    K090710
    Device Name
    MERCURY CPAP
    Manufacturer
    Date Cleared
    2009-08-20

    (155 days)

    Product Code
    Regulation Number
    868.5965
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    MERCURY CPAP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mercury CPAP is intended to provide CPAP to spontaneously breathing patients in the hospital and pre-hospital environment.

    Device Description

    The Mercury CPAP is a respiratory aid device intended for use with a facemask and gas supplying device to elevate pressure in the patient's lungs. Alternatively the device may be used in conjunction with an endotracheal tube to generate and maintain constant positive airway pressure during standard intubation procedures.

    AI/ML Overview

    The Mercury CPAP device is considered substantially equivalent to its predicate device (Vygon Boussignac CPAP - K013884) based on comparative performance testing. The acceptance criteria are implicit in the comparison to the predicate device's performance characteristics.

    1. Table of Acceptance Criteria and Reported Device Performance

    FeatureAcceptance Criteria (based on predicate Vygon Boussignac CPAP - K013884)Reported Device Performance (Mercury CPAP)
    Indications for UseIdentical to predicate: Provide CPAP to spontaneously breathing patients in hospital and pre-hospital environment.Identical to predicate.
    TechnologySimilar technology used to generate CPAP pressure.Similar technology used to generate CPAP pressure.
    MaterialsMaterials in patient contact are identical to predicate devices.Materials in patient contact are identical to predicate devices.
    Environment of UseIdentical to predicate: Hospital and pre-hospital.Identical to predicate.
    Gas flow provided byWall gas or cylinder.Wall gas or cylinder.
    Requires a flow meter with ~30 Lpm rangeYes.Yes.
    Typical flow rate range10-25 lpm.10-25 lpm.
    CPAP pressure rangeUp to 10 cm H₂O.Up to 10 cm H₂O.
    In-line pressure manometerRecommended: a manometer be connected.Integral manometer (Mercury K954486).
    Excessive pressure reliefNo excessive pressure relief (for predicate).Integrated pop-off at 25 cm H₂O.
    Patient interfaceFace mask with head harness, connects to ET tube with adapter.Face mask with head harness, connects to ET tube without adapter.
    Single patient use, disposableYes.Yes.
    Contraindications and WarningsNone.None.
    Comparative Flow vs. CPAP PressureComparable to predicate results.Demonstrated substantial equivalence.
    Performance of pressure relief valveComparable to predicate (though predicate had none, Mercury CPAP tested its own integral valve).Demonstrated substantial equivalence (for the Mercury CPAP's integral valve).

    2. Sample size used for the test set and the data provenance

    The document states "We have performed comparative performance testing, Comparative Flow vs. CPAP Pressure" and "Performance of the pressure relief valve." However, specific sample sizes for these tests are not provided in the summary. The data provenance is also not explicitly stated in terms of country of origin or whether it was retrospective or prospective. It implies the testing was conducted by or for Mercury Medical, Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable or not provided. The study is a comparative performance test of a medical device (CPAP machine) against a predicate device, focusing on physical and performance characteristics like gas flow, pressure, and mechanical features. It does not involve human interpretation or subjective assessment that would require expert ground truth establishment in the typical sense (e.g., image interpretation by radiologists).

    4. Adjudication method for the test set

    This information is not applicable or not provided. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation where disagreements need to be resolved. This study focuses on device performance metrics.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study concerns the physical and performance characteristics of a CPAP device, not an AI or diagnostic imaging system that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, this is not an AI-driven device, and therefore the concept of "algorithm only" or "human-in-the-loop" performance is not relevant. The testing performed was on the physical device itself.

    7. The type of ground truth used

    The "ground truth" for this type of device comparison is the established performance characteristics and specifications of the predicate device (Vygon Boussignac CPAP - K013884), as well as engineering standards for CPAP functionality and safety. The Mercury CPAP's performance was compared against these established benchmarks.

    8. The sample size for the training set

    This information is not applicable. This is a hardware medical device and does not involve machine learning algorithms that require a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable as there is no training set for this type of device.