    Why did this record match?
    Device Name :

    MICROBLAB, MICROLOOP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MicroLab / MicroLoop spirometer is intended, for prescription use only, to measure the maximal volume and flow of air that can be moved in and out of a patient's lungs and for pulse oximetry measurements. The system is intended for use with pediatric (4 to 17 years) and adult (18 to 99 years) patients in hospitals, physician offices, laboratories and occupational health testing environments.

    The optional Nonin IPOD® Integrated Pulse Oximetry Device is designed to measure pulse rate and oxygen saturation in adult patients. The sensor is designed for use on the fingers of patients weighing more than 30 kilograms, where the finger tissue is between 5 and 21 millimeters.

    Device Description

    The MicroLab / MicroLoop is a mains/battery operated desktop spirometer. It has context sensitive help screens, accessed at the touch of a button, that explain its features and navigational aides, making it easy to use. The results may be uploaded to a PC using the optional "Spirometry PC" software and patient details may be downloaded to the MicroLab / MicroLoop. Using spirometry PC software (SPCS) and the MicroLab / MicroLoop, live blows can be performed with the PC directly controlling the operation of the MicroLab / MicroLoop. The results and graphs produced are displayed directly on the PC screen. Stored data on the devices can be printed on an external printer using the USB cable supplied or uploaded to the PC. In addition the MicroLab is able to print the data on its integral thermal printer. Optional a Nonin Ipod® SpO2 sensor can be connected to the MicroLab / MicroLoop.

    The scientific concept which forms the basis of the MicroLab / MicroLoop is the CareFusion Digital Volume Transducer, a stable form of volume transducer, which measures expired air directly at B.T.P.S (Body Temperature and Pressure with Saturated water vapour) thus avoiding the inaccuracies of temperature corrections. The transducer is insensitive to the effects of condensation and temperature and avoids the need for individual calibration prior to performing a test.

    AI/ML Overview

    The provided document is a 510(k) Pre-market Notification for the MicroLab/MicroLoop spirometer and its optional SpO2 module. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a novel study to prove device meets acceptance criteria through specific quantitative performance metrics beyond what's stated for regulatory compliance.

    However, based on the Summary of Device Testing (Section 5.15) and the Summary Discussion of Bench Performance Data in the document, I can infer the acceptance criteria and study approach.

    Key takeaway: The submission emphasizes compliance with recognized standards rather than detailing specific de novo performance studies with novel clinical endpoints or expert ground truth.

    Here's the breakdown based on the information provided:


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Inferred from Standards)Reported Device Performance
    For Pulmonary Function (Spirometry):
    Accuracy: +/- 3% to ATS Recommendations (for volume and flow)"The proposed device passes the applicable tests and standards."
    Resolution: 0.01 liter"similar" to predicate
    Volume Range: 0.1 - 8 Litres"similar" to predicate
    Flow Range: 0.2 - 15 Litres/Second"similar" to predicate
    Compliance with ATS/ERS Standard of lung function testing"The proposed device passes the applicable tests and standards."
    For Oximetry Function (SpO2):
    Accuracy SpO2 70-100% (no motion, adults): +/- 2 digits"Accuracy SpO2 70-100%: - no motion (adults) +/- 2 digits"
    Accuracy SpO2 70-100% (motion, adults): +/- 3 digits"Accuracy SpO2 70-100%: - motion (adults) +/- 3 digits"
    Accuracy SpO2 70-100% (low perfusion, adults): +/- 3 digits"Accuracy SpO2 70-100%: - low perfusion (adults) +/- 3 digits"
    Heart Rate Accuracy (no motion, adults 18-300 BPM): +/- 3 digits"Heart Rate: - no motion (adults) (18 - 300 BPM) +/-3 digits"
    Heart Rate Accuracy (motion, adults 40-240 BPM): +/- 5 digits"Heart Rate: - motion (adults) (40 - 240 BPM) +/- 5 digits"
    Heart Rate Accuracy (low perfusion, adults 40-240 BPM): +/- 3 digits"Heart Rate: - low perfusion (adults) (40 - 240 BPM) +/- 3 digits"
    Compliance with Basic Safety (IEC 60601-1)"The proposed device passes the applicable tests and standards."
    Compliance with EMC Compatibility (IEC 60601-1-2)"The proposed device passes the applicable tests and standards."
    Compliance with Risk Management (ISO 14971)"The proposed device passes the applicable tests and standards."
    Compliance with Usability (EN 62366)"The proposed device passes the applicable tests and standards."
    Compliance with Software life cycle (ISO 62304)"The proposed device passes the applicable tests and standards."
    Compliance with Biocompatibility (ISO 10993-1)"The proposed device passes the applicable tests and standards."
    Measurement accuracy for the new oximetry module (Nonin IPOD®)"The proposed device passes the applicable tests and standards."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample sizes for the test sets used in the bench testing. It refers to "applicable tests and standards" and "validation and verification testing" but does not provide details on the number of subjects or samples for any specific performance measurement. The focus is on demonstrating compliance with standards rather than a clinical study with a specific patient cohort. The testing was non-clinical (bench testing) and likely conducted in Germany where the company is based.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. The testing described is primarily bench testing against established engineering and performance standards (e.g., ATS/ERS for spirometry, Nonin's own specifications for oximetry accuracy) and regulatory safety standards. There is no mention of human expert-derived ground truth as would be required for diagnostic image analysis or similar AI/ML-driven devices.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods typically relate to expert consensus on ground truth in a clinical study. The document describes non-clinical bench testing.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. The device is a diagnostic spirometer and pulse oximeter, not an AI-assisted diagnostic tool for interpretation by human readers. The submission explicitly states: "Clinical testing was not performed with this device."


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the performance testing described is standalone algorithm/device performance. The device provides direct measurements (spirometry parameters, SpO2, pulse rate). The "Summary of Device Testing" (Section 5.15) and "Summary Discussion of Bench Performance Data" (Section 11) describe non-clinical bench tests confirming the device's technical performance against established standards, independent of human interpretation or assistance beyond operating the device.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For spirometry, the ground truth would be precise, calibrated flow and volume simulators/generators that meet the reproducibility and accuracy requirements of standards like ATS/ERS.

    For oximetry, the ground truth for SpO2 and pulse rate accuracy would be from a reference oximeter or a controlled human desaturation study (though the document states "clinical testing was not performed," implying it relies on the Nonin module's pre-established accuracy or simulated scenarios). The "Accuracy Testing" for the oximetry module would compare its measurements to highly accurate reference instruments under various conditions.

    In essence, the ground truth is derived from validated engineering measurement standards and reference devices.


    8. The sample size for the training set

    Not applicable. This device does not appear to employ machine learning or AI that would require a "training set" in the conventional sense of AI development. It's a traditional medical device based on established physical principles (digital volume transducer for spirometry, light absorption for oximetry).


    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an AI/ML algorithm.