Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K230623
    Manufacturer
    Date Cleared
    2023-09-01

    (179 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KITE Distal Fibula Kit is indicated for:

    • Diaphyseal and metaphyseal fractures of the distal fibula
    • Intra and extra articular fractures of the distal fibula
    • Non-unions of the distal fibula
    Device Description

    The KITE Distal Fibula Kit is intended for the stabilization of distal lateral and posterior fibular fractures including those of the diaphysis and metaphysis and intra-articular fractures. The system is comprised of various plates and screws which allow it to be customized to a patient's anatomy and fracture pattern. The plate is secured in place via autolocking and cortical screws.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the KITE Distal Fibula Kit, formatted as requested:

    KITE Distal Fibula Kit Acceptance Criteria and Study Analysis

    Below is a summary of the acceptance criteria and the studies performed to demonstrate the device meets these criteria, based solely on the provided FDA 510(k) summary.


    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance Criteria SourceReported Device Performance
    Plates
    Static 4-point BendingASTM F382-17 Annex A1 and "Orthopedic Fracture Fixation Plates - Performance Criteria for Safety and Performance Based Pathway"Device performance demonstrates alignment with acceptance criteria. (Specific quantitative results not provided in the summary)
    Screws
    Torsional TestASTM F543-17 and "Orthopedic Non-Spinal Metallic Bone Screws and Washers - Performance Criteria for Safety and Performance Based Pathway"Device performance demonstrates alignment with acceptance criteria. (Specific quantitative results not provided in the summary)
    Insertion & Removal TestASTM F543-17 and "Orthopedic Non-Spinal Metallic Bone Screws and Washers - Performance Criteria for Safety and Performance Based Pathway"Device performance demonstrates alignment with acceptance criteria. (Specific quantitative results not provided in the summary)
    Pullout TestASTM F543-17 and "Orthopedic Non-Spinal Metallic Bone Screws and Washers - Performance Criteria for Safety and Performance Based Pathway"Device performance demonstrates alignment with acceptance criteria. (Specific quantitative results not provided in the summary)
    Self-Tapping TestASTM F543-17 and "Orthopedic Non-Spinal Metallic Bone Screws and Washers - Performance Criteria for Safety and Performance Based Pathway"Device performance demonstrates alignment with acceptance criteria. (Specific quantitative results not provided in the summary)
    Sterility/Biocompatibility
    LAL TestingFDA Guidance document "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile Guidance for Industry and Food and Drug Administration Staff" issued January 21, 2016Testing was performed according to the guidance. (Specific results like endotoxin levels or pass/fail not explicitly stated but implied success for clearance)

    Notes on Reported Performance: The summary states that performance data "demonstrates alignment" or "demonstrate substantial equivalence" and "demonstrate alignment acceptance criteria," indicating that the device met the specified criteria. However, explicit quantitative results for each test (e.g., specific bending strength values, torque values, pullout forces) are not provided in this 510(k) summary.


    2. Sample Size Used for the Test Set and Data Provenance

    The provided text describes bench testing for mechanical properties and sterility, not clinical testing with human subjects or patient data. Therefore, the concepts of "test set" in terms of patient data, data provenance, and retrospective/prospective studies are not applicable here.

    • Sample Size for Mechanical Testing: The specific number of plates or screws tested for each mechanical test (e.g., how many plates for 4-point bending, how many screws for torsional tests) is not specified in the provided summary.
    • Data Provenance: Not applicable as it's mechanical bench testing, not clinical data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The ground truth for mechanical and sterility testing is established by the specified ASTM standards and FDA guidance documents, which define the testing methodology and acceptance criteria. This does not involve human expert adjudication of results in the way clinical image analysis or diagnostic studies would.


    4. Adjudication Method for the Test Set

    Not applicable. This is not a study involving human readers or interpretation of results that would require an adjudication method like 2+1 or 3+1 consensus. The tests are either pass/fail based on objective mechanical properties or sterility criteria defined in standards and guidance documents.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices (e.g., AI for medical imaging) to assess the impact of AI assistance on human reader performance. The KITE Distal Fibula Kit is a bone fixation appliance, and its clearance is based on mechanical performance and biocompatibility.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a bone fixation kit, not an algorithm or AI-based diagnostic tool. The performance studies described are for the physical components of the kit.


    7. The Type of Ground Truth Used

    The ground truth used for this submission is based on:

    • Mechanical Standards: ASTM F382-17 (for plates) and ASTM F543-17 (for screws). These standards define the methods for measuring specific mechanical properties.
    • FDA Guidance Documents: "Orthopedic Fracture Fixation Plates - Performance Criteria for Safety and Performance Based Pathway" and "Orthopedic Non-Spinal Metallic Bone Screws and Washers - Performance Criteria for Safety and Performance Based Pathway" for mechanical performance, and "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile Guidance for Industry and Food and Drug Administration Staff" for LAL testing. These documents define the acceptance criteria that demonstrate safety and performance for these types of devices.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that requires a training set. The device is a physical medical implant.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for an AI/machine learning algorithm, there is no ground truth to establish for it.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1