Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K012458
    Manufacturer
    Date Cleared
    2001-08-14

    (13 days)

    Product Code
    Regulation Number
    870.3680
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    MODEL 6492 UNIPOLAR TEMPORARY ATRIAL PACING LEAD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Model 6492 Unipolar Temporary Atrial Pacing Lead is designed for temporary atrial pacing and sensing for a contemplated implant duration of 7 days or less. The device is supplied sterile and intended for single use only.

    Device Description

    The Medtronic Model 6492 Unipolar Temporary Atrial Pacing Lead consists of an electrode and an insulated multi-filament conductor which are crimped together. A blue monofilament proximally coiled for fixation of the lead is attached to the electrode and terminates distally in an atraumatic myocardial curved needle. A blue monofilament coil provides fixation while the lead is implanted in myocardial tissue. An atraumatic chest needle at the proximal end of the conductor wire permits exiting the pacing lead through the chest wall. To remove the pacing lead, gentle traction should be applied. No part of the lead remains in the body.

    AI/ML Overview

    This document describes a Special 510(k) Premarket Notification for the Medtronic Model 6492 Unipolar Temporary Atrial Pacing Lead. This type of submission is for modifications to a legally marketed device where the modifications do not affect the intended use or alter the fundamental scientific technology. Therefore, the submission focuses on demonstrating that the modified device maintains substantial equivalence to the predicate device through specific testing. It does not contain an "acceptance criteria" table in the traditional sense for a new device's performance, but rather a "summary of studies" showing that the modified device still meets all specified requirements.

    Here's an analysis of the provided text in relation to your request:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a formal table of "acceptance criteria" and "reported device performance" as would be seen for a new device's clinical efficacy study. Instead, it states that "All system compatibility tests performed have demonstrated that the modified Model 6492 heartwire meets the specified requirements." This implies that the acceptance criteria are inherent in the "specified requirements" for each test.

    Acceptance CriteriaReported Device Performance
    Environmental Conditioning: Meets specified requirements.Demonstrated that the modified Model 6492 heartwire meets the specified requirements.
    Visual Verification: Meets specified requirements.Demonstrated that the modified Model 6492 heartwire meets the specified requirements.
    Electrical Testing: Meets specified requirements.Demonstrated that the modified Model 6492 heartwire meets the specified requirements.
    Mechanical Testing: Meets specified requirements.Demonstrated that the modified Model 6492 heartwire meets the specified requirements.
    Sterilization Validation: Achieves required sterility assurance level (SAL) through 100% Ethylene Oxide (ETO) process.Processes appropriate for sterilizing the devices were validated.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample size for the "system compatibility testing" (Environmental Conditioning, Visual Verification, Electrical Testing, Mechanical Testing) or for the Sterilization Validation. It also does not mention the data provenance (country of origin, retrospective/prospective). This is common for Special 510(k) submissions focusing on engineering and bench testing rather than clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device submission relies on engineering and bench testing, not expert-derived ground truth from human observations or interpretations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This submission concerns physical device testing, not assessment of human interpretation or diagnostic accuracy.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/software device and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/software device. The testing described (Environmental Conditioning, Visual Verification, Electrical Testing, Mechanical Testing, Sterilization Validation) represents the "standalone" performance of the physical device in meeting its engineering specifications.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the engineering and sterilization tests, the "ground truth" is defined by the specified requirements and validated test methods. For example, in electrical testing, the ground truth would be the expected electrical parameters as per the device's design specifications. For sterility, the ground truth is a validated Sterility Assurance Level (SAL).

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1