Search Results
Found 1 results
510(k) Data Aggregation
(591 days)
MOERAE SURGICAL MARKING PEN
The VasoPrep Surgical Marking Pen is intended for use prior to or during the harvesting and preparation of vein grafts used in bypass surgery. The pen is used to demarcate selected sites and orientation of the graft.
The VasoPrep Surgical Marking Pen is a single patient use sterile prescription use only marker intended for use on veins prior to or during Coronary-Assisted Bypass Graft (CABG) surgery. The marker (Figure 1) consists of a pen body, barrel, wick and cap with a wide chisel style applicator tip for delivery of ink to mark internal tissue. The formulation is non-toxic as used and is comprised of an ink material (FD&C Blue Dye #1) compounded into a carrier material (i.e., solvent). The wide chisel tip can deliver either a thin line of ink for precise marks or can be rotated 90° to deliver a wide stripe of ink.
This document describes a De Novo classification request for the VasoPrep Surgical Marking Pen, an internal tissue marker. The studies presented are non-clinical/bench studies, therefore, the concept of AI performance, ground truth establishment by experts, adjudication methods, multi-reader multi-case studies, and a standalone algorithm performance are not applicable.
1. Table of Acceptance Criteria and the Reported Device Performance:
The acceptance criteria and reported device performance are presented across several tables in the provided text. For clarity, they are consolidated and summarized below:
Test Category | Test Name | Purpose | Acceptance Criteria | Reported Device Performance |
---|---|---|---|---|
Biocompatibility | Cytotoxicity (ISO 10993-5) | To test and evaluate the cytotoxicity of the marker and ink formulation. | Non-cytotoxic | Non-cytotoxic |
Sensitization - Maximization Method (guinea pig) (ISO 10993-10) | To test and evaluate the potential for the marker and ink formulation to cause delayed contact sensitization. | No evidence of causing delayed dermal contact sensitization | No evidence of causing delayed dermal contact sensitization | |
USP Intracutaneous Reactivity (ISO 10993-10) | To test and evaluate the potential for the marker and ink formulation to cause local dermal irritant effects. | Nonirritant | Nonirritant | |
USP Systemic Toxicity | To test and evaluate the acute system toxicity of the marker and ink formulation. | No indications of systemic toxicity | No indications of systemic toxicity | |
Hemocompatibility (In vitro hemolysis) | To test and evaluate the hemocompatibility of the marker and ink formulation. | No significant hemolysis | No significant hemolysis | |
Material Mediated Pyrogenicity (ISO 10993-9) | To test and evaluate the pyrogenicity of the marker and ink formulation. | Non-pyrogenic | Non-pyrogenic | |
Shelf Life/Sterility/Packaging | Sterility Testing (ISO 11137) | To test and evaluate the sterility of the marker and ink formulation. | To ensure gamma radiation sterilization process is an adequate dose. Devices must have a sterility assurance of at least 10^-6. | Meets Acceptance Criteria |
Packaging Integrity (ASTM F1886/F1886M; ASTM F1929; ASTM F88/88M) | To test and evaluate the marking ability after undergoing accelerated aging and mechanical stress. | The packaging must pass the Visual Seal Examinations; Dye Leak Test; and Peel Test. | Meets Acceptance Criteria | |
Performance Testing – Bench | Internal Tissue Marking Ability | To test and evaluate device ability to mark human saphenous veins (HSV). | The marker shall provide a visible mark on wet or dry tissue that is 1-3 mm wide and up to 90 cm long with a single swipe. The mark shall remain visible on tissue for at least 4 hours. | Meets Acceptance Criteria (The marker provided a visible mark on wet or dry tissue that is 1-3 mm wide and up to 90 cm long with a single swipe. The mark remained visible on tissue for at least 4 hours.) |
Effect of Dye on Human Vein Tissue | To test and evaluate for patency effects caused by the ink on HSV. | Ex-vivo exposure to ink shall have no detrimental effect on the viability, smooth muscle contractility and endothelial-dependent relaxation of human saphenous vein grafts. | Meets Acceptance Criteria (The VasoPrep Surgical Marking Pen met all design requirements for compatibility and functional use.) | |
Effect of Dye on Animal Vein Tissue | Preliminary dosing experiments to test and evaluate demarcation ability of the ink on porcine saphenous veins, at a dose that will have no detrimental effect on further HSV testing. | Application shall demonstrate ink demarcation ability on porcine saphenous veins at an amount that would have no detrimental effect on the viability, smooth muscle contractility and endothelial-dependent relaxation of human saphenous vein grafts. | Meets Acceptance Criteria | |
Product Stability | Functionality (ASTM F1980) | To test and evaluate the marking ability after undergoing real time aging and shipping stress. | The marker shall provide a visible mark on wet or dry tissue that is 1-3 mm wide and up to 90 cm long with a single swipe. The mark shall remain visible on tissue for at least 4 hours. | Meets Acceptance Criteria (sterilized aged marking pens functioned as designed) |
Cytotoxicity (ASTM F1980; ISO 10993-5) | To test and evaluate the cytotoxicity of the marker and ink formulation after undergoing accelerated aging. | The marker and contents shall be non-cytotoxic. | Meets Acceptance Criteria (ink is non-cytotoxic) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Human Saphenous Vein (HSV) segments: Used for "Internal Tissue Marking Ability" and "Effect of Dye on Human Vein Tissue" studies. The exact number of segments is not specified, but it refers to "segments" which implies a limited number for individual tests. The origin of the human tissue is not specified (e.g., country of origin). These are ex-vivo experiments (retrospective in the sense that the tissue has been harvested, but prospective for the purpose of the experiment itself).
- Porcine Saphenous Vein: Used for "Effect of Dye on Animal Vein Tissue" and product stability testing. The exact number of veins or samples is not specified. The origin is not specified. These are ex-vivo or animal model experiments.
- Animals for Biocompatibility (Guinea Pig): Used for Sensitization testing. A "guinea pig method" is mentioned, implying a standard number of animals for such a test (typically a small cohort).
- Sterilized and Aged Markers: Four markers were tested for product stability functionality, and two real-time aged markers were used for demarcation effectiveness and fluid volume.
- Toxicological Assessment Literature Studies: Canine, porcine, and murine models were used in literature studies for lifetime toxicity/carcinogenicity of FD&C Blue #1 dye. The number of animals in these literature studies is not specified in this document.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
Not applicable. This is a non-clinical/bench study. There is no mention of experts establishing a "ground truth" in the context of human interpretation or diagnostic accuracy. Performance criteria are based on objective physical, chemical, and biological measurements.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. This is a non-clinical/bench study. Adjudication methods are typically used in studies involving subjective human interpretation, which is not the case here.
5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is a non-clinical/bench study for a physical device (marking pen), not an AI-powered diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is a non-clinical/bench study for a physical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" in these studies is based on objective scientific measurements and established standards for biocompatibility (e.g., non-cytotoxic, non-irritant), sterility (sterility assurance level), packaging integrity (visual, dye leak, peel tests), and performance characteristics (visible mark, no detrimental effect on tissue viability). It's a scientific and technical "ground truth" established through laboratory testing.
8. The sample size for the training set:
Not applicable. There is no AI or machine learning component, therefore no training set.
9. How the ground truth for the training set was established:
Not applicable. There is no AI or machine learning component, therefore no training set or ground truth for it. Any "training" or optimization of the pen's design or ink formulation would have been part of the product development process, not a formal training set as understood in AI/ML contexts.
Ask a specific question about this device
Page 1 of 1