MULTI-CHANNEL COMPRESSION BTE (SD)
To amplify and transmit sound to the ear.
Dahlberg Hearing Aid with multi-channel compression circuit
This document is a 510(k) summary for a Dahlberg Hearing Aid. It primarily focuses on demonstrating substantial equivalence to a previously cleared device (K964530) rather than presenting a detailed clinical study with specific acceptance criteria and performance metrics typically seen for AI/ML or new clinical diagnostic devices.
Therefore, many of the requested categories are not applicable or cannot be extracted from the provided text. I will provide what information is available and indicate where information is missing or not relevant to this type of submission.
1. A table of acceptance criteria and the reported device performance
Based on the provided text, the "acceptance criteria" for a 510(k) submission like this are generally demonstrating substantial equivalence to a predicate device. The performance data is primarily technical and measured against industry standards rather than clinical outcomes.
Acceptance Criteria (Implied for 510(k) Substantial Equivalence) | Reported Device Performance |
---|---|
Technical Equivalence to Predicate Device: | |
- Identical intended use | Met: "To amplify and transmit sound to the ear." |
- Similar technological characteristics | Met: "Dahlberg utilizes the ANSI S3.22-1987 methodology to test and obtain performance data for hearing aids." |
- Similar materials | Met: "The hearing aid shells will be manufactured from materials that are commonly used in the industry." |
- Similar assembly methods | Met: "The hearing aid will be assembled utilizing similar assembly practices as Dahlberg's current products." |
Any differences do not raise new questions of safety and effectiveness. | (Implied by FDA clearance) |
Note: The document does not provide specific numerical performance data (e.g., gain, frequency response curves) but states that the methodology used is ANSI S3.22-1987, which is a standard for hearing aid measurement.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable/Not Provided. This 510(k) summary does not describe a clinical study with a "test set" of patient data. The evaluation for this type of device is primarily based on technical specifications and comparison to a predicate device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable/Not Provided. No clinical ground truth establishment is described in this document.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable/Not Provided. No clinical test set or adjudication process is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable/Not Provided. This device is a hearing aid, not an AI/ML diagnostic tool, so an MRMC study related to human reader performance with AI assistance is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not Applicable/Not Provided. This is a physical medical device (hearing aid), not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not Applicable/Not Provided (for clinical ground truth). The "ground truth" equivalent in this context is the adherence to ANSI S3.22-1987 standards for electroacoustic performance and the established safety and effectiveness of the predicate device.
8. The sample size for the training set
- Not Applicable/Not Provided. No "training set" in the context of AI/ML is relevant to this submission.
9. How the ground truth for the training set was established
- Not Applicable/Not Provided. No training set or associated ground truth establishment is relevant to this submission.