MODIFICATION TO PENTRA 60 AND PENTRA 60C+
The ABX PENTRA 60 and 60C+ Hematology Analyzer are fully automated (microprocessor controlled) hematology analyzer used for the in vitro diagnostic testing of whole blood specimens.
The ABX PENTRA 60 and 60C+ Hematology Analyzer are able to operate either in complete blood count (CBC) mode or in CBC + 5 differential leucocyte count (5DIFF) mode.
The ABX PENTRA 60 & PENTRA 60C+ are benchtop, clinical laboratory instrument which analyzes in-vitro samples of whole blood to provide complete blood count and leucocyte differential count using principles of cytochemistry, focused flow impedance and light transmission using a halogen light source.
The PENTRA 60 is an standalone device using a keyboard and LCD screen integrated into the device. Whereas the PENTRA 60C+ is the PENTRA 60 connected to a dedicated workstation with user interfacing software.
The ABX PENTRA 60 and PENTRA 60C+ Hematology Analyzers underwent a study to demonstrate their performance against pre-defined acceptance criteria, primarily focusing on expanded linearity ranges and maintaining accuracy.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Parameter | Acceptance Criteria | Reported Device Performance |
---|---|---|
Precision | In accordance with EP5-A (NCCLS guidelines) | Good precision for all parameters, "entirely acceptable". |
Linearity Range (WBC) | Not explicitly stated as a numerical range acceptance criterion, but the modification covers "extension of the linearity ranges for all parameters." | 0-120 x 10^3/μL |
Linearity Range (RBC) | Not explicitly stated as a numerical range acceptance criterion, but the modification covers "extension of the linearity ranges for all parameters." | 0 - 8.0 x 10^5/μL |
Linearity Range (HGB) | Not explicitly stated as a numerical range acceptance criterion, but the modification covers "extension of the linearity ranges for all parameters." | 0 - 24g/dl |
Linearity Range (HCT) | Not explicitly stated as a numerical range acceptance criterion, but the modification covers "extension of the linearity ranges for all parameters." | 0 - 67% |
Linearity Range (PLT with Hgb>2g/dl) | Not explicitly stated as a numerical range acceptance criterion, but the modification covers "extension of the linearity ranges for all parameters." | 0 - 1,900 x 10^3/μL |
**Linearity Range (PLT with Hgb 0.95 | ||
Accuracy (Inter-procedural Correlation for other parameters) | Not explicitly stated, but implies historical good correlation from previous submissions. | Good correlation demonstrated in previous submissions. |
Carry-over |