Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K014280
    Date Cleared
    2002-03-20

    (83 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EXAMINATION GLOVE (POWDERED) IS A DISPOSABLE DEVICE MADE OF NATURAL LATEX INTENDED FOR MEDICAL PURPOSE THAT IS WORN ON THE EXAMINERS HAND OR FINGER TO PREVENT CONTAMINATION BETWEEN PATIENT AND EXAMINER AND IS POWDERED WITH A DONNING POWDER ABSORBABLE USP CORN STARCH.

    Device Description

    Class I Patient Examination Gloves (powdered) 80L that meets all 'the requirements of ASTM D3578.

    AI/ML Overview

    Here's the breakdown of the acceptance criteria and study information for the Latex Examination Gloves (Powdered) from Brightway Gloves Pvt. Ltd., as extracted from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsSIZEAcceptance Criteria (ASTM D3578)Reported Performance (BGPL Value)Meets Criteria?
    LengthEX-S220 mm minimum235-240 mmYes
    S220 mm minimum235-240 mmYes
    M230 mm minimum235-240 mmYes
    L230 mm minimum235-240 mmYes
    WidthEX S70 +/- 6 mm70 mmYes
    S80 +/- 6 mm82 mmYes
    M95 +/- 6 mm93 mmYes
    L111 +/- 6 mm107 mmYes
    ThicknessEX S0.08 mm minimum0.10 mmYes
    S0.08 mm minimum0.10 mmYes
    M0.08 mm minimum0.10 mmYes
    L0.08 mm minimum0.10 mmYes
    Tensile Strength (Before Ageing)N/A14 mpa min20 - 22 mpaYes
    Elongation at break % (Before Ageing)N/A700% min800 - 850%Yes
    Tensile Strength (After Ageing)N/A14 mpa min18 – 20 mpaYes
    Elongation at break % (After Ageing)N/A500% min750-800%Yes
    Freedom from Holes (AQL)N/A41.5 (Reported as 'followed by BGPL', lower AQL is better)Yes
    Dimension (AQL)N/A44Yes
    Physical Property (AQL)N/A44Yes
    Powder ContentN/ANot explicitly stated in table, but noted in "Conclusion"120 +/- 20 mg per gloveYes (implied by conclusion)
    Protein ContentN/ANot explicitly stated in table, but noted in "Conclusion"80 +/- 20 ppmYes (implied by conclusion)
    Moisture ContentN/ANot explicitly stated in table, but noted in "Conclusion"0.8% maxYes (implied by conclusion)
    BiocompatibilityN/ANot explicitly stated in table, but noted in "Conclusion"Biologically CompatibleYes (implied by conclusion)
    Pin hole RequirementN/AFDA Pin hole RequirementNot explicitly quantified, but stated "Meet FDA Pin hole Requirement"Yes

    Note: For characteristics like Powder Content, Protein Content, Moisture Content, and Biocompatibility, no specific ASTM D3578 numerical requirement was explicitly listed in the tables provided for direct comparison. However, the "Conclusion of Performance Test Data" explicitly states that the gloves "Meet or exceed the ASTM D3578" and "Meet FDA Pin hole Requirement" and "Meet labelling claim," implying that the stated performance values for these attributes align with the relevant standards.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample size used for the test set for each parameter. It lists the measured values for various glove sizes and properties. The data provenance is from India (manufacturer is M/s. BRIGHTWAY GLOVES PVT.LTD. in Tamil Nadu, India). The study appears to be retrospective in the sense that it presents the results of tests already conducted (e.g., "The performance test data... given below").

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    N/A. This is a medical device (patient examination gloves) that is evaluated against established technical standards (ASTM D3578) and regulatory requirements (FDA pinhole requirement). The ground truth is objective measurement data against these standards, not expert interpretation of subjective information.

    4. Adjudication Method for the Test Set

    N/A. As mentioned above, the determination is based on objective measurements against predefined standards, not subjective assessments requiring adjudication among experts.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. This is a physical medical device (gloves), not an AI-powered diagnostic or interpretive system that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    N/A. This is a physical medical device, not an algorithm.

    7. The type of ground truth used

    The ground truth used for this device is based on established industry standards and regulatory requirements, specifically:

    • ASTM D3578 (Standard Specification for Rubber Examination Gloves)
    • FDA Pinhole Requirement

    These provide objective benchmarks for physical, chemical, and performance properties.

    8. The sample size for the training set

    N/A. This is a physical medical device. The concept of a "training set" is relevant for AI algorithms, not for the manufacturing and testing of examination gloves. The product undergoes quality control and testing against specifications, not machine learning model training.

    9. How the ground truth for the training set was established

    N/A. As above, the concept of a "training set" and its ground truth establishment is not applicable to this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1