Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K153470
    Manufacturer
    Date Cleared
    2016-08-19

    (261 days)

    Product Code
    Regulation Number
    868.5170
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MADgic™ Laryngo-Tracheal Mucosal Atomization Device is intended for the application of topical anesthetics to the oropharynx and upper airway region.

    Device Description

    The MADgic™ Laryngo-Tracheal Mucosal Atomization Device is a disposable nonsterile device that converts a solution of topical anesthetic into a fine particle spray for application to mucosal surfaces. The device consists of an atomizer tip, a semirigid tubular extension, a standard luer lock adapter and a syringe. The clinician draws up the desired volume of topical anesthetic into the syringe, attaches it to the luer lock fitting of the atomizer, and manipulates the tubing extensions into the desired position. As the syringe plunger is compressed, the anesthetic is forced into the tubular extension and out of the atomizer tip. The tip takes the pressurized fluid column and begins spinning it, allowing the fluid to exit the small hole at the end in a cone shaped spray. The anesthetic mist is gently distributed onto the mucosal surface in front of this tip. Topical anesthesia application can begin in the mouth and pharynx, and then proceed to the hypopharynx, epiglottis and vocal cords, larynx and trachea via direct visualization using a laryngoscope.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the MADgic™ Laryngo-Tracheal Mucosal Atomization Device:

    Device: MADgic™ Laryngo-Tracheal Mucosal Atomization Device
    FDA 510(k) Number: K153470
    Predicate Device: MADgic™ Laryngo-Tracheal Mucosal Atomization Device (K002255)

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Biocompatibility Testing
    Cytotoxicity - L929 MEM Elution AssayGrade of 0, 1, or 2 (not more than 50% of cells are round, devoid of intracytoplasmic granules, and no extensive cell lysis)Acceptable
    Sensitization - Kligman Maximization AssayDifference between test article mean score and vehicle control mean score is 1.0 or less (non-irritant).Acceptable
    Irritation - Intracutaneous Injection AssayGrade of 1, 0 or less using the Kligman scoring system.Acceptable
    Performance Data (Non-Clinical)
    Visual InspectionProper assembly, presence of all components, no cracks in tip and luer, no short shots in molded components, solvent presence around bonding surface covering minimum one third total area, no gaps between insert post and tubing.Pass
    Flow TestFlow rate ≥ 225 sccm and ≤ 700 sccm.Pass
    Leak TestLeak value ≤ 0.05 psi.Pass
    Hydrostatic TestDevice must remain assembled after being subjected to a minimum 300 psi hydrostatic pressure.Pass
    Atomization Test Post HydrostaticNo streaming, no excessive dripping, or occlusions upon atomization.Pass
    TensileTensile strength ≥ 6.0 lb.Pass
    Specification Verification (Not direct performance, but conformance)
    Typical Particle Size30-100 micronsN/A. Specification Study.
    System Dead SpaceMAD600/600OS = 0.24mL; MAD700-730/730OS = 0.18mL; MAD720 = 0.12mLN/A. Specification Study.
    Tip Diameter0.19 inches (4.8 mm)N/A. Specification Study.
    Applicator LengthMAD600/6000S/700 = 8.9 inches; MAD720/730/730OS = 4.9 inchesN/A. Specification Study.
    Conical Fittings with 6% TaperClauses 4.1-4.1 of ISO 540-2:1998Certificate of Conformance from the Supplier
    Sterile Hypodermic Syringes of Single UseISO 7886-1:1993Certificate of Conformance from the Supplier

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample sizes used for any of the biocompatibility or performance tests. The data provenance is not explicitly mentioned, but these are all non-clinical, bench-top tests conducted to verify product specifications and safety, likely performed internally by the manufacturer (Teleflex Medical). No human or animal data is described for these tests; they are related to material properties and device functionality.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable as the described tests are non-clinical and do not involve expert interpretation or human ground truth. They are objective measurements against defined chemical, physical, and engineering criteria.

    4. Adjudication method for the test set

    This section is not applicable as the described tests are objective measurements against defined criteria, not subjective assessments requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC comparative effectiveness study, nor is there any AI component to this device. This device is a manual, single-use medical device for topical anesthetic application.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This section is not applicable as the device is not an algorithm or AI-based system. It is a physical medical device.

    7. The type of ground truth used

    The "ground truth" for the biocompatibility tests is established by recognized international standards (ISO 10993-1) and specific grading scales (e.g., Kligman scoring system). For performance tests, the "ground truth" is the predefined quantitative and qualitative acceptance criteria based on engineering specifications and functional requirements. For specification verification, it's conformance to established international standards (e.g., ISO 540-2, ISO 7886-1) or internal product specifications.

    8. The sample size for the training set

    This section is not applicable as there is no training set mentioned for this type of device.

    9. How the ground truth for the training set was established

    This section is not applicable as there is no training set described.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1