Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K152014
    Manufacturer
    Date Cleared
    2015-10-20

    (91 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MTS300 System is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterlization of the enclosed materials and maintain sterility for up to 30 days until used. The unit must be used with the MTS300 Transfer Cart, MTS300 filters and integrity locks.

    The MTS300 System is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270°F and exposure time of 4 minutes. Use no more than 3 trays per shelf or 25 lbs. per tray.

    The MTS300 System was tested and validated with rigid instruments containing lumens with an inner diameter of 3.8mm and an overall length of 370mm. Do not use with instruments containing lumens with an inner diameter smaller than 3.8mm and an overall length longer than 370mm.

    Use only uncovered, perforated or wire mesh general delivery trays within the MTS300 System.

    Device Description

    The MTS300 Multiple Tray Sterilization System (also referred to as the MTS300 System) includes the stainless steel sterilization cabinet and adjustable transfer cart. Single use filters and integrity locks with sterilization indicator dots are used with the system for each sterilization cycle. An optional STEAMPlus TM tray record card with STEAMPlus sterilization integrator (SPS Medical) may be utilized by the healthcare provider. The MTS System is intended to sterilize a maximum weight of 300 pounds of devices and instruments for a single patient surgery. The loaded cabinet is deployed from the adjustable transfer cart and processed in a pre-vacuum steam sterilizer and then dried for 30 minutes. Sterility is maintained for up to 30 days.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the MTS300 Multiple Tray Sterilization System, based on the provided text:

    Important Note: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. As such, it often references studies done to meet existing standards or guidance documents rather than detailing a single, comprehensive study testing a device against specific, novel acceptance criteria. The information below reflects what is extractable from this type of document.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Performance ClaimedReported Device Performance (from studies/validation)
    Sterilization EfficacyDemonstrated through verification studies in accordance with ANSI/AAMI ST77:2013 and FDA Draft Guidance (2002).
    Sterility Maintenance DurationUp to 30 days
    Compatibility with Prevacuum Steam SterilizersPrevacuum cycle of 270°F, 4 minutes exposure time
    Maximum Tray Load per ShelfNo more than 3 trays per shelf
    Maximum Weight per Tray25 lbs. per tray
    Lumen Instrument CompatibilityRigid instruments with inner diameter ≥ 3.8mm and overall length ≤ 370mm
    Tray Type CompatibilityUncovered, perforated, or wire mesh general delivery trays
    System Stability (Cabinet on Transfer Cart)Stable during deployment and use.
    User Needs/Intended Use (Usability)Meets user needs and intended use.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific "sample size" of items tested for the sterilization efficacy studies (e.g., number of sterilization cycles, number of biological indicators). It refers to "sterilization efficacy verification studies" performed.

    • Data Provenance: The studies were performed in accordance with ANSI/AAMI ST77:2013 and the recommendations of the Premarket Notification [510(k)] Submissions for Medical Sterilization Packaging Systems in Healthcare Facilities, Draft Guidance for Industry and FDA, March 7, 2002. This implies that the testing methodologies would follow standard industry practices and regulatory guidance, likely conducted in a controlled laboratory or healthcare setting. The document does not specify the country of origin of the data or whether it was retrospective or prospective, but given it's for a 510(k) submission, it would typically involve prospective testing conducted by the manufacturer or a contract research organization.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide details on the number or qualifications of experts used to establish ground truth. For sterilization efficacy, ground truth is typically established by laboratory methods (e.g., culturing biological indicators to confirm sterility), not usually by expert consensus in the typical sense of reading medical images.

    4. Adjudication Method for the Test Set

    Not applicable in the context of this device. Adjudication methods like 2+1 or 3+1 are typically used for interpreting results from human readers (e.g., radiologists) in diagnostic studies, not for the technical performance testing of a sterilization system. The verification studies follow established protocols where results (e.g., growth/no growth of biological indicators) are objective.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for medical imaging devices where human readers interpret results, comparing their performance with and without AI assistance. The MTS300 is a sterilization system, not an imaging or diagnostic device involving human interpretation in that manner.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in a sense. The "sterilization efficacy verification studies" and "engineering analysis" are standalone performance evaluations of the device itself (the "algorithm" in this context refers to the device's functional performance). There isn't an "AI algorithm" involved in the way typically discussed in diagnostic imaging. The device is intended to perform its function without constant human intervention during the sterilization cycle. The "usability study" involves human interaction but evaluates the system's ease of use and ability to meet human needs, not human diagnostic performance.

    7. The Type of Ground Truth Used

    For sterilization efficacy, the ground truth is primarily based on microbiological testing (e.g., culturing biological indicators to confirm the absence of viable microorganisms after the sterilization process). For stability and usability, the ground truth would be based on engineering analysis (meeting design specifications) and potentially user feedback/observation in the usability study demonstrating successful and safe operation.

    8. The Sample Size for the Training Set

    Not applicable. This device does not use machine learning or AI in a way that requires a "training set." Its performance is based on physical and chemical principles and validated through engineered design and testing.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1