Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K231617
    Date Cleared
    2023-11-09

    (160 days)

    Product Code
    Regulation Number
    892.1000
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    MAGNETOM Free.Max; MAGNETOM Free.Star

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MAGNETOM Free.Max system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal, and oblique cross-sectional images that display the internal structure and/or function of the head. body, or extremities. Other physical parameters derived from the images may also be produced. Depending on the region of interest, contrast agents may be used. These images and the physical parameters derived from the images when interpreted by a trained physician vield information that may assist in diagnosis.

    MAGNETOM Free.Max may also be used for imaging during interventional procedures when performed with MR-compatible devices such as MR Safe biopsy needles.

    MAGNETOM Free.Star system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal, and oblique cross-sectional images that display the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images may also be produced. Depending on the region of interest, contrast agents may be used. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist in diagnosis.

    Device Description

    MAGNETOM Free.Max and MAGNETOM Free.Star with syngo MR XA60A include new and modified features compared to the predicate devices MAGNETOM Free.Max and MAGNETOM Free.Star with syngo MR XA50A (K220575, cleared on June 24, 2022).

    Below is a high-level summary of the new and modified hardware and software features compared to the predicate devices MAGNETOM Free.Max and MAGNETOM Free.Star with syngo MR XA50A:

    Hardware
    New hardware features:

    • Contour Knee coil
    • Respiratory Sensor

    Modified hardware features:

    • myExam 3D Camera
    • Host computer
    • MaRS

    Software
    New Features and Applications:

    • Injector coupling
    • Respiratory Sensor Support
    • myExam RT Assist (only for MAGNETOM Free.Max)
    • myExam Autopilot Hip
    • Deep Resolve Boost
    • Complex Averaging
    • HASTE_Interactive (only for MAGNETOM Free.Max)
    • BEAT_Interactive (only for MAGNETOM Free.Max)
    • Needle Intervention AddIn (only for MAGNETOM Free.Max)

    Modified Features and Applications:

    • Deep Resolve Sharp
    • Deep Resolve Gain
    • SMS Averaging

    Other Modifications:

    • Indications for Use (only for MAGNETOM Free.Max)
    • MAGNETOM Free.Max RT Edition marketing bundle (only for MAGNETOM Free.Max)
    AI/ML Overview

    The provided text describes information about the submission of the MAGNETOM Free.Max and MAGNETOM Free.Star MRI systems for FDA 510(k) clearance, and references a specific AI feature called "Deep Resolve Boost." However, it does not contain acceptance criteria or a detailed study proving the device meets specific performance criteria for the AI feature.

    The section titled "Test statistics and test results" for Deep Resolve Boost (Table 1, page 7) mentions that the impact of the network was characterized by quality metrics such as PSNR and SSIM, and visual inspection. It also states: "After successful passing of the quality metrics tests, work-in-progress packages of the network were delivered and evaluated in clinical settings with cooperation partners." This suggests internal testing and evaluation, but does not provide the specific numerical acceptance criteria or the detailed results of these tests.

    Therefore, I cannot fully complete the requested table and answer all questions due to the lack of this specific information in the provided document.

    However, I can extract the available information regarding the AI feature "Deep Resolve Boost" as much as possible:

    1. Table of acceptance criteria and the reported device performance:

    Metric / CriteriaAcceptance Criteria (Stated or Implied)Reported Device Performance (Specifics not provided in document)
    Deep Resolve Boost
    Peak Signal-to-Noise Ratio (PSNR)Must pass initial quality metrics tests.Quantified, but specific numerical values are not reported.
    Structural Similarity Index (SSIM)Must pass initial quality metrics tests.Quantified, but specific numerical values are not reported.
    Visual Inspection for ArtifactsMust ensure potential artifacts are detected that are not well captured by PSNR/SSIM.Images visually inspected.
    Clinical EvaluationMust be evaluated in clinical settings with cooperation partners."work-in-progress packages of the network were delivered and evaluated in clinical settings with cooperation partners." (No specific results or findings reported in this document.)

    2. Sample size used for the test set and the data provenance:

    • Test Set (Validation set for AI feature Deep Resolve Boost):
      • Sample Size: 1,874 2D slices.
      • Data Provenance: "in-house measurements and collaboration partners." The document does not specify the country of origin.
      • Retrospective or Prospective: Retrospectively created from ground truth by data manipulation and augmentation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of experts: Not specified.
    • Qualifications of experts: The document states the "acquired datasets represent the ground truth for the training and validation," but it does not specify how this ground truth was established in terms of expert involvement for the test set. It mentions "clinical settings with cooperation partners" for evaluation, but this is distinct from ground truth establishment.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not specified. The document states "acquired datasets represent the ground truth," suggesting pre-existing data or a different method of ground truth establishment than explicit reader adjudication for this AI feature.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • The document states "No clinical tests were conducted to support substantial equivalence for the subject device" (page 10). It mentions that "work-in-progress packages of the network were delivered and evaluated in clinical settings with cooperation partners," but this is not described as an MRMC comparative effectiveness study, nor are any results on human reader improvement reported.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • The performance of the "Deep Resolve Boost" AI feature was characterized by "quality metrics such as peak signal-to-noise ratio (PSNR) and structural similarity index (SSIM)" and visual inspection, which suggests a standalone evaluation of the algorithm's output against a reference standard. Specific results are not provided.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For Deep Resolve Boost: "The acquired datasets represent the ground truth for the training and validation." Input data for training was "retrospectively created from the ground truth by data manipulation and augmentation." This implies that high-quality, likely clinical-grade, MRI scans acquired without the AI feature were considered the "ground truth" to which the AI-processed images were compared. It's not explicitly stated if this "ground truth" itself was established by expert consensus, but it infers it from high-quality clinical acquisition.

    8. The sample size for the training set:

    • For Deep Resolve Boost: 24,599 2D slices.

    9. How the ground truth for the training set was established:

    • "The acquired datasets represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation and augmentation. This process includes further under-sampling of the data by discarding k-space lines, lowering of the SNR level by addition of noise and mirroring of k-space data."
    • This indicates that "ground truth" was established by using full, high-quality MR images. The "input data" for the AI model (which the AI then "boosts") was intentionally degraded (under-sampled, noised) from this high-quality ground truth. The AI's task is to reconstruct the degraded input data back to resemble the original high-quality "ground truth."
    Ask a Question

    Ask a specific question about this device

    K Number
    K220575
    Date Cleared
    2022-06-24

    (116 days)

    Product Code
    Regulation Number
    892.1000
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    MAGNETOM Free.Max, MAGNETOM Free.Star

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGNETOM MR system is indicated for use as a magnetic resonance diagnostic device (MRDD), which produces transverse, sagittal, coronal, and oblique cross sectional images that display the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images may also be produced. Depending on the region of interest, contrast agents may be used.

    These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist in diagnosis.

    Device Description

    The subject device software version, syngo MR XA50A, can support the following two MRI systems:

    • MAGNETOM Free.Max, which has been cleared with its initial software version syngo MR XA40A, through K210611 on July 1, 2021;
    • MAGNETOM Free.Star, a new product.

    With the introduction of MAGNETOM Free.Star, we extend the Free. platform, which consists of two products with a field strength of 0.55 Tesla on our high-value MRI platform. The main difference between these two products is the bore size, MAGNETOM Free.Star is equipped with a 60 cm patient bore while the MAGNETOM Free.Max is equipped with an 80 cm patient bore. The Gradient system, body coil and the system cover for MAGNETOM Free.Star are modified based those of the predicate device MAGNETOM Free.Max with syngo MR XA40A (K210611) to accommodate the smaller bore diameter. The other main components for the new device MAGNETOM Free.Star are the same as those of MAGNETOM Free.Max as cleared with K210611.

    Apart from the hardware adaption applied for MAGNETOM Free.Star for the smaller bore diameter, the new / modified hardware and software features for the subject devices comparing to the predicate device MAGNETOM Free.Max with software version syngo MR XA40A (K210611, cleared on July 1, 2021) are listed below:
    MAGNETOM Free.Max software version syngo MR XA50A
    New/Modified Hardware
    Common for both subject devices:

    • Scanner User Interface (SUI): introduce option to have two SUI set on both sides of the scanner, while there is only one set available on the left hand side as the standard configuration for the predicate device MAGNETOM Free.Max with software version syngo MR XA40A (K210611); Swap the orientation of patient pictogram on Select&GO is supported syngo MR XA50A.
    • myExam 3D Camera: auto registration with detection of patient height. weight and orientation are supported in the subject device software version syngo MR XA50A. The hardware remains unchanged as cleared with K210611 on July 1, 2021.
    • New Patient Video: A new patient video with 1920×1080 pixels is introduced.

    Applicable to the following subject device(s)
    MAGNETOM Free.Max
    MAGNETOM Free.Star
    New Local Coils
    Contour M Coil
    Contour M Coil
    New Patient table – High-load patient table: a new fixed patient table with vertical movement for heavy load patient is introduced.
    N/A

    Software Features
    Common for both subject devices:
    New Software Platform/Workflow
    myExam Autopilot is extended the supporting body region to shoulder:

    • myExam Shoulder Autopilot: it helps users to automate a shoulder examination.
      Migrated Software feature
    • EP2D FID: Single-shot FID EPI pulse sequence type optimized for perfusion-imaging in the brain.
    • Inline Perfusion: Automatic real-time calculation of parameter maps with Inline technology based on image data acquired with the ep2d fid pulse sequence type.
    • Access-i: Provides an interface to enable the connection of a 3rd party workstation to the MR syngo Acquisition Workplace via a network router and secure local network connection.

    Modified Software Platform/Workflow
    Modify in Scan assistance: Modified guidance of off-center imaging is provided to users who encounter the scan suspension by too off-centered shim volume.

    AI/ML Overview

    The provided text is a 510(k) summary for the MAGNETOM Free.Max and MAGNETOM Free.Star with syngo MR XA50A. This document asserts substantial equivalence to a predicate device (MAGNETOM Free.Max with syngo MR XA40A) and does not describe specific acceptance criteria or a study designed to prove the device meets those criteria in detail as requested.

    Instead, it states that:

    • "The results from each set of tests demonstrate that the devices perform as intended and are thus substantially equivalent to the predicate device to which it has been compared."
    • "No clinical tests were conducted to support substantial equivalence for the subject device; however, as stated above, sample clinical images were provided."
    • "The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device."

    Therefore, it is not possible to provide the specific details requested in your prompt based solely on the provided text, as this document focuses on demonstrating substantial equivalence to a previously cleared device through non-clinical testing, rather than reporting on a study with detailed acceptance criteria for standalone performance or comparative effectiveness.

    The document mentions non-clinical tests conducted, which include:

    • Sample clinical images: To evaluate coils, new and modified software features, and pulse sequence types.
    • Performance bench test: For SNR and image uniformity measurements for coils, and heating measurements for coils.
    • Software verification and validation: Mainly for new and modified software features.

    These tests were conducted in accordance with guidance documents like "Guidance for Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices" and "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

    In summary, the provided document does not contain the specific information requested in your prompt regarding detailed acceptance criteria and a study proving the device meets them, sample sizes, expert qualifications, adjudication methods, or MRMC study results for human reader improvement. The document focuses on demonstrating substantial equivalence through non-clinical testing against established standards and predicate devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1