510(k) Data Aggregation

    K Number
    K050068
    Manufacturer
    Date Cleared
    2005-02-09

    (28 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    LNOPV AD-L AND PD-L OXIMETRY SENSORS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LNOPv Ad-L and Pd-L Oximetry Sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) (measured by an SpO2 sensor) for use with Masimo SET pulse oximeters for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.

    Device Description

    The LNOPv Ad L and Pd L Oximetry Sensors are fully compatible disposable sensors for use with Masimo SET pulse oximeter monitors. They represent a design change to the Masimo LNOPv Oximetry Sensors. The LNOPv Ad L and Pd L disposable sensors are similar in construction to the predicate devices LNOPv In and LNOPv Ne except that the LNOPv Ad L and Pd L have a shorter tail and the emitter and detector position is switched. The LNOPv Ad-Land Pd-L use the same emitters (with Red wavelength of 658 nm and Infrared wavelength of 905 nm) as used in Masimo's LNOPv In and Ne sensors. The patient contacting materials in the LNOPv Ad-L and Pd-L disposable sensors are the same that is used in Masimo's LNOPv In and Ne sensors. The LNOPv Ad-L and Pd-L disposable sensors are supplied non sterile for single patient use. The LNOPv Ad L and Pd-L disposable sensors have the same electrical, optical, and material components as the LNOPv In and Ne disposable sensors.

    AI/ML Overview

    The provided text describes the performance testing of the LNOPv Ad-L and Pd-L Oximetry Sensors. Here's a breakdown of the acceptance criteria and study details based on the input:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implicit)Reported Device Performance
    Material Biocompatibility:
    Non-toxicDemonstrated non-toxic
    Non-irritatingDemonstrated non-irritating
    Non-sensitizingDemonstrated non-sensitizing
    Environmental Testing:
    ElectricalAll tests passed
    MechanicalAll tests passed
    EnvironmentalAll tests passed
    Clinical Accuracy (SpO2):