510(k) Data Aggregation

    K Number
    K150265
    Date Cleared
    2015-09-03

    (211 days)

    Product Code
    Regulation Number
    892.5050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CNMC Model 206 is a medical dosimeter, or medical electrometer that is capable of performing measurements of diagnostic and therapeutic amounts of ionizing radiation when an appropriate ionization chamber or diode dosimeter is connected. These instruments are used exclusively by qualified personnel, typically medical physicists, for the calibration and quality control of medical equipment that produces ionizing radiation.

    The CNMC Model 206 is a dosimetry electrometer intended for measuring the output charge of an Ionization chamber or dosimetry diode that is used in a radiotherapy beam.

    Device Description

    The CNMC Model 206 is a medical dosimeter, or medical electrometer that is capable of performing measurements of diagnostic and therapeutic amounts of ionizing radiation when an appropriate ionization chamber or diode dosimeter is connected. These instruments are used exclusively by qualified personnel, typically medical physicists, for the calibration and quality control of medical equipment that produces ionizing radiation.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the CNMC Model 206 Electrometer/Dosimeter:

    General Note: The provided document is a 510(k) Premarket Notification Summary from the FDA. This document primarily focuses on establishing "substantial equivalence" to a predicate device, rather than providing a detailed report of a new clinical study with specific acceptance criteria and performance data for a novel device. Therefore, many of the requested items (e.g., sample sizes, ground truth establishment, expert qualifications, MRMC studies) are not typically found in this type of regulatory submission because the device is not a complex AI/ML system requiring extensive clinical validation. Instead, the focus is on demonstrating that the new device performs comparably to an already approved device.


    Description of Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria

    Summary: The CNMC Model 206 Electrometer/Dosimeter is primarily evaluated for "substantial equivalence" to a predicate device (Sun Nuclear Model 1010/206 – K002444). The acceptance criteria are implicitly defined by the specifications of the predicate device and the new device's ability to meet those specifications. The "study" demonstrating this involves comparing the technical specifications and intended use of the CNMC Model 206 to the predicate device.


    1. A table of acceptance criteria and the reported device performance

    Since this is a substantial equivalence submission, the "acceptance criteria" are the performance specifications of the predicate device, and the "reported device performance" is how the CNMC Model 206 matches or is equivalent to those specifications.

    ParameterAcceptance Criteria (Predicate Device: Model 1010/206 K002444)Reported Device Performance (CNMC Model 206)
    Charge:
    Range0.0001 pC to 1999.9 nC0.0001 pC to 1999.9 nC
    Resolution10 fC10 fC
    Current:
    Range0.001 pA to 1999.9 uA0.001 pA to 1999.9 uA
    Resolution1 fA1 fA
    Bias voltage-300/-150 to +300/+150-300/-150 to +300/+150
    Leakage Current