Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    ANTHEM Total Knee System, Genesis Uni Knee System, JOURNEY BCS and II Knee Systems, JOURNEY II CR Knee
    System, JOURNEY II Uni Knee System, Journey II UK Knee System, JOURNEY II XR Bi-Cruciate Retaining Knee

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. Rheumatoid arthritis.
    2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
    3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
    4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
      The ANTHEM PS Total Knee System is indicated for use only with cement and is a single use device.

    Unicompartmental knee implants are indicated for restoring either compartment of a knee that has been affected by the following:

    1. Noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;

    2. Correction of functional deformity;

    3. Revision procedures where other treatments or devices have failed: and

    4. Treatment of fractures that are unmanageable using other techniques
      Unicompartmental knee implants are single use only and are intended for implantation only with bone cement. HA coated unicompartmental knee implants are available outside the US for use without bone cement.

    5. Rheumatoid arthritis.

    6. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.

    7. Failed osteotomies, unicompartmental replacement, or total knee replacement.

    8. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
      Smith & Nephew, Inc. Journey II BCS Knee components are indicated for use only with cement and are single use devices.

    9. Rheumatoid arthritis.

    10. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.

    11. Failed osteotomies, unicompartmental replacement, or total knee replacement.

    12. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
      The Journey II Total Knee system components are indicated for use only with cement and are single use devices.

    Unicompartmental knee implants are indicated for restoring either compartment of a knee that has been affected by the following:

    1. Noninflammatory degenerative joint disease including osteoarthritis, or avascular necrosis;
    2. Correction of functional deformity;
    3. Revision procedures where other treatments or devices have failed; and
    4. Treatment of fractures that are unmanageable using other techniques
      Unicompartmental knee implants are single use only and are intended for implantation only with bone cement. HA coated unicompartmental knee implants are available outside the US for use without bone cement.

    Unicompartmental knee implants are indicated for restoring either compartment of a knee that has been affected by the following:
    · Non-inflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis:
    · Correction of functional deformity;
    · Revision procedures where other treatments or devices have failed; and
    · Treatment of fractures that are unmanageable using other techniques.
    Unicompartmental knee implants are single use only and are intended for implantation only with bone cement.

    Total knee components are indicated for rheumatic arthritis, osteoarthritis, osteoarthritis, or degenerative arthritis; failed osteotomies, unicompartmental replacement, or total knee replacement.
    The Journey II Total Knee system components are indicated for use only with cement and are single use devices.

    1. Degenerative arthritis in the distal femur and patella;
    2. A history of patellar dislocation or patellar fracture; and
    3. Failed previous surgery (arthroplasty, tibial tubercule elevation, lateral release) where pain, deformity or dysfunction persists.
      The Smith & Nephew Patello-Femoral Implants are intended for implantation with bone cement.

    Unicompartmental knee implants are indicated for restoring either compartment of a knee that has been affected by the following:

    1. Noninflammatory degenerative joint disease including osteoarthritis, or avascular necrosis;

    2. Correction of functional deformity;

    3. Revision procedures where other treatments or devices have failed; and

    4. Treatment of fractures that are unmanageable using other techniques
      Unicompartmental knee implants are single use only and are intended for implantation only with bone cement. HA coated unicompartmental knee implants are available outside the US for use without bone cement.

    5. Rheumatoid arthritis.

    6. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.

    7. Failed osteotomies, unicompartmental replacement, or total knee replacement.

    8. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.

    9. Constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral collateral ligament) are incompetent.

    10. Hinge knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral ligament) are absent or incompetent.
      The Legion Hinge Knee System is for Cemented Use Only.

    11. Rheumatoid arthritis.

    12. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.

    13. Failed osteotomies, unicompartmental replacement, or total knee replacement.

    14. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.

    15. Constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral collateral ligament) are incompetent.

    16. Hinge knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral ligament) are absent or incompetent.
      The LEGION Knee System - Finned Tibial Wedges are for single use only and are intended for implantation with bone cement.

    These devices are indicated for patients with:

    1. Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis.
    2. Previous tibial condyle or plateu fractures with loss of anatomy or function.
    3. Varus or Valgus deformities.
    4. Revision of previous arthroplasty procedures.
      The devices are indicated for cemented use only.
      The Zimmer Unicompartmental Knee System is designed for use when load bearing ROM is expected to be less than or equal to 155 degrees.

    The Engage™ Partial Knee System is intended for medial unicompartmental knee arthroplasty to treat one or more of the following conditions:

    • Moderately disabling joint disease of the knee resulting from painful osteo- or post traumatic arthritis.
    • Revision of previous unsuccessful surgical procedures, including prior unicompartmental knee arthroplasty
    • As an alternative to tibial osteotomy in patients with Unicompartmental osteoarthritis.
      The femoral component and tibial tray are intented fixation. The porous surfaces of both the femoral and tibial tray components provide biological fixation when used in a cementless application. When the tibial tray is implanted without the use of bone cement, the tibial anchor should be used. When the tibial tray is implanted with bone cement, the tibial anchor should not be used.

    Total Knee components are indicated for:

    1. Rheumatoid arthritis
    2. Post- traumatic arthritis, osteoarthritis, or degenerative arthritis
    3. Failed osteotomies, unicompartmental replacement, or total knee replacement
    4. The posterior stabilized knee system is designed for use in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
      LEGION Porous CR Narrow Femoral Components are indicated for use without bone cement and are single use devices.
      The Porous Patella and the Porous Tibia Baseplate are indicated for use with or without bone cement, and are single use devices.
    Device Description

    The purpose of this Traditional 510(k) is to add MR safety information to the labeling (instructions for use, product labeling and patient implant card) of the subject Smith & Nephew Knee Systems. The technological characteristics, function of the devices, packaging and sterilization remain unchanged. As these products have been in distribution for several years, some parts have undergone changes in the technological characteristics that were assessed in compliance with the FDA's guidanceon "Deciding When to Submit a 510(k) for a Change to an Existing Device" or the FDA predecessor guidance on how to review changes to an existing device. These are not cumulative changes for all Smith & Nephew Knee Systems, but are rather specific changes that only affect certain parts. In the context of this submission, no modifications have been made to the device design, materials, sterilization, or the manufacturing process of the previously cleared devices due to the addition of MR Labeling for this 510(k).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for various Smith & Nephew knee systems. The purpose of this submission is not to introduce a new device or significantly change an existing one that would require a new performance study to demonstrate clinical effectiveness or safety based on patient outcomes. Instead, the core of this submission is to add MR safety information to the labeling of existing, already cleared devices.

    Therefore, the typical "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of diagnostic or AI-driven devices (which you are asking about) are not applicable to this 510(k) submission.

    The "performance data" section in the document, despite its heading, refers specifically to Magnetic Resonance Imaging (MRI) compatibility testing, not clinical performance of the knee implants themselves. This testing aims to demonstrate that the presence of the knee implants does not pose an undue risk or create excessive artifacts during MRI scans.

    Here's why the questions you asked cannot be fully answered or are not relevant to this specific regulatory submission:

    • No new device performance for clinical outcomes: This submission isn't about proving a device (like an AI algorithm) is effective at diagnosing or treating a condition. It's about ensuring an existing physical implant is safe in an MRI environment.
    • No diagnostic or AI algorithm: The "device" in question here is a knee implant, not a software algorithm that generates clinical insights or diagnoses. Therefore, terms like "multi reader multi case (MRMC) comparative effectiveness study," "standalone (algorithm only) performance," "ground truth," or "training set" are not applicable.

    However, I can extract the relevant information regarding the MR safety testing which serves as the "study" for this particular 510(k) submission:


    Acceptance Criteria and Study for MR Safety Labeling (as described in the document):

    Since this 510(k) is specifically for adding MR safety information to the labeling of already cleared devices, the "acceptance criteria" and "study" revolve around demonstrating the device's compatibility with MRI environments, not its clinical performance as a knee implant.

    1. Table of Acceptance Criteria and Reported Device Performance (for MR Safety)

    Acceptance Criteria (Implied)Reported Device Performance (from document)
    Device does not pose an unacceptable magnetic hazard due to displacement force. (ASTM F2052)"MR Safety testing/assessments support the appropriate MR parameters and symbols found in the subject device labeling."
    Device does not pose an unacceptable magnetic hazard due to torque. (ASTM F2213)"MR Safety testing/assessments support the appropriate MR parameters and symbols found in the subject device labeling."
    Device does not cause unacceptable RF-induced heating. (ASTM F2182-19e2, IEC 60601-2-33, ISO/TS 10974:2018E)"MR Safety testing/assessments support the appropriate MR parameters and symbols found in the subject device labeling."
    Device does not create unacceptable MR image artifact. (ASTM F2119)"MR Safety testing/assessments support the appropriate MR parameters and symbols found in the subject device labeling."

    Summary of Device Performance: The document states that "MR Safety testing/assessments support the appropriate MR parameters and symbols found in the subject device labeling." This implies that the devices met the requirements of the referenced ASTM and IEC standards for MR compatibility, allowing defined MR conditions (e.g., specific field strengths, gradients, RF levels) for imaging patients with these implants.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document does not specify the sample size for the MR safety testing. These tests are typically conducted on representative samples of the implant devices themselves, not on human patients.
    • Data Provenance: The testing references industry standards (ASTM, IEC, ISO/TS), indicating that the tests were performed in a controlled laboratory environment in accordance with these established protocols. The country of origin of the data is not specified but is presumed to be where Smith & Nephew conducted their R&D and testing. Since it's device testing, not patient data, the retrospective/prospective distinction is not applicable.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Not Applicable: For MR safety testing of physical implants, "ground truth" as established by human experts (like radiologists interpreting images) is not the primary method. The "ground truth" for MR safety is determined by the physical properties of the materials and the design of the device, measured against established engineering and safety standards. The experts involved would be engineers and physicists specializing in MRI safety and material science.

    4. Adjudication Method for the Test Set:

    • Not Applicable: Adjudication methods (like 2+1 or 3+1 for clinical interpretations) are not relevant here. The results of MR safety testing are quantitative measurements (e.g., temperature rise, displacement force, torque, artifact size) subject to direct measurement and comparison against defined thresholds in the standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No: This is not applicable. MRMC studies are used to assess the impact of a diagnostic tool (like an AI algorithm) on human reader performance. This submission is about the physical safety of an implant in an MRI scanner.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No: This is not applicable as there is no algorithm being tested in this submission.

    7. The Type of Ground Truth Used:

    • Technical Standards and Physical Measurements: The "ground truth" for MR safety is established by international and national technical standards (ASTM, IEC, ISO/TS) that define acceptable limits for magnetic force, torque, heating, and image artifact. The studies involve physical measurements of these parameters using specialized testing equipment.

    8. The Sample Size for the Training Set:

    • Not Applicable: There is no "training set" in the context of MR safety testing of physical implants, as this is not an AI/machine learning submission.

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable: As there is no training set, this question is not relevant.

    In summary, this 510(k) submission is a prime example of a regulatory clearance based on engineering and performance testing against recognized standards for physical device properties (MR compatibility), rather than clinical efficacy studies or AI performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121443
    Date Cleared
    2012-08-13

    (90 days)

    Product Code
    Regulation Number
    888.3560
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    JOURNEY II CR KNEE SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Total knee components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis; failed osteotomies, unicompartmental replacement, or total knee replacement. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact. The Journey II Total Knee system components are indicated for use only with cement and are single use devices.

    Device Description

    Subject of this Abbreviated Premarket Notification are the Journey II Knee system. This 510(k) was prepared in accordance with the Agency's, "Draft Guidance for the Preparation of Premarket Notifications (510(k)s) for Cemented, Semi-Constrained Total Knee Prostheses," dated April 1993. The subject device is a cruciate retaining (CR) total knee system which provides the ability for greater flexion to those patients who have the anatomical capability to allow a greater flexion range. Components of this premarket notification include the following components: Cruciate retaining femoral components which will initially be available in sizes 1-10 in right and left designs in OXINIUM material. Cruciate retaining femoral components which will initially be available in sizes 1-9 in right and left designs in cobalt chrome material Cruciate retaining articular inserts which will initially be available in sizes 1-2, 3-4, 5-6, and 7-8 in right and left designs. Journey II CR articular inserts will be offered in 9-21 mm thicknesses and manufactured from cross-linked polyethylene or UHMWPE. The Journey II CR Knee system will use existing cemented Journey tibial tray and patellar components currently used with the Journey BCS Knee System (K042515), Journey II Deep Dished articular insert components (K113482), and existing patellar components of the Genesis II Knee System (K951987)

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Journey II CR Knee System." The core of this submission is to demonstrate substantial equivalence to previously cleared predicate devices, rather than conducting a de novo study against explicit acceptance criteria for device performance. Therefore, the information requested regarding acceptance criteria, performance metrics, sample sizes for test and training sets, expert involvement, and ground truth establishment, as typically found in studies for novel algorithms or AI devices, is not applicable in this context.

    Here's an analysis based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    • Not Applicable. This submission is based on demonstrating substantial equivalence to predicate devices, not on meeting specific, predefined performance acceptance criteria for a novel device. The "performance" assessment here is a comparison of design and mechanical characteristics to existing products.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. No human or animal test sets are mentioned. The "testing" referred to is mechanical testing of the device components.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. No test set with ground truth established by experts is mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No test set requiring expert adjudication for ground truth is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a mechanical device submission, not an AI/software device. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a mechanical device submission, not an AI/software device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not Applicable. For a mechanical device submission demonstrating substantial equivalence, the "ground truth" is typically the established performance and safety profiles of the predicate devices and the results of various mechanical tests performed on the new device (e.g., patellofemoral contact area, tibiofemoral constraint). The document states: "A review of the mechanical data indicates that the Journey II CR Knee System is capable of withstanding expected in vivo loading without failure."

    8. The sample size for the training set

    • Not Applicable. This is a mechanical device submission, not a machine learning or AI device.

    9. How the ground truth for the training set was established

    • Not Applicable. This is a mechanical device submission, not a machine learning or AI device.

    Summary of the Study and "Acceptance Criteria" (in the context of a 510(k) for a mechanical device):

    The "study" conducted for the Journey II CR Knee System was primarily a mechanical testing and comparison study to demonstrate substantial equivalence to previously cleared predicate devices. The "acceptance criteria" are implied by the regulatory requirements for substantial equivalence, meaning the device must be as safe and effective as a legally marketed predicate device.

    Device Performance Measures (Mechanical Testing):

    The document explicitly states that the following mechanical testing was performed to demonstrate the device's capability of withstanding expected in vivo loading:

    • Patellofemoral Contact Area Analysis
    • Tibiofemoral Contact Area Analysis
    • Patellofemoral Resistance to Lateral Subluxation
    • Tibiofemoral Constraint Testing

    The report concludes: "A review of this testing has demonstrated that there are no new issues related to the safety and effectiveness of the subject devices."

    Table of "Acceptance Criteria" (Substantial Equivalence Aspects) and Reported Device "Performance" (Comparison):

    Design Aspect ReviewedAcceptance Criteria (Implied by Predicate)Journey II CR Knee System Performance (Comparison to Predicate)
    Indications for UseMust be similar to predicate devices.Similar to predicate (rheumatoid, post-traumatic, osteoarthritis, degenerative arthritis; failed osteotomies, unicompartmental replacement, total knee replacement; primary/revision surgery where cruciate ligaments are incompetent and collaterals intact; cemented use only; single use).
    Insert Locking MechanismMust have a similar mechanism to predicate devices.Similar ("Y" in comparison table).
    Sterilization MethodMust be similar to predicate devices.Similar ("Y" in comparison table).
    MaterialMust use clinically accepted materials, similar to predicates.Femoral - OXNIUM and CoCr; Insert - XLPE and UHMWPE (similar to K111711, K101499 uses UHMWPE for insert).
    Manufacturing ProcessMust be similar to predicate devices to ensure similar quality/performance.Similar ("Y" in comparison table).
    Mechanical Performance (In Vivo Loading)Must be capable of withstanding expected in vivo loading without failure.Testing performed for: Patellofemoral Contact Area Analysis, Tibiofemoral Contact Area Analysis, Patellofemoral Resistance to Lateral Subluxation, Tibiofemoral Constraint Testing. "No new issues related to the safety and effectiveness" found.

    Missing Information & Key Points:

    • No specific numerical acceptance criteria for the mechanical tests are provided in the document. The conclusion is a qualitative statement of "no new issues." This is typical for a 510(k) where the primary goal is to show similarity to an already cleared device, implying that if the new device performs within the established range of the predicate, it is deemed acceptable.
    • No clinical data was required or presented to support the safety and effectiveness, as explicitly stated in the document ("Clinical data was not needed to support the safely and effectiveness of the subject devices"). This reinforces that the submission relies on substantial equivalence through material and design similarity, and mechanical testing.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1