Search Results
Found 2 results
510(k) Data Aggregation
(86 days)
The Microlife Automatic Blood Pressure Monitor, Model BP 3 AC1-1 P C i s a d evice intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive MAM (Microlife Average Mode) technique in which an inflatable cuff is wrapped around the upper arm.
The device detects the appearance of irregular heartbeat during measurement, and gives a warning signal with the reading once the irregular heartbeat is detected.
Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC, is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive MAM (Microlife Average Mode) technique in which an inflatable cuff is wrapped around the Upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic capacitive pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method".
The device detects the appearance of irregular heartbeat during measurement, and the arrhythmia symbol " device can be used in connection with your personal computer (PC) running the Microlife Blood Pressure Analyzer (BPA) software. The memory data can be transferred to the PC by connecting the monitor via cable with the PC.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC:
Disclaimer: This device submission is from 2006. Modern medical device regulations and testing standards, especially for AI/ML-based devices, are significantly more rigorous than what is presented here. This analysis is based solely on the provided text, which describes a relatively straightforward, non-AI medical device by today's standards. The "AI" mentioned in the questions is not explicitly covered in the document as it pertains to advanced machine learning, but rather refers to the "software algorithm" for oscillometric measurements.
Based on the provided 510(k) summary for the Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC:
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not explicitly state specific numerical acceptance criteria for blood pressure accuracy (e.g., mean difference and standard deviation compared to a reference standard). Instead, it states that the device's clinical performance "will remain unchanged" from its predicate device. The primary acceptance criteria appear to be compliance with relevant consensus standards and internal reliability tests.
Acceptance Criteria Category | Specific Criteria (Implicit/Explicit) | Reported Device Performance |
---|---|---|
Clinical Performance | Non-inferiority to predicate device (BP3BT0-AP) for systolic & diastolic blood pressure and pulse rate measurement accuracy. | "Clinical performance of the modified device will remain unchanged." |
Mechanical/Electrical | Compliance with FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, outlining Electrical, Mechanical and Environmental Performance Requirements. | "The Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC, meets all relevant requirements of the aforementioned tests." This implies acceptance criteria were met. |
Reliability | Successful completion of Storage, Operation, Vibration, Drop, and Life tests. | "None of the tests demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards." Implies compliance. |
EMC (Electromagnetic Compatibility) | Compliance with relevant EMC standards. | "EMC Test" performed; implied compliance with acceptance criteria. |
Software (PC-link) | Functionality of BPA version 3.1.6 and data transfer. | "PC-link software BPA version 3.1.6 test report" performed; implied successful operation. |
Voluntary Standards | Compliance with ANSI/AAMI Voluntary Standard, SP10-2002. | "as well as... the ANSI/AAMI Voluntary Standard, SP10-2002 testing results" - implies compliance. |
Irregular Heartbeat Detection | Detection and warning signal given with reading. | Stated as a function, and that it is present in both predicate and new device, implying it performs as intended. No specific accuracy metrics provided for this feature. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document explicitly states: "Clinical performance of the modified device will remain unchanged; therefore another clinical test for the modified device, BP3AC1-1PC is not required." This means no new clinical test set (or patient sample) was used for this modified device. The submission relies entirely on the clinical performance of the predicate device (Microlife Automatic Blood Pressure Monitor, Model BP3BT0-AP).
- Data Provenance: Not applicable for a new clinical test, as none was performed. For the predicate device, the provenance is not detailed in this submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. Since no new clinical tests were conducted for the BP3AC1-1 PC device, there was no new "ground truth" to establish for a test set. The submission relies on the established performance of the predicate device.
4. Adjudication Method for the Test Set
Not applicable, as no new clinical test set was used.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
- No MRMC study was done. This device is a standalone blood pressure monitor.
- No "AI" in the modern sense (machine learning) is described. The "software algorithm" mentioned refers to the standard oscillometric method used to derive blood pressure and pulse rate. The submission does not involve human readers interpreting AI outputs or any AI assistance to human tasks.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, the testing described is effectively a standalone performance assessment for the device itself. The device (algorithm + hardware) directly measures and displays blood pressure and pulse rate without human intervention in the measurement process (other than putting on the cuff and starting the device). The "clinical performance unchanged" statement refers to the device's standalone accuracy against a reference standard (likely auscultation) as established for the predicate device.
7. The Type of Ground Truth Used
For the predicate device's clinical validation (upon which this submission relies), the ground truth for blood pressure measurement would almost certainly have been established using expert auscultation simultaneous with the device's measurement, often with a double-blinded approach (e.g., using a Y-tube and two independent observers, with a third to break ties if necessary, and a mercury manometer as the reference standard). This standard methodology is typical for AAMI and BHS (British Hypertension Society) protocols.
8. The Sample Size for the Training Set
The concept of a "training set" as understood in modern machine learning is not applicable here. This device uses a pre-defined oscillometric algorithm, not a trainable AI model. Therefore, no distinct training set was used.
9. How the Ground Truth for the Training Set Was Established
As above, the concept of a training set is not applicable. The underlying algorithm (oscillometric method) is based on well-established physiological principles and signal processing, not on iterative training with a labeled dataset. The parameters of the oscillometric algorithm would typically be refined through engineering and calibration rather than machine learning training.
Ask a specific question about this device
(464 days)
The BP-2BHO Digital Blood Pressure Monitor is a device intended to measure the systolic, diastolic and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.
The BP-2BHO Digital Blood Pressure Monitor is designed to measure the systolic, diastolic and pulse rate of an individual by using a non-invasive technique which an inflatable curr is wrapper als manufactory of the upper arm. Our technique which an infiatable curr is wrapper als similar to the auscultatory method but uses an electronic capactive sensor rather than stethoscope and mercury manonical. - no onnalyzing those signals to define the cuff pressure to electrical signals, by andryzing the subscripmarket so called "oscillometric method".
The MicroLife Automatic Blood Pressure Monitor, Model BP-2BHO, was evaluated for both non-clinical and clinical performance to demonstrate its safety and effectiveness.
1. Table of Acceptance Criteria and Reported Device Performance
Test Category | Acceptance Criteria | Reported Device Performance |
---|---|---|
Non-Clinical Tests | ||
General Functions | Meets requirements of "Reviewer Guidance for Premarket Notification Submissions", DCRND (Nov 1993) | Met all relevant requirements |
Reliability (Operation Conditions) | Meets requirements of "Reviewer Guidance for Premarket Notification Submissions", DCRND (Nov 1993) | Met all relevant requirements |
Reliability (Drop Testing) | Meets requirements of "Reviewer Guidance for Premarket Notification Submissions", DCRND (Nov 1993) | Met all relevant requirements |
Reliability (Storage) | Meets requirements of "Reviewer Guidance for Premarket Notification Submissions", DCRND (Nov 1993) | Met all relevant requirements |
Reliability (Vibration Testing) | Meets requirements of "Reviewer Guidance for Premarket Notification Submissions", DCRND (Nov 1993) | Met all relevant requirements |
EMC | Meets requirements of "Reviewer Guidance for Premarket Notification Submissions", DCRND (Nov 1993) | Met all relevant requirements (EMC Test Report submitted) |
EN1060 Regulations | Meets EN1060 Regulations | Met all relevant requirements (EN1060 Test Report and CE Certification submitted) |
Clinical Tests | ||
Blood Pressure Measurement Accuracy | Meets relevant sections of ANSI/AAMI SP-10-1992 "National Standard for Electronic or Automated Sphygmomanometers" | Met all relevant requirements of the standard |
2. Sample Size and Data Provenance
The document does not explicitly state the sample size used for the test set or training set.
The data provenance (e.g., country of origin, retrospective/prospective) for the clinical study is not specified in the provided text.
3. Number of Experts and Qualifications for Ground Truth Establishment
The document does not specify the number of experts used to establish ground truth for the test set or their qualifications. The clinical testing was performed according to ANSI/AAMI SP-10-1992, which typically involves comparison against reference measurements from trained observers.
4. Adjudication Method
The document does not describe the adjudication method used for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No multi-reader multi-case (MRMC) comparative effectiveness study evaluating the improvement of human readers with AI vs. without AI assistance was reported. The device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool for human readers.
6. Standalone Performance Study
Yes, a standalone performance study was done for the device. The clinical testing against ANSI/AAMI SP-10-1992 demonstrates the algorithm-only performance of the MicroLife Automatic Blood Pressure Monitor.
7. Type of Ground Truth Used
For the clinical study, the ground truth was established by comparison to the "National Standard for Electronic or Automated Sphygmomanometers" (ANSI/AAMI SP-10-1992). This standard outlines procedures for accuracy testing, typically involving simultaneous measurements by the device and a trained human observer using a reference method (e.g., auscultation with a mercury sphygmomanometer). Therefore, the ground truth would be best described as expert consensus/reference method (as defined by the standard).
8. Sample Size for the Training Set
The document does not specify the sample size for the training set.
9. How Ground Truth for the Training Set was Established
The document does not provide details on how the ground truth for the training set was established. The device utilizes an "oscillometric method" based on a software algorithm to determine blood pressure. This implies that the algorithm itself might have been developed using various data, but the specifics of that development and ground truth establishment for training are not described in this summary.
Ask a specific question about this device
Page 1 of 1