Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K073074
    Manufacturer
    Date Cleared
    2008-07-10

    (253 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    KEY LASER III 1243 US WITH DETECT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For removal of subgingival calculi in periodontal pockets with periodontitis by closed or open curettage. Removal of subgingival calculi in periodontal pockets which have been affected by periodontitis, with open or closed curettage and using the detection function of the KEY Laser 1243.

    Device Description

    This device represents the combination into one device of two previously cleared devices, the Diagnodent Laser Fluorescence Caries Detection Device with the KEY Laser III 1243.

    AI/ML Overview

    Acceptance Criteria and Study for KEY Laser III 1243 US without Feedback, with Detect

    Based on the provided 510(k) Summary (K073074), the KEY Laser III 1243 US without Feedback, with Detect is a combination of two previously cleared devices: the Diagnodent Laser Fluorescence Caries Detection Device and the KEY Laser III 1243. The acceptance criteria and "study" are focused on demonstrating that the performance of the individual components has not changed as a result of their combination and that the combined device meets established safety standards for laser and medical devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred)Reported Device Performance
    Functional PerformancePerformance of detection function is unchanged from predicate."Laboratory testing has revealed that the performance of the detection function... have not changed."
    Performance of laser surgical functions is unchanged from predicate."Laboratory testing has revealed that the... performance of the laser surgical functions have not changed."
    Safety - ElectricalCompliance with IEC 601-1/VDE 0750 Part 1Device has been tested in accordance with IEC 601-1/VDE 0750 Part 1 and UL Std No 60601-1 (1st Edition).
    Compliance with UL Std No 60601-1 (1st Edition)Device has been tested in accordance with IEC 601-1/VDE 0750 Part 1 and UL Std No 60601-1 (1st Edition).
    Safety - LaserCompliance with IEC 825-1 / DIN-VDE 0837 Part 1Device has been tested in accordance with IEC 825-1 / DIN-VDE 0837 Part 1, Regulations 21 CFR 1040, ANSI Z 136.1, and ANSI Z 136.3.
    Compliance with Regulations 21 CFR 1040Device has been tested in accordance with IEC 825-1 / DIN-VDE 0837 Part 1, Regulations 21 CFR 1040, ANSI Z 136.1, and ANSI Z 136.3.
    Compliance with ANSI Z 136.1Device has been tested in accordance with IEC 825-1 / DIN-VDE 0837 Part 1, Regulations 21 CFR 1040, ANSI Z 136.1, and ANSI Z 136.3.
    Compliance with ANSI Z 136.3Device has been tested in accordance with IEC 825-1 / DIN-VDE 0837 Part 1, Regulations 21 CFR 1040, ANSI Z 136.1, and ANSI Z 136.3.
    Safety - EMCDemonstrated Electromagnetic CompatibilityDevice has been tested in accordance with relevant EMC standards (implied by "EMC and electrical safety testing").
    BiocompatibilityCompliance with ISO 7494Device has been tested in accordance with ISO 7494.
    Overall EquivalenceIndividual device performance and indications for use have not changed; device is substantially equivalent to predicates."The indications for use and the applied standards have not changed, nor has the performance of the individual devices changed as a result of combining two devices into one package."
    "In all important respects, the KEY Laser III 1243 US without Feedback, with Detect is substantially equivalent to the individual predicates."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide specific details on the sample size used for any in-vitro or bench testing. It only states "Laboratory testing has revealed..." and "bench, and in-vitro studies." The data provenance (e.g., country of origin, retrospective/prospective) is also not specified, as this was a premarket notification for a combination device, and detailed clinical study results are typically not required in the same way as for novel devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided. The "ground truth" for this submission appears to be derived from the established performance and safety profiles of the individual predicate devices and the relevant industry standards. There is no mention of human expert evaluation in establishing ground truth for the "laboratory testing" or "bench/in-vitro studies."

    4. Adjudication Method for the Test Set

    Not applicable. There is no mention of a test set requiring adjudication in the context of human expert review. The "studies" were technical laboratory and bench tests against pre-defined performance and safety standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-assisted diagnostic tool for "human readers" in the typical sense that an MRMC study would be performed. It's a medical device with a laser for surgical procedures and a fluorescence detection function. The comparison is to predicate devices, not to human performance with or without AI.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This statement is partially applicable in that the device's functional integrity (laser operation, detection sensitivity) was assessed in isolation. The "laboratory testing" and "bench, and in-vitro studies" essentially evaluated the technical performance of the device components without necessarily considering a "human-in-the-loop" performance study for this type of submission. However, it's not an "algorithm only" device; it's a physical medical instrument. The "detection function" performance was evaluated in what could be considered a standalone manner from the surgical function, but both are physical components of the device.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for this submission appears to be:

    • Validated performance parameters of the predicate devices: The functionality of the detection and laser surgical components are compared to their previously cleared, individual versions.
    • Established industry and regulatory standards: Compliance with various IEC, ISO, UL, VDE, ANSI, and 21 CFR standards for electrical safety, laser safety, and biocompatibility serves as the ground truth for safety and technical performance.

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI algorithm in the context of requiring a training set for model development. The device combines existing, cleared technologies.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as this device does not utilize a training set in the AI machine learning sense.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1