510(k) Data Aggregation

    K Number
    K043299
    Date Cleared
    2004-12-10

    (10 days)

    Product Code
    Regulation Number
    880.5725
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    MEDLEY PCA MODULE, MODEL 8120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medley PCA Module is intended for use in today's growing professional healthcare environment for facilities that utilize infusion devices for the delivery of medications or fluids. The Medley PCA Module is indicated for use on adults, pediatrics, and neonates for continuous or intermittent delivery through clinically acceptable routes of administration; such as intravenous (IV), subcutaneous, or epidural.

    The addition of the PCA/Monitoring Protocol provides an optional and hospitalconfigurable feature that is intended to align with healthcare facilities' current protocols that require monitoring of patients while on PCA therapy. All device programming, data entry and validation of PCA/Monitoring Protocol parameters is performed by the trained healthcare professional according to hospital-defined protocol or a physician's order.

    The Medley PCA Module with PCA/Monitoring Protocol is intended for use with patients that are prescribed PCA pain management therapy with opioid medications, specifically: Fentanyl, Demerol, Morphine, and Hydromorphone. The Medley PCA Module with PCA/Monitoring Protocol is intended for use by healthcare professionals in clinical environments. The Protocol does not replace clinician assessment or therapy decision making, but adds an additional safety net for the clinician at the point of care. All device programming, data entry and validation of the Medley PCA Module with PCA/Monitoring Protocol is performed by the trained healthcare professional according to hospital-defined protocol or a physician's order.

    Device Description

    Due to the potential for opioid-induced respiratory depression, hospitals require regular monitoring and assessment of patients prescribed opioid medications. By allowing for the combination of a PCA Module and monitoring module (EtCO2 Module and/or SpO2 module) on the same platform, the Medley System provides a readily available way for the clinician to continuously monitor the patient's respiratory response while receiving opioid infusion therapy.

    This submission will utilize the current system modularity to add a new hospitalconfigured Guardrails parameter (PCA/ Monitoring Protocol) to the Medley PCA Module. This Protocol is designed to help ensure patient safety when administering opioid medications. This Protocol will be available for use when a PCA Module and monitoring module(s) are attached to the same Medley Point of Care Unit (PCU). The PCA/Monitoring Protocol will be part of the Guardrails data set and will consist of a specified list of opioid medications and hospital-specified pre-configured monitoring limits. The Protocol will activate only when hospital-specified protocol indicates an unsafe state of respiratory depression is detected.

    AI/ML Overview

    The provided document discusses the Medley™ PCA Module with PCA/Monitoring Protocol and its substantial equivalence to a predicate device, rather than presenting a detailed study with acceptance criteria and performance data in the typical sense for a diagnostic or AI-powered device.

    This clearance is for a medical device (an infusion pump with a specific protocol feature), which undergoes different types of validation than, for example, an AI diagnostic tool. The "performance data" mentioned in the document refers to engineering and functional testing to ensure the device operates as intended and is safe, rather than clinical performance metrics like sensitivity, specificity, or AUC against a ground truth.

    Therefore, many of the requested categories (like sample size for test/training sets, number of experts for ground truth, MRMC studies, standalone performance with metrics like AUC/sensitivity/specificity) are not applicable or not detailed in this type of 510(k) summary for an infusion pump.

    However, I can extract what is available from the document regarding the "acceptance criteria" (implied functional requirements) and what the document states about "performance data".

    Here's an attempt to answer based on the provided text, highlighting what is available and what is not:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied / Functional Requirements)Reported Device Performance
    Functional Safety & Performance:
    • Ability to deliver continuous or intermittent medications/fluids.
    • Compatibility with various routes of administration (IV, subcutaneous, epidural).
    • Integration of a PCA Module and monitoring module (EtCO2 Module and/or SpO2 module) on the Medley System platform.
    • Activation of PCA/Monitoring Protocol based on hospital-specified parameters and detection of an "unsafe state of respiratory depression."
    • All device programming, data entry, and validation of parameters performed by trained healthcare professionals according to protocol/physician's order.
    • The Protocol acts as an "additional safety net" for clinicians without replacing assessment or decision-making.
    • The device must meet specified requirements for safety and effectiveness. | The document states:
      "The Medley PCA Module is intended for use... for continuous or intermittent delivery through clinically acceptable routes of administration; such as intravenous (IV), subcutaneous, or epidural."

    "By allowing for the combination of a PCA Module and monitoring module (EtCO2 Module and/or SpO2 module) on the same platform, the Medley System provides a readily available way for the clinician to continuously monitor the patient's respiratory response..."

    "This Protocol is designed to help ensure patient safety when administering opioid medications... The Protocol will activate only when hospital-specified protocol indicates an unsafe state of respiratory depression is detected."

    "All device programming, data entry and validation of PCA/Monitoring Protocol parameters is performed by the trained healthcare professional according to hospital-defined protocol or a physician's order."

    "The Protocol does not replace clinician assessment or therapy decision making, but adds an additional safety net for the clinician at the point of care."

    "The performance data indicate that the Medley™ PCA Module with PCA/Monitoring Protocol meets specified requirements, and is substantially equivalent to the predicate device." |

    2. Sample size used for the test set and the data provenance

    • Not explicitly stated in the document. The "performance data" mentioned would typically refer to engineering and functional testing of the device's hardware and software, rather than clinical trials with patient data. This is common for a 510(k) medical device clearance where substantial equivalence to a legally marketed predicate is established.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable/not explicitly stated. For a medical device like an infusion pump, "ground truth" in the diagnostic sense is not typically established by experts reviewing cases. Performance is usually assessed against engineering specifications and functional output.

    4. Adjudication method for the test set

    • Not applicable/not explicitly stated. Clinical adjudication, as seen in diagnostic studies, is not typically part of the clearance for an infusion pump's functional performance.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is an infusion pump with a safety protocol (not an AI diagnostic tool for image or data interpretation by human readers). Therefore, MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Partially applicable, but for device function, not diagnostic performance. The "algorithm" here refers to the PCA/Monitoring Protocol. Its standalone functionality (i.e., whether it correctly detects 'unsafe states' based on its programming and triggers alerts) would have been tested as part of the "performance data." However, its "performance" is not measured in terms of diagnostic metrics like AUC, sensitivity, or specificity against a clinical ground truth, but rather its adherence to design specifications and safety requirements. The document states it is "intended to align with healthcare facilities' current protocols."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not explicitly stated/not applicable in the traditional sense. For this device, the "ground truth" for validation would be its adherence to engineering specifications, safety standards, and its ability to correctly implement the defined "PCA/Monitoring Protocol" logic (e.g., detecting specified physiological parameters (EtCO2, SpO2) and acting in accordance with pre-configured monitoring limits). These are validated through functional and performance testing, not against clinical outcomes or expert consensus on patient cases.

    8. The sample size for the training set

    • Not applicable. This device is not an AI/ML model that requires a training set in the conventional sense. Its "protocol" is programmed logic, not learned from data.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.

    Summary of the Study (as described in the 510(k) submission):

    The document describes the Medley™ PCA Module with PCA/Monitoring Protocol and references a "performance data" section, which is typical for 510(k) submissions. This "performance data" is designed to demonstrate that the device meets its specified requirements and is substantially equivalent to its predicate device (Medley PCA Module, K032233).

    The study that proves the device meets the acceptance criteria (i.e., functional requirements and safety) is implicitly described as the "Performance Data" section of the 510(k) submission. This data would have involved:

    • Bench testing: Verifying the accuracy of fluid delivery, the functionality of the monitoring module integration (EtCO2, SpO2), and the correct execution of the PCA/Monitoring Protocol logic (e.g., proper activation in response to simulated "unsafe states" and adherence to pre-configured monitoring limits).
    • Software validation: Ensuring the software correctly implements the protocol and user interface.
    • Electrical safety and EMC testing: Adherence to relevant standards.

    The primary goal of this submission was to demonstrate Substantial Equivalence to a previously cleared device (Medley PCA Module, K032233) due to the addition of a new hospital-configurable Guardrails parameter (the PCA/Monitoring Protocol). The document states, "The results of this comparison demonstrate that the Medley™ PCA Module with PCA/Monitoring Protocol is equivalent to the marketed predicate device in technological characteristics." and "The performance data indicate that the Medley™ PCA Module with PCA/Monitoring Protocol meets specified requirements, and is substantially equivalent to the predicate device."

    This means the "study" involved comparing the new device's functional and safety performance against the established requirements and the predicate device's performance, ensuring the added protocol feature did not introduce new safety concerns or compromise existing functionality.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032233
    Date Cleared
    2003-09-09

    (50 days)

    Product Code
    Regulation Number
    880.5725
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    MEDLEY PCA MODULE, MODEL 8120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medley PCA Module is intended for use in today's growing professional healthcare environment for facilities that utilize syringe pumps for the delivery of medications or fluids.

    The Medley PCA Module is indicated for use on adults, pediatrics, and neonates for continuous or intermittent delivery through clinically acceptable routes of administration; such as, intravenous (IV), subcutaneous, or epidural.

    Device Description

    The Medley PCA Module functions as part of the Medley™ Medication Safety System. In combination with the Medley" Programming Module (PM), the PCA Module will deliver fluids in a manner similar to current PCA pumps on the market. The Medley" PCA Module uses standard non-dedicated, single-use, administration sets and syringes with luer-lock connectors, of type designed for use on syringe-type PCA pumps.

    AI/ML Overview

    The provided document is a 510(k) summary for the ALARIS Medical Systems® Medley™ PCA Module. It states that the device is substantially equivalent to predicate devices and meets specified performance requirements. However, it does not provide specific acceptance criteria or detailed results from a study proving these criteria are met.

    Therefore, I cannot populate the table or answer most of the questions using the information given. The document explicitly states: "The performance data indicate that the Medley PCA Module meets specified requirements, and is substantially equivalent to the predicate devices," but it does not detail those requirements or the specific performance data.

    Here's what can be inferred or stated based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not specified in documentMeets specified requirements (specific criteria not detailed)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for test set: Not specified.
    • Data provenance: Not specified (country of origin or retrospective/prospective nature of any studies).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Not specified, as no detailed study or ground truth establishment is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No mention of an MRMC study, AI, or human reader improvement. This device is a PCA infusion pump, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable/Not specified. This is a medical device (infusion pump), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable/Not specified, as no detailed study or ground truth establishment is described. The performance data would likely relate to engineering specifications for pump accuracy, safety features, etc.

    8. The sample size for the training set

    • Not applicable/Not specified, as no machine learning or AI training is described.

    9. How the ground truth for the training set was established

    • Not applicable/Not specified.