Search Results
Found 1 results
510(k) Data Aggregation
(56 days)
JUSHA-C43 LCD Monitor is intended to be used in displaying digital images for review and analysis by trained medical practitioners. It does not support the display of mammography images for diagnosis.
JUSHA-C43 LCD Monitoris the display system with the high resolution(2560 x 1600), high luminance(350cd/m²) and 1024 simultaneous shades of gray out of a palette of 4096, 4 DICOM look up table inside, the product is consisted of the following components:
- 30inch, Color Active Matrix Liquid Crystal Display
- FR4-v0.4
- JUSHA-C43 LCD Monitor software
- Power Adapter
- Data Cable.
The LCD Monitoris designed, tested, and will be manufactured in accordance with both mandatory and voluntary standards:
- IEC 60601-1Medical equipment medical electrical equipment - Part 1: General requirements for basic safety and essential performance 2005 + CORR. 1 (2006) +CORR. 2 (2007)
- IEC 60601-1-2 Edition 3:2007, medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
The provided document describes the JUSHA-C43 LCD Monitor, a display system for medical imaging review and analysis. The submission focuses on demonstrating substantial equivalence to a predicate device (RADIFORCE RX440; K130070) through bench testing for display performance and adherence to electrical safety and electromagnetic compatibility (EMC) standards.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of acceptance criteria with corresponding pass/fail results or reported device performance against specific numerical targets. Instead, it lists the types of measurements performed in bench testing and states that the device "meets all performance standards" without providing the specific standards or numerical outcomes.
However, based on the Discussion of Differences in the comparison table, we can infer some implied performance expectations and how the JUSHA-C43 LCD Monitor performs relative to its predicate.
Acceptance Criteria (Inferred from Predicate Comparison & Bench Testing) | Reported Device Performance (JUSHA-C43 LCD Monitor) |
---|---|
Display Performance | |
Screen Technology Performance (Resolution, Luminance, Viewing Angle) | Resolution: 2560 x 1600 / 1280 x 1600 x 2 (Same as predicate) |
Recommended Luminance: 300cd/m² (Same as predicate) | |
Viewing Angle: Horizontal 178°, Vertical 178° (Wider than predicate) | |
Contrast Ratio: 1200:1 (Higher than predicate, "can display image more clarity") | |
Display Area: 641.28 (H) x 400.8 (V) mm (Same as predicate) | |
Pixel Pitch: 0.2505 x 0.2505 mm (Same as predicate) | |
Angular Dependency of Luminance Response | Measurement performed, device "meets all performance standards." (No specific results provided) |
Luminance Non-uniformity characteristics (TG18 guideline) | Measurement performed, device "meets all performance standards." (No specific results provided) |
Chromaticity Non-uniformity characteristics (TG18 guideline) | Measurement performed, device "meets all performance standards." (No specific results provided) |
Small-spot Contrast Ratio | Measurement performed, device "meets all performance standards." (No specific results provided) |
Temporal Response | Measurement performed, device "meets all performance standards." (No specific results provided) |
Luminance Stability | Measurement performed, device "meets all performance standards." (No specific results provided) |
Display Colors | 12-bit, 68.7 billion colors (Better than predicate, uses "color extension technology to improve image display quality") |
Scanning Frequency (H; V) | 103.8kHz; 50Hz (Different from predicate, but display area and effect are stated to be "same") |
Dot Clock | 148MHz (Different from predicate, but display area and effect are stated to be "same") |
Electrical Safety and EMC | |
IEC 60601-1 (Safety) | Complies with IEC 60601-1:2005 + CORR. 1 (2006) + CORR. 2 (2007) (Newer edition than predicate) |
IEC 60601-1-2 (EMC) | Complies with IEC 60601-1-2 Edition 3:2007 (Same edition as predicate) |
Other | |
Power Consumptions/Save Mode | 82W (less than predicate's 84W); less than 1.5W in save mode (different but not affecting display function) |
Indications for Use | Same as predicate: displaying digital images for review and analysis by trained medical practitioners; does not support mammography. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document describes bench testing only. There is no mention of a test set involving medical images or patient data, nor is there any information about sample size for such a test set, country of origin, or whether it was retrospective or prospective. The testing was performed on the device itself.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. Since there was no test set involving medical images or human interpretation, there were no experts used to establish ground truth for a test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. No test set involving human interpretation was performed.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an LCD monitor, not an AI-powered diagnostic tool. No MRMC study was conducted, and the concept of "human readers improve with AI vs without AI assistance" does not apply here.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a display monitor, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the device's performance relies on engineering specifications and established international standards. For example, luminance and chromaticity non-uniformity were measured against TG18 guidelines, and electrical safety and EMC were evaluated against IEC 60601-1 and IEC 60601-1-2 standards.
8. The sample size for the training set
Not applicable. This device is a hardware display monitor; it does not involve machine learning or a training set.
9. How the ground truth for the training set was established
Not applicable. No training set was used.
Ask a specific question about this device
Page 1 of 1