Search Results
Found 4 results
510(k) Data Aggregation
(67 days)
LIFEPAK 12 DEFIBRILLATOR/MONITOR
Manual Defibrillation:
Indications
Defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Energy delivered in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia.
Contraindications
Defibrillation is contraindicated in the treatment of Pulseless Electrical Activity (PEA), such as indioventricular or ventricular escape rhythms, and in the treatment of asystole.
Automated External Defibrillation:
Indications
AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing normally before using the defibrillator to analyze the patient's ECG rhythm. In AED mode, the LIFEPAK 12 defibrillator/monitor is not intended for use on pediatric patients less than 8 years old.
Noninvasive Pacing:
Indications
Noninvasive pacing is indicated for symptomatic bradycardia in patients with a pulse.
Contraindications: Noninvasive pacing is contraindicated for the treatment of ventricular fibrillation and asystole.
12-lead Electrocardiography:
Indications
The 12-lead electrocardiogram is used to identify, diagnose and treat patients with cardiac disorders and is useful in the early detection and prompt treatment of patients with acute myocardial infarction.
Pulse Oximetry:
Indications
Pulse Oximetry is indicated for use in any patient who is at risk of developing hypoxemia.
Noninvasive Blood Pressure Monitoring:
Indications
Noninvasive blood pressure monitoring is indicated for detection in trends of hypertension or hypotension. These include patient conditions indicated by abnormalities in various physiologic parameters such as shock, evaluation of perfusion during dysrhythmias, major fluid shifts, evaluation of response to fluid therapy, and titration of vasoactive and cardiotonic drugs. Noninvasive blood pressure monitoring may be useful during ECG monitoring or for post-defibrillation recovery analysis.
End-Tidal CO2 monitoring:
Indications
EtCO2 monitoring is indicated for detection of trends in the level of expired CO2. It is used for monitoring breathing efficacy and treatment effectiveness in acute cardiopulmonary care, for example, to determine if adequate compressions are being performed during CPR or to rapidly detect whether an endotracheal tube has been placed successfully. It is intended for use on adult and pediatric patients.
Invasive Pressure Monitoring:
Indications
The LIFEPAK 12 invasive pressure monitor is indicated for use in measuring arterial, venous, intracranial and other physiological pressures using and invasive catheter system with a compatible transducer. It may be used on the adult or pediatric patient.
The LIFEPAK 12 defibrillator/monitor was designed for use in a variety of hospital and pre-hospital settings including emergency departments, critical care areas, and air and ground ambulances. The LIFEPAK 12 defibrillator/monitor is powered by either a battery or AC power. LIFEPAK 12 defibrillator/monitor features include manual and automated external defibrillation, noninvasive pacing, ECG monitoring (3-lead, 7-lead and interpretive 12-Lead), pulse oximetry, synchronized cardioversion. noninvasive blood pressure monitoring, end-tidal CO2 monitoring, and invasive pressure monitoring.
The LIFEPAK 12 defibrillator/monitor has been modified with new setup options in automated external defibrillation (AED) mode to give medical directors or physicians flexibility in establishing their AED protocols including consistency with the recently updated 2005 AHA Guidelines for CPR and ECC.
The provided text describes a 510(k) summary for the Medtronic LIFEPAK® 12 Defibrillator/Monitor, which was modified with new setup options in automated external defibrillation (AED) mode. However, the document does not contain any study data, acceptance criteria for performance, or details about how the device's performance was evaluated against specific metrics.
It primarily focuses on:
- Device Description and Intended Use: Explaining what the LIFEPAK 12 is, its features, and for what purposes it's used.
- Indications for Use and Contraindications: Detailing when each function (Manual Defibrillation, AED, Pacing, ECG, Oximetry, NBP, EtCO2, Invasive Pressure) should be applied and when it should not.
- Substantial Equivalence: Stating that the modified device is substantially equivalent to a previously cleared version, with the only change being new setup options for AED mode.
- Regulatory Information: Classification, contact person, and the FDA's clearance letter.
Therefore, I cannot provide the requested information regarding acceptance criteria and study details because it is not present in the provided text. The document indicates a "Summary of Design Controls" and a "declaration of conformity to design controls," implying that internal testing and validation were performed to support substantial equivalence, but the specifics of these tests (e.g., performance metrics, sample sizes, ground truth establishment) are not included in this 510(k) summary.
Ask a specific question about this device
(29 days)
LIFEPAK 12 DEFIBRILLATOR/MONITOR
AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing spontaneously before using the defibrillator to analyze the patient's ECG rhythm. In AED mode, the LIFEPAK 12 defibrillator/monitor is not intended for use on pediatric patients less than 8 years old.
Manual Defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Energy delivered in the synchronized mode is a method for treating atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia.
Noninvasive Pacing is indicated for patients with symptomatic bradycardia or asystole.
12-Lead Electrocardiography is useful in the early detection and prompt treatment of patients with acute myocardial infarction.
Pulse Oximetry is used to check the saturation of oxygen in arterial blood. It is indicated for use in any patient who is at risk of developing hypoxemia.
Noninvasive Blood Pressure Monitoring is indicated for detection in trends of hypertension or hypotension. These include patient conditions indicated by abnormalities in various physiologic parameters such as shock, evaluation of perfusion during dysrhythmias, major fluid shifts, evaluation of response to fluid therapy, and titration of vasoactive and cardiotonic drugs. Noninvasive blood pressure monitoring may be useful during ECG monitoring or for post-defibrillation recovery analysis.
End-Tidal CO2 monitoring is indicated for detection of trends in the level of expired CO2. It is used for monitoring breathing efficacy in acute cardiopulmonary care, for example, to determine if adequate compressions are being performed during CPR or to rapidly detect whether an endotracheal tube has been placed successfully.
Invasive Pressure Monitoring is indicated for use in measuring arterial, venous, intracranial and other physiological pressures using an invasive catheter system with a compatible transducer.
The LIFEPAK 12 defibrillator/monitor is a complete acute cardiac response system, which consists of a battery or auxiliary powered defibrillator (manual or automated), external pacemaker, ECG monitor (3-lead, 5-lead and interpretive 12-Lead), pulse oximeter, noninvasive blood pressure monitor, end-tidal CO2 monitor, and invasive pressure monitor. Patient data can be transmitted from the LIFEPAK 12 defibrillator to a fax machine, computer, or to a receiving station. This modification changes the pulse oximetry module and sensors used in the LIFEPAK 12 defibrillator/monitor from Nellcor to Masimo.
The provided text does not contain detailed information about specific acceptance criteria and a study that proves the device meets those criteria for the pulse oximetry feature. The document is a 510(k) premarket notification for a modification to the LIFEPAK 12 Defibrillator/Monitor, specifically changing the pulse oximetry module and sensors from Nellcor to Masimo.
The submission claims substantial equivalence to previously cleared devices and mentions design control activities, but it does not include granular data on the performance of the new Masimo pulse oximetry module against predefined acceptance criteria from a study.
Therefore, most of the requested information cannot be extracted from the provided text.
Here's an attempt to answer based only on the provided information, noting what is missing:
1. Table of Acceptance Criteria and Reported Device Performance:
Feature/Metric | Acceptance Criteria | Reported Device Performance |
---|---|---|
Pulse Oximetry Accuracy | Not explicitly stated in the provided text. | Not explicitly stated in the provided text. The document indicates the modification involves changing the pulse oximetry module and sensors from Nellcor to Masimo, and states "The modified LIFEPAK 12 uses the same sensors and an updated version of the Masimo SET pulse oximetry module used in the currently marketed LIFEPAK 20 defibrillator/monitor." This suggests performance parity with the existing Masimo SET technology in the LIFEPAK 20, but specific accuracy metrics are not provided. |
2. Sample size used for the test set and the data provenance:
- Sample Size: Not specified in the provided text.
- Data Provenance (e.g., country of origin, retrospective/prospective): Not specified in the provided text.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not specified, as no study details (like ground truth establishment) are provided for the pulse oximetry performance.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not specified, as no study details are provided.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- A MRMC study is not mentioned. This device is a defibrillator/monitor with a pulse oximetry feature, not an AI diagnostic tool that human readers would use in conjunction with.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- A standalone performance study for the pulse oximetry accuracy is implied by the need for regulatory clearance for a new module, but the details (methodology, results, and whether it's "algorithm only") are not provided in this specific document. The document focuses on substantial equivalence via design control summaries and comparison to predicate devices, rather than a detailed performance study report.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not specified, as no study details for pulse oximetry accuracy are provided. For pulse oximetry, the typical ground truth would involve co-oximetry measurements from arterial blood gas analysis.
8. The sample size for the training set:
- Not applicable, as this is hardware/firmware modification of a medical device, not a machine learning algorithm that requires a "training set" in the conventional sense.
9. How the ground truth for the training set was established:
- Not applicable for the same reason as above.
Summary of what the document does provide regarding "proof":
The document primarily states that "This 510(k) includes a summary of design control activities and a declaration of conformity to design controls." and that "The information in this 510(k) notification demonstrates that the modified LIFEPAK 12 defibrillator/monitor is substantially equivalent to the predicate device."
The basis for the pulse oximetry change is that "The modified LIFEPAK 12 uses the same sensors and an updated version of the Masimo SET pulse oximetry module used in the currently marketed LIFEPAK 20 defibrillator/monitor." This implies that the performance of the new Masimo module is already established and cleared through the predicate LIFEPAK 20, and that the "proof" for the LIFEPAK 12 modification relies on this existing equivalence and design control adherence, rather than new, detailed performance study data presented within this 510(k) summary.
Ask a specific question about this device
(149 days)
LIFEPAK 12 DEFIBRILLATOR/MONITOR SYSTEM
Ask a specific question about this device
(210 days)
NIBP AND CO2 OPTIONS FOR THE PHYSIO-CONTROL LIFEPAK 12 DEFIBRILLATOR/MONITOR SYSTEM
NIBP
The LIFEPAK®12 NIBP monitor non-invasively measures blood pressure of the adult and pediatric patients by professionally trained health care providers. It is not designed for continuous, unsupervised monitoring.
EtCO2
The use of the LIFEPAK®12 EtCo2 monitor is indicated whenever professionally trained health care providers determine that a patient requires the continuous, non invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate.
The LIFEPAK®12 defibrillator / monitor series is a complete acute cardiac response system - battery or auxiliary powered defibrillator (manual and AED), monitor, pacemaker, 3 lead ECG, interpretive 12 lead ECG and pulse oximeter. Data can be transmitted by landline or cell phone to computer, fax, printer, or ECG storage system.
The users will be Advanced Life Support and Basic Life Support providers in a variety of hospital and pre-hospital settings. Emergency Medical Services users will include Paramedics and Emergency Medical Technicians trained and authorized to respond to medical emergencies. This device will be used in the pre-hospital setting and in the hospital, in critical areas (emergency departments, critical care, operating rooms, etc.) and on general duty floors (e.g. medical/surgical). It will also be used for patient transport (air and ground ambulance, in hospital transport, etc.)
Here's an analysis of the provided text regarding the acceptance criteria and study data for the NIBP and CO2 Options for the Physio-Control LIFEPAK®12 Defibrillator / Monitor System:
Based on the provided 510(k) summary, specific, quantitative acceptance criteria and the detailed study that proves the device meets those criteria are not explicitly described in the provided text. The document focuses primarily on establishing substantial equivalence to predicate devices and adherence to industry standards, rather than presenting a standalone performance study with detailed acceptance criteria and results.
However, we can infer some information from the text:
1. Table of Acceptance Criteria and Reported Device Performance
As noted, explicit quantitative acceptance criteria are not provided. The document states that the device is subject to "extensive safety and performance testing" and that "Final testing for the system includes various performance tests designed to ensure that the device meets all of its functional requirements and performance specifications."
The closest we get to "acceptance criteria" are the standards to which the device complies, implying that meeting the requirements of these standards constitutes acceptable performance.
Acceptance Criteria (Implied) | Reported Device Performance (Implied) |
---|---|
Compliance with ANSI/AAMI SP-10-1992 | The NIBP option is developed to comply with this standard. |
Compliance with IEC 601-2-30; 1995 | The NIBP option is developed to comply with this standard. |
Compliance with EN 864; 1997 | The CO2 option is developed to comply with this standard. |
Compliance with IEC 601-1 | The overall device is developed to comply with this general safety standard. |
Meets its functional requirements and performance specifications | "Final testing for the system includes various performance tests designed to ensure that the device meets all of its functional requirements and performance specifications." (No specific results provided) |
Substantial Equivalence to Predicate Devices | Stated to be substantially equivalent to listed predicate NIBP and CO2 monitors. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not specified. The document mentions "performance tests" and "safety testing" but does not provide any details about the number of subjects or cases used in these tests.
- Data Provenance: Not specified. There is no information provided about the country of origin of any testing data, nor if the data was retrospective or prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Number of Experts: Not specified.
- Qualifications of Experts: Not specified.
4. Adjudication Method for the Test Set
- Adjudication Method: Not specified. There is no mention of any expert consensus or adjudication process for establishing ground truth during testing.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size
- MRMC Study: No. The provided text does not mention any multi-reader multi-case comparative effectiveness study or any evaluation of human reader improvement with or without AI assistance. This document describes a modification to a medical device (defibrillator/monitor) with NIBP and CO2 options, which are sensors/measurement tools, not AI-driven diagnostic interpretation tools.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Standalone Performance Study: The document focuses on the device's performance in meeting standards and functional specifications. While "performance tests" are mentioned for the system, it's not described as an "algorithm only" study in the context of typical AI/software device standalone performance. The NIBP and CO2 functions are direct measurements, not an algorithm providing a diagnosis or interpretation that would typically have a "standalone" or "human-in-the-loop" comparison.
7. The Type of Ground Truth Used
- Type of Ground Truth: This is not explicitly stated as pathology, outcomes data, or expert consensus in the traditional sense. For NIBP and CO2 measurement, the "ground truth" would typically refer to accepted, gold-standard reference measurement methods. The document states compliance with standards like ANSI/AAMI SP-10-1992 and EN 864, which themselves define reference methods and accuracy requirements for such measurements. Therefore, the ground truth would be implicitly derived from these established reference measurement techniques as per the specified standards.
8. The Sample Size for the Training Set
- Sample Size for Training Set: Not applicable/Not specified. This document describes a hardware device with integrated measurement capabilities (NIBP and CO2). There is no mention of an AI algorithm that would require a "training set" in the context of machine learning.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: Not applicable/Not specified. As there's no mention of a machine learning component requiring a training set, this information is not relevant to the provided text.
In summary: The 510(k) summary provided focuses on establishing substantial equivalence to predicate devices and adherence to recognized performance and safety standards. It does not contain the detailed, quantitative efficacy study results, sample sizes, expert qualifications, or ground truth methodologies that would typically be found in submissions for novel diagnostic algorithms or AI-driven systems. The "studies" referred to are generally compliance tests against established engineering and medical device standards.
Ask a specific question about this device
Page 1 of 1