Search Results
Found 5 results
510(k) Data Aggregation
(108 days)
MONARCH Platform (MON-000008)
The MONARCHTM Bronchoscope and the MONARCHTM Platform and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
The MONARCH Platform is intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The MONARCH Platform enables robotic electro-mechanical articulation and precise control of a flexible bronchoscope under continuous and direct control by a physician operator.
The MONARCH Platform allows for precise access of the lung anatomy and continuous visualization using the bronchoscope distal tip camera. The MONARCH Platform consists of four major components, (1) MONARCHTM Cart, (2) MONARCHTM Tower, (3) MONARCHTM Bronchoscope, (4) MONARCHTM Controller and working channel instruments and accessories. The MONARCH Cart provides support for the effector arms. It includes up to three robotic arms and the electronic systems required to power and operate the robotic system. The robotic arms possess multiple degrees of freedom. The MONARCH Tower is the primary user (i.e. physician) procedural display interface. It contains a monitor for user viewing and computers running the system software. The tower provides connectivity for the bronchoscope camera and lighting, as well as the fluidics system. The user controls the system with an endoscopic controller which transmits user inputs through the electromechanical system to the bronchoscope. The flexible MONARCH Bronchoscope has a working channel and a camera at the distal end. The working channel of the Bronchoscope is used for irrigation, aspiration and to deliver the working channel instruments.
Additionally, the Monarch Platform includes electromagnetic (EM) navigation and integrates a pre-operative computed tomography (CT) scan into an intra-operative interface, displaying the modeled bronchoscope tip location relative to the pre-operative scan anatomy.
The MONARCH Platform is being updated to receive 3D imaging data via an ethernet connection directly from a Cone Beam CT (CBCT) System with 3D imaging technology. This will enable real time navigation updates during a procedure with the device. This integration feature is optional for the user, but if selected, use of the MONARCHTM Window Field Generator is required.
The modified device includes software updates for navigation and presentation of the anatomy for display and planning purposes, firmware updates for a revised controller, and hardware and software revisions for the purposes of cybersecurity.
The document provided does not contain a table of acceptance criteria and reported device performance. It outlines various non-clinical performance tests conducted to verify the modified MONARCH™ Platform's safety and effectiveness.
Here's the information that could be extracted from the document:
1. A table of acceptance criteria and the reported device performance
The document does not provide a specific table with acceptance criteria and corresponding reported device performance values. However, it lists several non-clinical tests performed and states that the device "passed all tests in accordance with appropriate test criteria and standards."
Non-Clinical Performance Tests Performed:
- Proof of Design electrical tests: Verified all hardware modules perform within specifications.
- Location accuracy tests: For navigation purposes.
- Software functional tests: Covered complete system functionality, including error handling, usability, time, and ability to receive, process, and display imaging data from cone beam CT Systems.
- Safety, EMC, and mechanical tests: Performed by an independent nationally recognized testing laboratory to verify compliance with safety and EMC standards for medical devices to address hardware changes.
- Simulated use accuracy test: A complete CT image registration and instrument navigation workflow was performed to verify the overall accuracy of the system.
- Usability (Human Factors) Pre-clinical (cadaver) tests: Designed to mimic surgical procedures using the MONARCH™ Platform in a simulated clinical environment to assess the execution of a complete robotically assisted bronchoscopy procedure workflow and to qualitatively estimate the system clinical accuracy.
Reported Performance (General Statement):
"The proposed MONARCH Platform passed all tests in accordance with appropriate test criteria and standards, and the modified device did not raise new questions of safety or effectiveness."
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document mentions "Simulated use accuracy test" and "Usability (Human Factors) Pre-clinical (cadaver) tests."
- For the simulated use accuracy test, no specific sample size (number of cases or measurements) is provided.
- For the usability (human factors) pre-clinical tests, it states "cadaver tests," indicating ex-vivo human anatomical models were used.
- The document does not specify the country of origin or whether the data was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
The document does not provide information on the number or qualifications of experts used to establish ground truth for any of the tests. For the human factors/cadaver testing, it states the tests were "designed to mimic surgical procedures...to qualitatively estimate the system clinical accuracy," implying expert involvement, but no details are given.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
The document does not mention any adjudication method used for either the test set (if referring to the simulated or cadaver testing) or for establishing ground truth.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document does not describe an MRMC comparative effectiveness study or any study evaluating the improvement of human readers with vs. without AI assistance. The focus of this submission is on updates to the MONARCH™ Platform for direct Cone Beam CT integration and cybersecurity, rather than a new AI-driven diagnostic or assistive feature assessed with human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
The document specifically describes the MONARCH™ Platform as enabling "robotic electro-mechanical articulation and precise control of a flexible bronchoscope under continuous and direct control by a physician operator." The tests performed are in the context of this human-in-the-loop operation, for example, "Simulated use accuracy test, in which a complete CT image registration and instrument navigation workflow was performed." Therefore, a standalone algorithm-only performance study without human-in-the-loop was not performed or described as relevant to this device update. The closest to an 'algorithm only' test would be software functional tests and location accuracy tests, but these are components of the overall human-controlled system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The document refers to "location accuracy tests," and "simulated use accuracy test" which implies a quantitative ground truth for positional accuracy. For the "Usability (Human Factors) Pre-clinical (cadaver) tests," the goal was to "qualitatively estimate the system clinical accuracy," suggesting subjective expert assessment in a simulated environment, without specifying a formal ground truth like pathology or outcomes data.
8. The sample size for the training set
The document does not mention a training set or its sample size. This submission focuses on modifications to an existing device (software updates for navigation and display, firmware for a revised controller, and cybersecurity revisions) and integrates new input from CBCT, rather than describing a new AI model that would typically require a training set.
9. How the ground truth for the training set was established
As no training set is mentioned, information on how its ground truth was established is not provided.
Ask a specific question about this device
(205 days)
Monarch Platform, Urology
The Monarch® Platform, Urology and its accessories are intended to access and visualize anatomical locations within the urinary tract and interior of adolescents and adults, aged 12 and up, for diagnostic and therapeutic procedures with transurethral access or transurethral access in conjunction with percutaneous access.
The Monarch® Platform, Urology, the Ureteroscope, and endourology accessories are indicated to provide endoscopic visualization and access of organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) with transurethral access or transurethral access in conjunction with percutaneous access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.
The Monarch® Platform, Urology ("Proposed Device") is a capital equipment platform that enables electro-mechanical articulation and precise control of a flexible ureteroscope and/or a flexible suction catheter for visualization and access to the urinary tract for diagnostic and therapeutic procedures. The ureteroscope and suction catheter move only under continuous and direct physician control, via the Monarch Controller.
The proposed device consists of Monarch® Tower, Monarch® Cart, Monarch® Fluidics Pump, Monarch® Software, Monarch® Controller, Monarch® Ureteroscope and endourology instruments and accessories needed to perform Monarch Urology diagnostic and therapeutic procedures. The device as well as its instruments and accessories are used together during urological diaenostic and therapeutic procedures and are bundled in this 510(k) submission so they can be addressed during one review. The summary descriptions of each component are outlined below.
Monarch® Tower: The Monarch® Tower (Tower) contains the primary procedural display interface for the physician that is provided by a touchscreen monitor for physician viewing and computers running the Monarch software. The monitor has a touchscreen and allows for user input during setup and intra-operative use. The tower has integrated electromagnetic (EM) tracking and video signal processing components. In addition, the tower provides connectivity for the endoscope camera as well as the fluidics pump.
Monarch® Cart: The Monarch® Cart (Cart) contains a touchscreen, three robotic arms and electronic systems required to power and operate the robotic arms. The Cart touchscreens accepts user input and shows instructions for arm deployment and cart functions. The robotic arms possess multiple degrees of freedom. The cart transmits the physician directed movement to the Ureteroscope and other instruments during a procedure.
Monarch® Fluidics Pump: The Monarch® Fluidics Pump (Fluidics Pump) controls the irrigation and suction during a procedure. The Fluidics Pump is connected to the Monarch Tower via an umbilical cable allowing electrical and data communication. The Fluidics Pump provides the physician with the ability to control irrigation and suction levels during a procedure. The Urology Irrigation Cartridge and Suction Set are irrigation and suction tubing provided as sterile accessories to connect the Fluidics Pump to the Ureteroscope and Instruments during ureteroscopy or PCNL procedures.
Monarch® Software: The Monarch® Software provides the user with the ability to safely drive the Monarch® Ureteroscope and Monarch® Mini-PCNL suction catheter around the kidney and within the calyces of interest as well as provides optional navigation guidance to the user to insert the needle to the target during Monarch® PCNL procedure. It receives user input from the Monarch® Controller, computes the appropriate robotic motion to coordinate the movement of the robotic arms and endoscopic devices loaded onto the Cart. It provides a graphical user interface where the endoscopic camera view is shown in real time and displays important system status information.
Monarch® Controller: The Monarch® Controller) is a handheld device that serves as the user interface that allows the physician to control the system during a procedure. An Umbilical cord connects the Controller to the Tower.
Monarch® Ureteroscope: The Monarch® Ureteroscope (Ureteroscope) is comprised of an endoscope that provides vision, illumination, and a working channel to the distal tip of the device. The Ureteroscope can be navigated by the user within the bladder, urinary tract, and kidney. It contains a working channel to accommodate compatible commercially available working channel instruments. The Ureteroscope can be articulated 2-directions (along a single plane) when manually driven or can be articulated 4-directions (along two planes) when mounted to the electromechanical arms under command by the physician using the endoscopic controller. The shaft of the ureteroscope can also be rolled 155° in either direction to reorient the camera and working channel. The Ureteroscope is packaged within the Monarch® Ureteroscopy Kit. The Monarch® Ureteroscopy Kit also includes Ureteroscope Valve, Ureteroscope Driver and Laser Driver (accessories for Ureteroscopy procedure).
Procedural specific instruments and accessories for the proposed device are outlined below, these devices are single-use devices and sterilized via EO.
- Monarch® Mini-PCNL Suction Catheter: The Monarch® Mini-PCNL Suction Catheter ● (Suction Catheter) is part of the Monarch® Mini-PCNL Kit. The Suction Catheter is a flexible device that is inserted into the kidney via antegrade tract under direct visualization from the ureteroscope. It has a center lumen which allows for the suction of stone debris and can articulate in 4 directions. The suction catheter is manipulated by one robotic arm under the control of a physician who is concurrently controlling the Ureteroscope.
- Dilation Set: The Dilation Set is part of the Monarch® Mini-PCNL Kit. The Dilation Set ● includes:
- 8Fr dilator
- 10Fr catheter
- Percutaneous Sheath Dilator ●
- Percutaneous Sheath: The Percutaneous Sheath is part of the Monarch® Mini-PCNL Kit. The Percutaneous Sheath consists of an inner metal sheath with a threaded connection and a metal outer sheath. It is designed to establish percutaneous access and designed to allow simultaneous irrigation and suction of fluid to support removal of kidney stones. The outer sheath has a molded hub with an angled luer connection for irrigation and an insertion depth marking on the distal end.
The provided text describes performance data but does not specify quantifiable acceptance criteria for the Monarch® Platform, Urology device. Instead, it states that the device was tested against internal design specifications and applicable performance standards to demonstrate safety and effectiveness, and that "all intended user needs evaluated met their acceptance criteria." However, what those specific quantitative acceptance criteria were for different aspects (e.g., accuracy, reliability, image quality) and the reported performance against them are not detailed in the provided document.
Therefore, the following information is based on what can be extracted or inferred from the provided sections, recognizing the absence of explicit quantitative acceptance criteria and specific performance metrics for the device itself.
1. Table of Acceptance Criteria and Reported Device Performance
As mentioned above, the document does not provide a table of explicit acceptance criteria with specific quantitative targets and corresponding reported device performance values. The general statements are that all tests "Passed" or "met acceptance criteria."
Therefore, I cannot construct the requested table with specific numerical values for acceptance criteria and reported device performance.
2. Sample Size Used for the Test Set and Data Provenance
The document mentions "live porcine model" for Design Validation and "Acute Animal Safety" studies.
- Sample Size for Test Set: Not explicitly stated as a numerical count for either the Design Validation or Acute Animal Safety studies. The phrasing "live porcine model" and "live animal model" are general.
- Data Provenance: The studies were conducted using "live porcine model" and "live animal model," indicating animal testing (pre-clinical). The location or country of origin is not specified, but it suggests a controlled laboratory setting. The studies are prospective in nature, as they involve active testing and data collection.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document mentions "intended users" and "representative users" in the context of Human Factors and Usability evaluation and Design Validation. However, it does not specify the number of experts used to establish ground truth for any test set or their specific qualifications (e.g., "radiologist with 10 years of experience").
4. Adjudication Method for the Test Set
The document does not describe any adjudication method (e.g., 2+1, 3+1) for establishing ground truth for any test set. The term "ground truth" itself is not explicitly used in the context of expert review in the provided text.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
The document does not mention or describe a Multi Reader Multi Case (MRMC) comparative effectiveness study. The Monarch® Platform is described as a capital equipment platform with electromechanical articulation and control, and software for navigation, but there is no indication of an AI component designed to assist human readers in interpretation or diagnosis, nor any study comparing human performance with and without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
The document describes the Monarch® Platform as a system that requires "continuous and direct physician control" and where "Device movement is only at the command of the physician." The software provides "optional navigation guidance." This indicates that the device is designed for human-in-the-loop operation, and the description does not support the idea of a standalone (algorithm-only) performance study.
7. The Type of Ground Truth Used
Based on the descriptions:
- Design Validation and Acute Animal Safety: The "ground truth" for these studies would implicitly be established through direct observation, physiological measurements, and potentially post-mortem examination in the animal models, assessing safety and functionality in a simulated use environment. This is a form of direct-measurement or observation-based ground truth specific to animal model performance.
- Other tests (Reprocessing, Sterility, Biocompatibility, Electrical Safety, Software, Verification Testing, Human Factors): These rely on adherence to established standards and internal specifications, where "ground truth" is defined by the objective pass/fail criteria of technical tests rather than expert consensus on diagnostic images or pathology.
8. The Sample Size for the Training Set
The document describes pre-market testing and validation for a medical device (Monarch® Platform, Urology). It does not mention a "training set" in the context of machine learning or AI, as the primary focus seems to be on the electromechanical system and software components rather than a learning algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
As no "training set" for an AI/ML algorithm is mentioned, this question is not applicable based on the provided text.
Ask a specific question about this device
(57 days)
Monarch Platform
The Monarch Platform and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
The Monarch Platform is intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The Monarch Platform enables robotic electro-mechanical articulation and precise control of a flexible bronchoscope under continuous and direct control by a physician operator. The Monarch Platform allows for precise access of the lung anatomy and continuous visualization using the bronchoscope distal tip camera.
The Monarch Platform consists of three major components, (1) Monarch Cart, (2) Monarch Tower, and (3) Monarch Bronchoscope, and working channel instruments and accessories. The Monarch Cart provides support for the robotic arms. It includes up to three robotic arms and the electronic systems required to power and operate the robotic system. The flexible Monarch Bronchoscope has a working channel and a camera at the tip. The working channel of the Bronchoscope is used for irrigation, aspiration and to deliver the working channel instruments.
Additionally, the Monarch Platform includes electromagnetic (EM) navigation that integrates a pre-operative computed tomography (CT) scan into an intra-operative interface. displaying the modeled bronchoscope tip location relative to the preoperative scan anatomy. Two options of field generators are available to enable electromagnetic navigation. The current Monarch Navigation Field Generator and a new Window Field Generator (WFG). The new WFG is being added to increase compatibility of the Monarch system with advanced imaging systems (Cone Beam and Mobile CT).
Here's an analysis of the provided text regarding the Monarch Platform's acceptance criteria and study information:
Device Acceptance Criteria and Performance Study Analysis
The document describes the Monarch Platform, a robotic bronchoscopy system. The premarket notification (K211493) focuses on modifications to an existing cleared device (K193534), specifically the introduction of a "Window Field Generator (WFG)" as an optional accessory and minor software changes. Therefore, the performance data presented is primarily to demonstrate that these changes do not raise new questions of safety or effectiveness compared to the predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not provide a formal table of quantitative acceptance criteria with specific thresholds (e.g., accuracy > 95%). Instead, it generally states that the device was tested to ensure it performs "as intended" and that changes "do not raise different questions of safety or effectiveness."
Here's a summary of the reported device performance based on the general statements:
Aspect Tested | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Overall System Performance | Performs as intended, no new safety/effectiveness questions raised by modifications. | "Overall, device performance testing showed that the differences in technological characteristics do not raise different questions of safety or effectiveness." |
Electrical Safety | Meets relevant electrical safety standards. | "Verification testing included electrical safety..." |
Electromagnetic Compatibility (EMC) | Meets relevant EMC standards. | "...and electromagnetic compatibility..." |
Software Functionality | Software changes operate correctly and do not introduce new risks. | "...software, and performance specifications verification testing." (Specifically mentions "minor software changes" including arm kinematics, smoother virtual views, new 'Segmentation Failure' notification trigger, and reduced red saturation in live endoscopy view.) |
Window Field Generator (WFG) Functionality | Performs the same function as the existing Monarch Field Generator, with increased mechanical compatibility. | "This new WFG has the same core technological characteristics as the current Monarch Field Generator but is secured under the patient bed to optimize mechanical compatibility. Both EM field generators perform the same function for the Monarch Platform." |
Clinical Input Requirements (CIR) | Meets defined CIRs for its intended use. | "confirmatory validation testing was performed to ensure the device meets its Clinical Input Requirements (CIR) for its intended use." |
Impact on Workflow | Modifications do not negatively impact the pertinent parts of the workflow. | "Validation Testing included evaluations of the pertinent parts of workflow impacted by this change with accredited bronchoscopists." |
2. Sample Size and Data Provenance
The document does not specify the sample size used for the test set. It mentions "evaluations of the pertinent parts of workflow impacted by this change with accredited bronchoscopists," but provides no numbers for patients, cases, or procedures.
The data provenance is not explicitly stated as retrospective or prospective, nor is the country of origin. Given the context of a 510(k) submission for a medical device cleared in the U.S., it's highly likely the data would come from U.S. clinical settings, but this is an inference.
3. Number and Qualifications of Experts
The document states that "accredited bronchoscopists" were involved in the "Validation Testing including evaluations of the pertinent parts of workflow." It does not specify the number of experts or their years of experience/specific qualifications beyond being "accredited bronchoscopists."
4. Adjudication Method
The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) for the test set or for establishing ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No mention is made of a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance. The study described focuses on verifying the safety and effectiveness of device modifications rather than assessing the AI's impact on human reader performance.
6. Standalone Performance (Algorithm Only)
The document does not explicitly state that a standalone (algorithm only without human-in-the-loop performance) study was done for the navigation updates. The "navigation was updated for smoother virtual views when transitioning between predicted airways" and "An additional condition for which a "Segmentation Failure" notification would be triggered" suggest algorithm internal testing, but no formal standalone performance study results are provided in the summary. The Monarch Platform is an integrated robotic system where the navigation is an aid to the physician, not a standalone diagnostic tool.
7. Type of Ground Truth Used
The type of ground truth for the device's performance, particularly for the navigation and segmentation aspects, is not explicitly defined. However, given the context of bronchoscopic visualization and access, the "ground truth" for the device's ability to maintain its intended function would likely involve:
- Imaging data: Pre-operative CT scans are used for EM navigation, and live endoscopy views provide real-time visualization.
- Expert assessment: "Accredited bronchoscopists" evaluated workflow, implying their professional judgment against clinical standards would serve as a form of ground truth for usability and functional performance.
- Engineering specifications: Performance verification against design specifications would constitute a ground truth for many technical aspects.
Pathology or extensive outcomes data are not mentioned as being used for the performance testing described.
8. Sample Size for the Training Set
The document does not provide any information regarding the sample size for a training set. This is because the submission is for modifications to an already cleared device, and the focus is on verification and validation of the changes, not a de novo AI algorithm development where training data would typically be detailed.
9. How Ground Truth for the Training Set Was Established
Since no training set information is provided, there is no information on how its ground truth would have been established.
Ask a specific question about this device
(117 days)
Monarch Platform
The Monarch Platform and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
The Monarch Platform is intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The Monarch Platform enables robotic electro-mechanical articulation and precise control of a flexible bronchoscope under continuous and direct control by a physician operator. The Monarch Platform allows for precise access of the lung anatomy and continuous visualization using the bronchoscope distal tip camera.
The Monarch Platform consists of three major components, (1) Monarch Cart, (2) Monarch Tower, and (3) Monarch Bronchoscope, and working channel instruments and accessories. The Monarch Cart provides support for the effector arms. It includes three robotic arms and the electronic systems required to power and operate the robotic system. The robotic arms possess multiple degrees of freedom. The Monarch Tower is the primary user (i.e. physician) procedural display interface. It contains a monitor for user viewing and computers running the system software. The tower provides connectivity for the bronchoscope camera and lighting, as well as the fluidics system. The user controls the system with an endoscopic controller which transmits user inputs through the electromechanical system to the bronchoscope. The flexible Monarch Bronchoscope has a working channel and a camera at the tip. The working channel of the Bronchoscope is used for irrigation, aspiration and to deliver the working channel instruments. The singleuse manually controlled Auris instruments compatible with the Monarch Platform include the Aspirating Biopsy Needle, Biopsy Forceps, and Cytology Brush.
Additionally, the Monarch Platform includes electromagnetic (EM) navigation that integrates a pre-operative computed tomography (CT) scan into an intra-operative interface, displaying the modeled bronchoscope tip location relative to the pre-operative scan anatomy.
This document is a 510(k) summary for the Auris Health, Inc. Monarch Platform, a bronchoscope. It describes the device, its intended use, and a comparison to a predicate device, focusing on functional aspects and basic performance testing rather than specific AI/ML performance metrics typically found in diagnostic device submissions.
Based on the provided text, there is no mention of acceptance criteria or a study that evaluates the device's performance in terms of AI/ML or diagnostic accuracy, which would typically involve metrics like sensitivity, specificity, or accuracy compared to a ground truth. The performance data section focuses on general device functionality, safety, and usability.
Therefore, many of the requested elements for AI/ML performance studies cannot be extracted from this document, as it describes a robotic-assisted bronchoscope rather than an AI-powered diagnostic tool.
However, I can extract information related to the general performance testing and ground truth establishment for the device's functionality:
1. A table of acceptance criteria and the reported device performance
The document does not provide a formal table of acceptance criteria with specific quantitative performance metrics typically associated with AI/ML diagnostic devices (e.g., sensitivity, specificity, AUC). Instead, it describes general verification and validation testing outcomes.
Acceptance Criteria (Inferred from testing descriptions) | Reported Device Performance |
---|---|
Biocompatibility requirements per ISO 10993-1 | Met |
EMC compliance and electrical safety per AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 60601-2-18 | Fully evaluated and tested, found compliant |
Overall functionality (pre-operative planning, workflow, latency, functional safety, irrigation/aspiration, general functionality) | Conforms to design specifications, meets intended user needs, satisfied all sub-system specifications, exhibits electrical, mechanical, and functional integrity |
User needs and safety/effectiveness under simulated use clinical conditions | Design meets intended user requirements and facilitates safe and effective use |
Safe and effective use by representative users during a simulated bronchoscopy procedure after training | Assessed as safe and effective |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: Not explicitly stated for specific quantitative metrics or AI model evaluation. The document mentions "animal and cadaver testing" and "simulated use testing."
- Data Provenance: Not specified. Animal and cadaver testing
and simulated use testing are mentioned, implying experimental data rather than patient data from a specific country. This is prospective testing in a controlled environment.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable to this type of device submission's performance data. The device's "performance" here refers to its mechanical and functional integrity and usability, not diagnostic accuracy requiring expert ground truth in the traditional sense. The "user needs" and "safe and effective use" would implicitly involve clinical experts as testers/evaluators.
4. Adjudication method for the test set
Not applicable. The performance testing described is mechanical, electrical, and usability verification, not a clinical trial requiring adjudication of diagnostic outcomes.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a robotic-assisted bronchoscope, not an AI diagnostic tool. No AI component is described as assisting human readers in interpreting medical images or data. The "AI" in this context refers to robotic control and electromagnetic navigation, not diagnostic decision support.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a human-controlled robotic system. Its function inherently involves a human operator ("physician operator").
7. The type of ground truth used
The "ground truth" for the verification and validation tests are:
- Biocompatibility: Established by adherence to ISO 10993-1 standards and the results of specific biological evaluations.
- Electrical Safety and EMC: Established by compliance with international standards (AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-18).
- System Level Tests: Established by meeting predefined "design input requirements" and "sub-system specifications."
- Validation Testing (Animal/Cadaver): Established by the device meeting "intended user requirements" and facilitating "safe and effective use" in simulated clinical scenarios. This is essentially functional ground truth based on simulated procedural outcomes.
- Human Factors and Usability Testing: Established by the device being assessed for "safety and effective use" by representative users in simulated procedures, likely against predefined usability objectives.
8. The sample size for the training set
Not applicable. This device is described as a robotic platform for physical procedures, not an AI/ML model that requires a training set of data.
9. How the ground truth for the training set was established
Not applicable. As no AI/ML training set is described.
Ask a specific question about this device
(101 days)
Monarch Endoscopy Platform (Monarch Platform)
The Monarch Endoscopy Platform (Monarch Platform) and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
The Monarch™ Endoscopy Platform (Monarch Platform) is intended to be used by qualified physicians to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The Monarch Platform consists of three major components, (1) Monarch™ Endoscopy Cart, (2) Monarch™ Endoscopy Tower, and (3) Monarch™ Bronchoscope System, and working channel instruments.
The Monarch Cart provides support for the effector arms. It includes two robotic arms and the electronic systems required to power and operate the robotic system.
The Monarch Tower is the primary user (i.e. physician) procedural display interface. It contains a monitor for user viewing and computers running the system software. The tower provides connectivity for the bronchoscope camera and lighting, as well as the fluidics system. In addition, the tower includes an endoscopic controller that allows the user to control the system during a procedure.
The user controls the system with an endoscopic controller which transmits user inputs through the electromechanical system to the end effectors. The Monarch Bronchoscope System comprising of a bronchoscope and sheath is attached at the end effector of a robotic arm with multiple degrees of freedom. The flexible bronchoscope has a working channel and a camera at the tip. The bronchoscope has an articulated tip that can bend in all directions. The working channel of the bronchoscope is used for irrigation, aspiration and to deliver the working channel instruments.
The single-use manually controlled working channel instruments compatible with the Monarch Platform include the Aspirating Biopsy Needle, Biopsy Forceps, and Cytology Brush. All three have the same technological characteristics as the respective predicates and reference devices, that is, they are intended to operate in the same manner for collection of tissue or cells in the patient's lungs through a flexible endoscope or other working channel.
The Aspirating Biopsy Needle is comprised of a polymeric shaft with a needle joined to the distal end. A Luer fitting at the proximal end of the shaft allows the user to provide suction to the needle via the supplied syringe. A handle mechanism interfaces with the shaft to provide extension and retraction of the needle, and an outer jacket provides protection when the needle is not in use.
The Cytology Brush is comprised of an outer sheath and an inner catheter assembly. The inner catheter assembly consists of a thumb ring at the proximal end and a shaft to connect to the distal end which terminates in a brush. When the catheter is inserted into a working channel, the distal brush would be in a retracted position inside the outer sheath. When the catheter is in position, the brush can be extended into the tissue to obtain samples by advancing the proximal thumb ring.
The Biopsy Forceps instrument is comprised of a coil pipe with a mechanical jaw assembly joined to the distal end. The jaws are opened by sliding the outer handle component distally relative to the inner, longer handle component. The jaws are closed by squeezing the two components together.
The Monarch Endoscopy Platform (Monarch Platform) and its accessories were evaluated for safety and effectiveness through various performance tests, including biocompatibility, electrical safety, electromagnetic compatibility, software verification and validation, specific component performance testing, a navigational accuracy comparative study, animal and cadaver testing, and usability/human factors testing.
Here's a breakdown of the acceptance criteria and the studies performed, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present acceptance criteria in a table format with specific quantitative targets for the Monarch Endoscopy Platform as a whole, or for each individual performance test. Instead, it states that the device "successfully tested for function, performance, and safety per FDA-recognized standards" and "complies with" or "conforms to" specific standards and design specifications. For the Navigational Accuracy Comparative Study, it mentions "substantially equivalent segmental, sub-segmental accuracy and on target accuracy, and thus equivalent Navigation Yield" compared to the predicate.
Therefore, a direct table of "acceptance criteria" and "reported performance" like for a diagnostic accuracy study is not directly extractable for every aspect. However, we can synthesize the information for the Navigational Accuracy Comparative Study:
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Navigational Yield (Accuracy) substantially equivalent to predicate device (superDimension™ System) | Substantially equivalent segmental, sub-segmental accuracy and on-target accuracy to the superDimension™ System, leading to equivalent Navigational Yield. |
Distance to Target accuracy substantially equivalent to predicate device (superDimension™ System) | Substantially equivalent to the superDimension™ System. |
For all other performance testing categories (Biocompatibility, Electrical Safety, EMC, Software V&V, specific component testing, animal/cadaver, usability/human factors), the acceptance criterion was "compliance with recognized standards" and "meeting intended user requirements/design specifications," and the reported performance was that the device met these.
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a distinct "test set" and corresponding sample sizes for most of the performance evaluations in terms of human subjects or distinct data points in a clinical context. The studies involved:
- Navigational Accuracy Comparative Study: The document doesn't explicitly state the sample size (number of cases or measurements) used in this comparative study.
- Data Provenance: Not specified (e.g., country of origin, retrospective/prospective).
- Animal and Cadaver Testing: Performed to evaluate the Monarch Platform under simulated use conditions.
- Sample Size: Not specified (number of animals/cadavers).
- Data Provenance: Prospective, simulated use.
- Usability/Human Factors Testing:
- Sample Size: Not specified (number of representative users).
- Data Provenance: Prospective, simulated use.
- Biocompatibility, Electrical Safety, EMC, Software V&V, and individual component performance testing: These are typically bench tests or engineering evaluations, not usually associated with a "test set" sample size in the clinical data sense.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This information is not provided in the document for any of the studies described. The studies appear to be primarily engineering, non-clinical (animal/cadaver), and simulated use evaluations rather than studies requiring expert adjudication of ground truth for diagnostic accuracy in patient data.
4. Adjudication Method for the Test Set
Since the document does not describe studies involving human medical experts establishing ground truth for a diagnostic test set, there is no mention of an adjudication method (e.g., 2+1, 3+1).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned in the provided text. The "Navigational Accuracy Comparative Study" compares the Monarch Platform's navigation metrics to the superDimension™ System, but it's not described as an MRMC study and therefore no effect size of human readers improving with AI vs. without AI assistance is provided. The Monarch Platform is described as a robotic endoscopy system that assists physicians, but the focus of the performance data section is on the device's inherent capabilities and comparison to existing technologies, not on augmented human performance in an MRMC setting.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
The performance testing section primarily describes standalone performance aspects of the Monarch Platform and its accessories (e.g., biocompatibility, electrical safety, software V&V, mechanical performance of instruments, navigational accuracy of the system). The navigational accuracy study comparing Monarch and superDimension™ evaluates the system's navigation performance itself. The animal, cadaver, and usability studies involve human users but evaluate the system's safety and effectiveness with human interaction, not a standalone algorithm's diagnostic performance in the absence of human input.
Therefore, while the individual component and software testing represent "standalone" evaluations of those specific parts, a standalone diagnostic algorithm performance study (without human-in-the-loop) for a clinical outcome is not explicitly detailed as described for typical AI/ML devices. The Monarch Platform is a surgical robotic system with navigation features, so its "standalone" performance relates to its ability to perform its function (e.g., navigate, manipulate instruments) rather than to interpret images or diagnose conditions like a standalone AI diagnostic software.
7. The Type of Ground Truth Used
For the specific tests mentioned:
- Navigational Accuracy Comparative Study: The "ground truth" for navigation metrics (segmental, sub-segmental accuracy, on-target accuracy, distance to target) would likely be established through precise measurements against a known anatomical model or imaging data, but the document does not specify the exact method for this ground truth establishment.
- Animal and Cadaver Testing: Ground truth established through observation of the system's performance and impact on the tissues/anatomy, likely assessed by skilled medical professionals involved in the study.
- Usability/Human Factors Testing: Ground truth related to user needs, safety, and effectiveness is established through defined task completion rates, error rates, and user feedback against pre-defined success criteria.
- Other performance tests (biocompatibility, electrical safety, etc.): Ground truth is established by adherence to recognized international standards and engineering specifications.
8. The Sample Size for the Training Set
The document does not mention a training set size. This is expected as the Monarch Platform is described as a robotic surgical system with navigation, not a machine learning or AI diagnostic device that typically requires a large training set of annotated data for model development. The software verification and validation are mentioned, but this refers to traditional software development and testing, not AI/ML model training.
9. How the Ground Truth for the Training Set Was Established
As no training set is mentioned in the context of an AI/ML model, there is no information on how its ground truth was established.
Ask a specific question about this device
Page 1 of 1