510(k) Data Aggregation

    K Number
    K181974
    Manufacturer
    Date Cleared
    2018-08-23

    (30 days)

    Product Code
    Regulation Number
    884.1690
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MyoSure XL Tissue Removal Device for Fluent in intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: Submucous myomas Endometrial Polyps Retained products of conception.

    Device Description

    The MyoSure XL Tissue Removal Device for Fluent is a sterile, single-use device that is intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: Submucous myomas, Endometrial Polyps, Retained products of conception. The MyoSure XL Tissue Removal Device for Fluent is designed to be used with a hysteroscope and to connect to the Fluent Fluid Management System. The MyoSure XL Tissue Removal Device for Fluent uses mechanical resection which allows the surgeon to have precise control over the locations and extent of tissue resected by drawing the targeted tissue into the cutting window under suction while the inner blade cuts the tissue.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "MyoSure XL Tissue Removal Device for Fluent." This device is a Class II medical device intended for hysteroscopic resection and removal of intrauterine tissue such as submucous myomas, endometrial polyps, and retained products of conception.

    The document focuses on demonstrating substantial equivalence to a predicate device (MyoSure Hysteroscopic Tissue Removal System and MyoSure Tissue Removal Devices, K172566). As such, the "acceptance criteria" discussed are primarily related to functional performance testing and biocompatibility to confirm that the modified device performs similarly to the predicate and meets established safety standards for medical devices.

    Here's an analysis of the provided information based on your questions:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in a quantitative, pass/fail manner with specific numerical thresholds for performance metrics. Instead, it describes types of functional tests conducted to demonstrate equivalence to the predicate device. The performance is reported as meeting design specifications and being substantially equivalent to the predicate.

    Acceptance Criterion (Type of Test)Reported Device Performance
    Biocompatibility
    Cytotoxicity (ISO 10993-5:2009)Complies with requirements
    Sensitization (ISO 10993-10:2010)Complies with requirements
    Irritation (ISO 10993-10:2010)Complies with requirements
    Acute Systemic Toxicity (ISO 10993-11:2006)Complies with requirements
    Functional Testing
    Connector CompatibilityPerformance demonstrates substantial equivalence to predicate device.
    Cutting PerformancePerformance demonstrates substantial equivalence to predicate device.
    Design verification (mechanical testing of handpiece)Complies with design specifications; Performance demonstrates substantial equivalence to predicate device.
    Reciprocation Rate TestPerformance demonstrates substantial equivalence to predicate device.
    Risk Management (ISO 14971:2007)Risks mitigated to an acceptable level.

    2. Sample size used for the test set and the data provenance:

    The document does not specify the sample sizes (e.g., number of devices) used for the biocompatibility or functional performance tests. It also does not explicitly state the provenance (country of origin, retrospective/prospective) of any data used for these tests. Given that this is a 510(k) for a hardware device with minor modifications, the tests are typically benchtop (in vitro) and simulated use, not clinical studies with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable and not provided in the document. For a 510(k) submission focused on minor device modifications and substantial equivalence through bench testing, ground truth is established by engineering specifications, validated test methods, and industry standards, rather than expert clinical interpretation of data.

    4. Adjudication method for the test set:

    This information is not applicable and not provided. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies involving human interpretation of medical images or disease diagnosis, which is not the nature of the testing described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable and not provided. The device is a surgical tissue removal device, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study and AI-related effect sizes are irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable and not provided. The device is a physical, mechanical tissue removal instrument, not a standalone algorithm.

    7. The type of ground truth used:

    For biocompatibility testing, the "ground truth" is adherence to international standards like ISO 10993. For functional testing, the "ground truth" is the device's design specifications and the demonstrated performance of the predicate device. This is based on engineering specifications and validated test methods, not clinical "expert consensus, pathology, or outcomes data" in the typical sense of a diagnostic or therapeutic clinical trial.

    8. The sample size for the training set:

    This information is not applicable and not provided. There is no AI component mentioned, so there is no "training set."

    9. How the ground truth for the training set was established:

    This information is not applicable and not provided, as there is no training set mentioned.