no reaction. The SC (Sodium chloride) and SO (Sesame oil) test extracts were considered nonirritants to the rectal tissue of the rabbit model.
    Sensitization (no evidence of dermal reaction, no delayed dermal contact sensitization)No observed evidence of dermal reaction with all sites scoring a 0 reaction. The test article showed no evidence of causing delayed dermal contact sensitization in the guinea pig. The test article (Myriad Prostate Balloon) was not considered a sensitizer in the guinea pig maximization test.
    Acute Systemic Toxicity (no mortality, normal animal appearance, weight gain, no systemic toxicity)For the test subject and the control, there was no mortality, the animals appeared normal and all animals gained weight. There was no mortality or evidence of systemic toxicity from the extracts injected into the mice. The results showed no difference between the test and control animals.
    Shelf Life/Microbial Contamination:
    Shelf-life: 2 years (device functions properly after 2 years storage)The sponsor has provided shelf-life testing data, which ensures that the subject device functions properly after 2 years of storage.
    Bioburden (non-sterile, acceptable limits per USP & USP ): Total Aerobic Count: ≤ 1000 CFU/g; Fungi Count: ≤ 100 CFU/gTotal Recoverable Aerobic Bioburden: average of 75 CFU (maximum 168 CFU). Total Recoverable Fungi Bioburden: average of 4 CFU (maximum 8 CFU). Bioburden results are well within acceptable limits.
    Performance Testing - Bench:
    Bioburden Testing (Maximum acceptable limit: 1000 CFU for aerobic, 100 CFU for fungi)Total Recoverable of Aerobic Bioburden was an average of 75 CFU with a maximum level of 168 CFU. The maximum acceptable limit is 1000 CFU. The Total Recoverable Fungi Bioburden was an average for 4 CFU with a maximum level of 8 CFU. The maximum acceptable limit is 100 CFU. The Bioburden results are well within acceptable limits.
    Leakage Testing (no leaks when inflated with 100cc water for 10 minutes)All units passed Leak Testing. All devices are subject to leak testing as the final quality control check (100% of Balloons inflated and held under pressure).
    Tensile-strength and Burst-strength (air): withstand 7.1 Lbs force at 100cc air inflationAll of the balloons passed the test with no signs of burst, leakage or joint failure. The test was repeated for the 2-Year Shelf Life Testing with no failures.
    Stopper Resistance Testing: withstand 3.57 Lbs tensile forceThe force to move the Stopper had an average value of 4.91 lbs. All of the units passed the test.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This document describes bench and biocompatibility testing, not a clinical study on a test set of patient data. Therefore, concepts like “test set,” “data provenance,” and “retrospective/prospective” are not applicable in this context. The samples used are devices themselves or animal models for biocompatibility.

    • Biocompatibility testing: Conducted at NAMSA, 6750 Wales Road, Northwood, Ohio 43619.
      • Cytotoxicity: Not specified individual sample count, but "data presented" supported safety.
      • Rectal Irritation: Rabbit model. Not specified sample count.
      • Sensitization: Guinea pig model. Not specified sample count.
      • Acute Systemic Toxicity: Mice. Not specified sample count.
    • Performance Testing - Bench:
      • Bioburden Testing: The testing process was repeated 3 times. The average of 75 CFU and maximum of 168 CFU suggests multiple device samples were used.
      • Leakage Testing: "A statistically appropriate number of devices" were tested, and "100 % of the Balloons are inflated" for final QC.
      • Tensile-strength and Burst-strength: Not specified individual sample count, but performed on "the Balloons."
      • Stopper Resistance Testing: Not specified individual sample count, but "All of the units passed the test."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. This device is a physical medical device, not an AI or imaging diagnostic tool. "Ground truth" in the sense of expert consensus on medical images is not relevant to its evaluation. Biocompatibility tests were conducted by a testing facility (NAMSA) according to established standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. There is no test set in the context of expert adjudication. Performance was evaluated based on physical measurements and biological responses in laboratory settings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. No MRMC study was performed as this is not an AI/CADe device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is established by:

    • Biocompatibility standards: International Organization for Standardization (ISO) 10993 series for biological evaluation of medical devices.
    • Microbial limits: USP and USP for Total Aerobic Count and Fungi.
    • Physical performance derived from engineering specifications and literature: For example, the 7.1 Lbs force for tensile/burst strength and 3.57 Lbs for stopper resistance are described as simulating worst-case usage per literature with a safety factor.

    8. The sample size for the training set

    Not applicable. This device does not use a training set for machine learning.

    9. How the ground truth for the training set was established

    Not applicable. This device does not use a training set for machine learning.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1