Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K962344
    Manufacturer
    Date Cleared
    1996-08-15

    (58 days)

    Product Code
    Regulation Number
    866.5680
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This in vitro diagnostic procedure is a solid phase immunoassay intended for the quantitative determination of Myoglobin in human serum or heparin plasma on the Technicon Immuno 1 system. When used in combination with other clinical data such as presenting symptoms and EKG values, measurement of Myoglobin aides in the early phase diagnosis of Myocardial Infarctions.

    Device Description

    The method described is an enzyme label sandwich assay using a monoclonal (mouse) capture and a polyclonal (goat) detector antibody. The monoclonal antibody is labelled with fluorescein and the polyclonal antibody labelled with alkaline phosphatase (ALP). The two reagents are the active compounds of the R1 and the R2 reagent, respectively. The solid phase consists of a suspension of magnetizable particles coated with antibody to fluorescein (mIMP reagent). Sample or calibrator, R1and R2 reagent and mIMP reagent are mixed simultaneously and incubated at 37 °C. In the presence of Myoglobin fluorescein-conjugate= Myoglobin=ALPa conjugate complex is formed and captured by the antiFluorescein antibodies on the magnetic particles. The particles are precipitated by an external magnetic field, washed and para-Nitrophenylphosphate is added as the enzyme substrate. The increase in absorbance due to the formation of p-Nitrophenolate is monitored spectrophotometrically at 405 and 450 nm. The concentration of Myoglobin in a sample. A Cubic Fit Through Zero is used to calculate the dose response curve. The assay is depicted schematically in fig. 1.

    AI/ML Overview

    This document describes the performance characteristics of the MYOGLOBIN METHOD FOR THE IMMUNO 1 SYSTEM.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical targets in the provided text. Instead, the document presents performance data to demonstrate the device's characteristics. The reported performance metrics are summarized below:

    Performance CharacteristicReported Device Performance
    Imprecision (Total CV)
    Sample 1 (14.8 ng/mL)5.5 %
    Sample 2 (52.6 ng/mL)3.6 %
    Sample 3 (75.6 ng/mL)3.9 %
    Sample 4 (131.3 ng/mL)3.6 %
    Sample 5 (247 ng/mL)4.4 %
    Sample 6 (278.1 ng/mL)3.0 %
    Sample 7 (639.7 ng/mL)3.6 %
    Sample 8 (1557.6 ng/mL)3.6 %
    Sample 9 (2718.9 ng/mL)3.4 %
    Correlation with Behring Nephelometer A
    Correlation Equation (y = Immuno 1, x = BNA)y = 1.02 × x + 1.05
    Slope (b)1.02 (Limits: 0.98 - 1.06)
    Intercept (a)1.05 (Limits: -3.2 - 4.5)
    Confidence of Correlation0.99314
    Interference (measured concentration as % of -2 pool)
    Bilirubin (25 mg/dL)100.3% - 101%
    Albumin (6.5 g/dL)100.9% - 101.5%
    Hemoglobin (1 g/dL)100.1% - 102%
    Gamma Globulins (5.3 g/dL)78.6% - 103.5% (The result for 1+ (95) and 2+ (103.5) are presented in a confusing manner, but the measured concentrations are within a reasonable range of the -2 pool)
    Triglycerides (supertrate)93.8% - 104.5%
    Heparin (65 IU/mL)99.9% - 101.5%
    Citrate (50 mg/mL)96% - 99%
    Urea and Creatine (200 mg/dL Urea, 2.5 mg/dL Creatine)99.8% - 101.2%
    Rheumatoid Factor (567 IU/mL)100.2% - 103.9%
    Linearity (deviation from calculated)
    Range-21% to 1.6%
    Sample Dilution (Recovery)
    Serum samples dilated with Immuno 1 Sample Diluent B93.9% - 108%
    Serum samples dilated with Immuno 1 Calibrator Level 193.2% - 107.2%
    Plasma samples dilated with Immuno 1 Sample Diluent B99.9% - 105.4%
    Plasma samples diluted with Immuno 1 Calibrator Level 188.9% - 111.8%
    Hook EffectNo erroneous results within the calibration range for Myoglobin content
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1