Search Results
Found 1 results
510(k) Data Aggregation
(31 days)
The RAUMEDIC® MPR2 is indicated for use by qualified neurosurgeons or neurointensivists for measurement of intracranial pressure, temperature as well as to determine the partial oxygen pressure ptiO2 of the interstitial fluid.
The MPR2 is a small bedside monitor for measuring, displaying and monitoring of physiological parameters in the human brain which is used only in conjunction with RAUMEDIC ® -catheters. Analog signals can be relayed to a series-connected bedside monitor via its 2 analog outputs. Optionally, the device can be operated in stand-alone mode. The power is supplied by either the internal rechargeable battery or by an external, medically approved wide range wall plug transformer.
The MPR2 is a diagnostic device with physiological limit value monitor for measuring, displaying and monitoring the following physiological parameters: oxygen partial pressure (pO2), intracranial pressure (ICP) and brain temperature (T). These parameters are determined exclusively by connecting RAUMEDIC® catheters for single channel ICP measurement or with RAUMEDIC® multi-parameter catheters for combined measurement of ICP, pO2 and temperature (thermistor-based with YSI400 characteristic) or combinations derived from them such as pO2 and temperature. Optionally the pO2 and ICP signals can be relayed via the 2 analog outputs to a bedside monitor with limit value monitoring.
The provided document is a 510(k) Premarket Notification from the FDA regarding the RAUMEDIC® MPR2 logO DATALOGGER. This document is focused on demonstrating substantial equivalence to previously marketed predicate devices, rather than presenting a performance study against a defined set of acceptance criteria for a new AI/Machine Learning device.
Therefore, many of the typical acceptance criteria and study design elements requested for an AI/ML device (e.g., sample size for test/training sets, data provenance, number of experts for ground truth, MRMC studies, standalone performance, etc.) are not present in this type of regulatory submission.
Instead, the document focuses on:
- Intended Use/Indications for Use: Comparing the intended use of the new device to predicates.
- Technological Characteristics: Comparing features, materials, and form/function.
- Performance Testing: Stating that "extensive performance testing" was conducted and results showed the device is "safe for its intended use," but without providing specific test data, methodologies, or acceptance criteria in the format typically used for AI/ML device validation. The performance parameters listed in Table 1 (Accuracy of ICP Measurement, Temperature Measurement, etc.) are characteristics of the device itself and are compared to the predicate devices, serving as a form of "acceptance criteria" for substantial equivalence.
Given this context, I will extract the information that is available and clearly state what information is not present.
Acceptance Criteria and Device Performance (Derived from Substantial Equivalence Comparison)
The acceptance criteria are not explicitly stated as pass/fail thresholds in the traditional sense for a de novo device. Instead, they are implied through the comparison to predicate devices, where the new device's performance characteristics must be "substantially equivalent" or better. The reported device performance is listed alongside the predicate devices.
Table 1: Device Performance Comparison (Acting as Acceptance Criteria Check)
| Feature | Acceptance Criteria (Predicate) | Reported Device Performance (MPR2 logO) | Met? |
|---|---|---|---|
| Intended Use | RAUMEDIC® EASY logO: measurement of intracranial pressure, temperature, and ptiO2. Integra® Camino® ICP Monitor: measurement of intracranial pressure and temperature. | The RAUMEDIC® MPR2 is indicated for use by qualified neurosurgeons or neurointensivists for measurement of intracranial pressure, temperature as well as to determine the partial oxygen pressure ptiO2 of the interstitial fluid. | YES (Substantially Equivalent - see SE-Discussion 1) |
| ICP Measurement | YES (both predicates) | YES | YES |
| Pressure Measurement Range (ICP) | EASY logO: -40 to +100 mmHg. Integra® Camino®: -10 to 125 mmHg. | -20 to +100 mmHg | YES (Within physiological range & EASY logO's - see SE-Discussion 2) |
| Accuracy of ICP Measurement | EASY logO: ±1 mmHg (-20 to 20 mmHg), ±5% (<-20 mmHg, >20 mmHg). Integra® Camino®: ±2 mmHg (-10 to 10 mmHg), ±3 mmHg (11 to 33 mmHg), ±(6% + 1 mmHg) (34 to 125 mmHg). | • Accuracy -20 to 20 mmHg: ±1 mmHg for measuring instrument and ±1 mmHg for catheter • Accuracy <-20 mmHg and >20 mmHg: ±5% for measuring instrument and ±5% for catheter | YES (Similar to EASY logO, effectively equivalent to Integra Camino) |
| Temperature Measurement | YES (both predicates) | YES | YES |
| Temperature Measurement Range | 15°C - 45°C (both predicates) | 15°C - 45°C | YES |
| Accuracy of Temperature Measurement | EASY logO: Detailed range-specific accuracies similar to MPR2. Integra® Camino®: ±0.3 °C. | 15 ... 25°C: ±0.2 K (+sensor); 25 ... 45°C: ±0.1 K (+sensor); 37 ... 39°C: ± 0.05 K (+sensor); 39 ... 45°C: ± 0.1 K (+sensor) | YES (Similar to EASY logO, effectively equivalent to Integra Camino) |
| pO2 Measurement | EASY logO: YES. Integra® Camino®: NO. | YES | YES |
| pO2 Measurement Range | EASY logO: 0 to 150 mmHg. | 0 to 150 mmHg | YES |
| Accuracy of pO2 Measurement | EASY logO: ±3 % of measured value or ±2.5 mmHg (< 120 mmHg), < 10 % (120 mmHg to 150 mmHg). | ±3 % of the measured value or ±2.5 mmHg - the higher value is applicable for the partial oxygen pressure < 120 mmHg < 10 % of the measured value for the partial oxygen pressure 120 mmHg to 150 mmHg | YES |
| User Interface | EASY logO: alphanumeric display. Integra® Camino®: graphic display, touch screen. | Multi Parameter Monitor with graphic display, Membrane Buttons | YES (Equivalent or better) |
| Monitoring | Continuous (both predicates) | Continuous | YES |
| Data Storage | EASY logO: NO. Integra® Camino®: YES. | YES | YES (Equivalent or better) |
| Power Source | EASY logO: A/C wall outlet. Integra® Camino®: A/C wall outlet or D/C battery. | A/C wall outlet or D/C battery | YES |
| Battery Operating Time | Integra® Camino®: 1.5 hours. | Min. 4 hours | YES (Better - see SE-Discussion 3) |
| Analog Output to Bedside Monitor | EASY logO: 2 pcs. for ICP and ptO2. Integra® Camino®: 2 pcs. for ICP and Temperature. | 2 pcs. analog outputs for ICP and ptO2 | YES |
| Digital Interface | RS 232 (both predicates) | RS 232 | YES |
| Data Extraction | Integra® Camino®: USB 1.1. | USB 1.1 | YES (Equivalent or better) |
| Calibration | Auto Zero (both predicates) | Auto Zero | YES |
| Alarms | EASY logO: Technical Alarms only. Integra® Camino®: Technical and Physiological Alarms for ICP. | Technical Alarms and Physiological Alarms for ICP, Temperature and ptO2 | YES (Equivalent or better - see SE-Discussion 4) |
| Safety Measures (electrical) | Defibrillation-proof type CF applied parts (both predicates) | Defibrillation-proof type CF applied parts | YES |
| Safety Measures (functional) | EASY logO: Control system and 2 watch dog timers. Integra® Camino®: unknown (but assumed to have adequate safety) | Control system and 2 watchdog timers | YES |
| Accessories | Listed accessories for both predicates. | All cables and catheters listed in chapter 9 of the IFU zwo-459 (includes EASY logO accessories and previously cleared RAUMEDIC accessories). | YES (Substantially Equivalent - see SE-Discussion 5) |
| Case Material | Plastic (both predicates) | Plastic | YES |
Study Details (Information Not Provided/Applicable for this Type of Submission)
This document is a 510(k) submission for a non-AI/ML medical device, focusing on substantial equivalence to predicate devices. As such, the typical detailed study information requested for a novel AI/ML device is not included.
- Sample size used for the test set and the data provenance: Not applicable/not provided for this type of device. The performance testing mentioned is likely bench testing or pre-clinical validation, not clinical data for an AI model.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for physiological measurements is based on reference methods and calibration, not expert consensus on images.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI/ML device.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a measurement device, not an algorithm.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): For a medical device measuring physiological parameters (ICP, temp, ptiO2), "ground truth" is established through:
- Calibration Standards: Using known physical standards for pressure, temperature, and gas concentration.
- Reference Devices: Comparing measurements to established, highly accurate reference instruments.
- Controlled Bench Testing: Testing device accuracy and performance under controlled laboratory conditions, including simulated physiological ranges.
The document states "extensive performance testing" was done, and the specific accuracy values in the table are the results of such testing.
- The sample size for the training set: Not applicable, as this is not an AI/ML device.
- How the ground truth for the training set was established: Not applicable.
Ask a specific question about this device
Page 1 of 1