Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K252056
    Manufacturer
    Date Cleared
    2025-07-24

    (23 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Isolator**®** Synergy™ EnCompass Clamp and Guide system (OLH, OSH, GPM100)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AtriCure Isolator Synergy EnCompass Clamp and Guide System is intended to ablate cardiac tissue during surgery.

    Device Description

    The Isolator Synergy EnCompass Clamp and Guide System (OLH, OSH, GPM100) is a single-use electrosurgical instrument offered in two configurations: standard length jaws (OSH), and long length jaws (OLH), each with an accessory Glidepath Magnetic Guide (GPM100). All Isolator devices are configured as vascular clamps and feature clamping jaws of various lengths and curvatures. The clamp features two pairs of opposing dual electrodes, an in-line handle with syringe-type actuation and button release mechanism. When activated, the generator delivers radiofrequency (RF) energy to the linear electrodes on the insulated jaws of the device. The Operator controls the application of this RF energy by pressing the Footswitch connected to the generator. The Guide is a single-use surgical accessory designed to facilitate the guidance of surgical instruments through tissue during cardiothoracic surgical procedures. The guide has a flexible, malleable shaft, and magnetic attachment ends that connect to the metal tip of the clamp jaws inside the jaw magnet cups.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the AtriCure Isolator® Synergy™ EnCompass Clamp and Guide System (K252056) describe modifications to an existing device, rather than a novel device requiring extensive clinical or AI-based performance studies. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are focused on demonstrating substantial equivalence to a predicate device through bench testing and non-clinical performance evaluations, rather than establishing performance against specific diagnostic accuracy or clinical outcome metrics.

    The document primarily focuses on ensuring the device modifications do not negatively impact the safety and effectiveness compared to the predicate device. As such, there is no mention of AI integration, MRMC studies, or extensive human reader performance evaluations typically associated with AI/CADe devices.

    Here's the breakdown of the information as requested, tailored to what is provided in this specific 510(k) summary:


    Acceptance Criteria and Study Proving Device Meets Acceptance Criteria for the AtriCure Isolator® Synergy™ EnCompass Clamp and Guide System (K252056)

    The acceptance criteria and supporting studies for this device are designed to demonstrate substantial equivalence to its predicate device (Isolator® Synergy™ EnCompass Clamp (OLH, OSH) and Guide (GPM100) System, K210477) following minor design and manufacturing changes. The focus is on ensuring the updated device maintains the same safety and effectiveness profile.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a substantial equivalence submission for minor modifications, the "acceptance criteria" are implied by the specific performance tests conducted to show that the modified device performs similarly to the predicate and meets relevant safety standards. The reported device performance indicates that the device met these implicit criteria.

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    Mechanical TestingMaintain structural integrity and functionality equivalent to predicate. Specific metrics for jaw aperture, clamp force, handle closure force, device-to-guide pull force, and snag resistance within acceptable ranges.Jaw Aperture Testing: Met predetermined criteria.
    Clamp Force Testing: Met predetermined criteria.
    Handle Closure Force Testing: Met predetermined criteria.
    Device-to-Guide Pull Force Testing: Met predetermined criteria.
    Snag Testing: Met predetermined criteria.
    Overall: "confirmed that the difference in technological characteristics... do not impact the safety and effectiveness of the device."
    Non-clinical Performance: Ex Vivo Ablation Comparison TestingAbility to create transmural lesions in bovine tissue equivalent to the predicate device."Results of this testing demonstrate that the Isolator Synergy EnCompass Clamp and Guide System... creates transmural lesions in a substantially equivalent manner as compared to the predicate..."
    Non-clinical Performance: Lifecycle (Reliability) TestingMeeting the predetermined design life for single patient use, maintaining electrical continuity and isolation, and physical integrity (no signs of breakdown) following specified cycles."The proposed Isolator Synergy EnCompass Clamp and Guide System... met the predetermined acceptance criteria, ensuring substantial equivalence to the previously cleared Isolator Synergy EnCompass Clamp... No new safety or performance issues were raised during testing."
    Non-clinical Performance: Usability TestingNo use errors, close calls, or use difficulties observed, and no new risks introduced by the changes."Zero use errors, close calls, or use difficulties were observed during the simulation, supporting the assessment that the use of the device is unaffected by the changes proposed, and that no new risks are introduced by these changes. No new risks were identified and no changes to the existing ratings were identified as a result of this change."
    Biocompatibility TestingCompliance with ISO 10993-1:2018 for cytotoxicity, sensitization, irritation, acute systemic toxicity, and material mediated pyrogenicity. No new or increased biocompatibility risks."Results demonstrated there were no new or increased biocompatibility risks and the proposed Isolator Synergy EnCompass Clamp and Guide System... complies with ISO 10993-1:2018." Tested for Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Material Mediated Pyrogenicity.
    Electrical Safety TestingCompliance with IEC 60601-1:2005+A1:2012+A2:2020 Ed. 3.2 and IEC 60601-2-2:2014+A1:2020."The system complies with IEC 60601-1:2005+A1:2012+A1:2020 Ed. 3.2 'General requirements for basic safety and essential performance' and IEC 60601-2-2:2014+A1:2020 'Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance.' The proposed... met the predetermined acceptance criteria..."
    Sterilization ValidationAchieving a Sterility Assurance Level (SAL) 10-6 in accordance with ANSI/AAMI/ISO 11135 and meeting residual requirements of ANSI/AAMI/ISO 10993-7."The sustainable EO cycle has been demonstrated to achieve the same Sterility Assurance Level (SAL) 10-6 as the predicate and has been validated in accordance with ISO 11135 and residuals for both EO and ECH have been evaluated to the requirements of the recognized standard ANSI/AAMI/ISO 10993-7." Endotoxin-mediated pyrogenicity demonstrated by LAL testing.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify exact numerical sample sizes for each test (e.g., number of clamps tested for mechanical properties, number of cadavers for usability). It refers to "bovine tissue" for ex vivo ablation and "cadaver lab" for usability. Given the nature of a 510(k) for minor device modifications, this level of detail regarding sample size for bench testing is typical.
    • Data Provenance: The studies are primarily bench testing and laboratory-based non-clinical performance studies.
      • Ex vivo ablation: "bovine tissue"
      • Usability testing: "cadaver lab"
      • Biocompatibility: In vitro and some in vivo tests (implied by ISO 10993 standards, typically not human data unless specified).
      • Sterilization: Tests conducted at Steris Isomedix Operation (Spartanburg, SC).
      • Retrospective/Prospective: Not applicable as these are laboratory and bench studies for device changes, not clinical data from patients.
      • Country of Origin of Data: Not explicitly stated for all tests, but given the manufacturer (AtriCure, Inc., Ohio, USA) and sterilization vendor location (Spartanburg, SC, USA), the direct tests were likely conducted in the USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is generally not applicable to a 510(k) submission for minor device modifications based on bench and non-clinical testing. "Ground truth" in this context refers to established scientific and engineering principles, and performance standards (e.g., ISO, IEC). For the usability study, "participants" are mentioned, implying a group of users (likely surgeons or clinical personnel) who operated the device in a simulated environment, but their number and specific qualifications beyond being "users" are not detailed.

    4. Adjudication Method for the Test Set

    Not applicable. As these are technical, mechanical, and biological tests, "adjudication" in the sense of expert review for ambiguous cases (like in diagnostic image interpretation studies) is not relevant. The results are objective measurements against predefined acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is specifically designed for evaluating the impact of a diagnostic aid (e.g., AI in image interpretation) on human reader performance. The Isolator® Synergy™ EnCompass Clamp and Guide System is an electrosurgical instrument, not a diagnostic device, and the modifications are minor, so such a study would not be relevant or required.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. This device does not incorporate an AI algorithm or software that operates independently to produce a diagnostic or analytical output. All testing pertains to the physical and functional performance of the medical device itself.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests conducted is based on:

    • Regulatory Standards: Compliance with recognized standards like ISO 10993 (Biocompatibility), IEC 60601 (Electrical Safety), and ANSI/AAMI/ISO 11135 (Sterilization).
    • Engineering Specifications: Predetermined design life, mechanical limits (e.g., force, jaw aperture), and functional requirements for the device.
    • Established Medical Practice: The ability to create "transmural lesions" in tissue (ex vivo), reflecting the intended therapeutic effect.
    • Predicate Device Performance: The underlying "ground truth" for demonstrating substantial equivalence is often the proven safety and effectiveness of the legally marketed predicate device. The new device must perform in a "substantially equivalent manner" in relevant aspects.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical electrosurgical clamp and guide system, not an AI/ML-driven software device that requires a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1