Search Results
Found 1 results
510(k) Data Aggregation
(739 days)
JTK Dental Implant System
JTK One-Piece Integrated Implant System:
The JTK one-piece integrated dental implant system is intended to replace single or multiple teeth in the fully or partially edentulous mandibular or maxillary alveolar process. The implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
JTK Two-Piece Implant System:
The JTK two-piece dental implant system is comprised of dental implant fixtures and prosthetic devices a two-piece implant system. The implants are intended for use in the mandla, in support of single unit or multiple unit cement or screw-receiving restorations and support of overdentures. The implants are intended for immediate placement and function for the support of single tooth restorations, recognizing bone stability and appropriate occlusal load requirements.
The JTK Dental Implant System is offered in two different models consist of a one-piece integrated implant system where the implant and abutment are machined from the same piece of titanium, and a traditional two-piece implant system that consists of an implant and abutment that are made from different pieces titanium and held together by way of a supplied titanium healing cap is also supplied with the two-piece implants and abutments are supplied individually packaged and sterile to the end user via gamma radiation. The healing cap is packaged with the implant and the attachment screw is packaged with the abutment. All implant System are made by traditional CNC manufacturing methods. For both models of the implant body, the threaded portion receives a surface modification using sandblasting and acid etching (SLA).
For the one-piece implant system, the implants are offered in two thread diameters (3.0 and 3.3 mm) and in three thread lengths (10, 13 and 15 mm). The integral abutment portion of the implant is contoured and machined, is without a surface modification, and is offered in two platform heights (4 and 6 mm).
For the two-piece implant system, the implants are offered in five thread diameters (3.8, 4.0, 4.5, 5.0 and 5.5 mm) and in four thread lengths (8, 10, 13 and 15 mm). The neck of the implant is conical in shape, 2.5 mm in height and has been machined and is without a surface modification. The two-piece implant bodies are considered a tissue level implant. The abutments that are used with the two-piece system are offered in two different models (straight and angled (15° and 25°). The straight abutments are offered in a diameter of 4.4 mm and in lengths of 4.2, 5.2, 6.2 and 8.2 mm. The 15° angled abutments are offered in a diameter of 4.4 mm and in lengths of 6.0 and 8.0 mm, while the 25° the angled abutments are offered in a diameter of 4.4 mm and in lengths of 6.0 and 10.5 mm.
The titanium used to manufacture the one-piece implants conform with ASTM F67, Standard Specification for Unalloyed Titanium, for Surgical Implant Applications while the titanium used to manufacture the abutments conform with ASTM F136, Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications.
The provided text is an FDA 510(k) Premarket Notification summary for the JTK Dental Implant System. It focuses on demonstrating substantial equivalence to predicate devices through comparisons of device characteristics and non-clinical performance data.
Crucially, this document describes a dental implant system, a physical medical device, not a software or AI/ML-based device. Therefore, the concepts of acceptance criteria for algorithm performance (like sensitivity, specificity, AUC), ground truth experts, adjudication methods, MRMC studies, or training/test sets in the context of an AI/ML algorithm are not applicable to this submission.
The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this document refer to the non-clinical performance testing required for medical devices to demonstrate safety and effectiveness and substantial equivalence to existing predicate devices.
Here's an interpretation based on the provided text, aligning with the type of device being discussed:
1. A table of acceptance criteria and the reported device performance:
The "acceptance criteria" here are generally international standards and FDA guidance documents that the device must conform to for safety and performance. The "reported device performance" indicates whether the device passed these tests.
Acceptance Criterion (Standard/Guidance) | Reported Device Performance |
---|---|
Biocompatibility Testing per ISO 10993-1 (cytotoxicity, rabbit pyrogen test) | Passed all testing |
Dynamic Fatigue Testing per ISO 14801 | Demonstrated 5x10^6 cycles without failure |
Sterilization Validation per ISO 11137-1, -2, -3 | Demonstrated a SAL of 10^-6 via gamma radiation |
Shelf-life Validation per ASTM F1980-07 | Stated shelf-life of 2 years |
Packaging Validation per ISO 11607 | Stated (implicit in shelf-life validation) |
Risk Analysis per ISO 14971 | Conforms to the ISO standard |
LAL testing for pyrogenicity (FDA Guidance) | Addressed all recommendations |
2. Sample size used for the test set and the data provenance:
For this type of physical device, "sample size" refers to the number of physical units tested. The document does not specify exact numbers for each test (e.g., how many implants were tested for fatigue). The provenance is "Jiangyin Jintech Biotech Co., Ltd." in "Jiangyin City, Jiangsu Province, China," as they conducted the performance testing. These are non-clinical (laboratory) studies.
- Sample Size: Not explicitly stated for each test (e.g., number of implants for fatigue testing), but implied to be sufficient to meet the requirements of the standards.
- Data Provenance: Non-clinical (laboratory) testing conducted by Jiangyin Jintech Biotech Co., Ltd. in China.
- Retrospective or Prospective: These are laboratory tests specifically conducted for the submission, so they are prospective in nature to generate data to support the 510(k).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This concept is not applicable here. Ground truth is typically established for diagnostic or AI/ML performance. For a dental implant, "ground truth" relates to its physical and biological performance as measured by objective tests (e.g., fatigue strength, biocompatibility, sterilization efficacy) against established international standards. The "experts" are the engineers, scientists, and technicians who design, conduct, and analyze these standardized tests.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This concept is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving human interpretation (e.g., radiologists interpreting images) to resolve disagreement. For physical device performance testing, the results are typically quantitative measurements against objective criteria, not subjective interpretations requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This concept is not applicable. An MRMC study is relevant for AI/ML-assisted diagnostic devices. This submission concerns a physical dental implant. There are no "human readers" or "AI assistance" in the context of this device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This concept is not applicable as this is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for this medical device is the adherence to established international consensus standards (e.g., ISO, ASTM) and FDA guidance for the physical, chemical, and biological properties of dental implants. For instance, the "ground truth" for fatigue is whether the implant can withstand 5x10^6 cycles without failure as per ISO 14801. The "ground truth" for biocompatibility is whether it passes cytotoxicity and pyrogen tests per ISO 10993. These are objective, quantifiable standards, not subjective interpretations.
8. The sample size for the training set:
This concept is not applicable. There is no "training set" for a physical dental implant system in the AI/ML sense. The "training" for such a device occurs during its design and manufacturing processes, where iterations and refinements are made to meet engineering specifications and regulatory requirements.
9. How the ground truth for the training set was established:
This concept is not applicable. Since there is no "training set" in the AI/ML context, there's no ground truth established for it. The "ground truth" for the device's design and manufacturing quality relies on engineering principles, material science, and adherence to quality management systems (e.g., 21 CFR Part 820 Quality System Regulation).
Ask a specific question about this device
Page 1 of 1