Search Results
Found 1 results
510(k) Data Aggregation
(28 days)
INDIGO Aspiration Catheters and Separators:
As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.
INDIGO Aspiration Tubing:
As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.
Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
The INDIGO® Aspiration System is comprised of several devices:
- INDIGO Aspiration Catheter ●
- Penumbra Aspiration Pump ●
- INDIGO Aspiration Pump Canister
- INDIGO Aspiration Tubing ●
- . INDIGO SeparatorTM
The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath and into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, and introducer. The INDIGO Separator may be provided with an introducer and torque device.
This document outlines the substantial equivalence of the "Indigo Aspiration System Lightning Aspiration Tubing" (subject device) to a predicate device (K193244) and a reference device (K192833). The review primarily focuses on demonstrating that the subject device is identical to the predicate device and that prior performance data for similar devices within the Indigo Aspiration System remain applicable.
Here's an analysis of the acceptance criteria and supporting studies based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document states: "The subject and predicate Indigo System devices are identical. Therefore, previous device performance data regarding substantial equivalence described below remain unchanged." This implies that the acceptance criteria for the subject device are met by virtue of it being identical to a previously cleared predicate device that had already satisfied those criteria.
| Acceptance Criteria Category | Reported Device Performance (Subject Device) | Supporting Evidence |
|---|---|---|
| Biocompatibility | No changes to previously provided biocompatibility data from K193244 (predicate device). | Implicitly, the subject device meets the biocompatibility criteria established for K193244. |
| Bench-Top Performance | No changes to previously provided bench-top data from K193244 (predicate device). | Implicitly, the subject device meets the bench-top performance criteria established for K193244. |
| Clinical Performance | No additional clinical data required or performed for these devices. Met by the EXTRACT-PE trial (from K192833, reference device) which demonstrated substantial equivalence for acute PE. Primary efficacy and safety endpoints were met. | Clinical study EXTRACT-PE, associated with reference device K192833. |
| Shelf-Life | No changes to previously provided shelf-life data from K193244 (predicate device). The shelf-life for the subject device is "SAME as Predicate Device (K193244)", which is 12 Months. | Implicitly, the subject device meets the shelf-life criteria established for K193244. |
| Packaging | No changes to the packaging material listing or the packaging process for the devices cleared in K193244 (predicate device). | Implicitly, the subject device meets the packaging integrity criteria established for K193244. |
2. Sample Size Used for the Test Set and Data Provenance
- Biocompatibility & Bench-Top Performance: Not explicitly stated for the subject device, as these tests were not re-performed. The document refers to data from the predicate device (K193244). The specific sample sizes and provenance for those original tests are not detailed in this document.
- Clinical Performance (EXTRACT-PE Trial): The sample size for the clinical study (EXTRACT-PE) is not specified in this document. The data provenance is also not specified (e.g., country of origin). It is stated as a "clinical study" and "trial," suggesting it was prospective. The document refers to the data from the reference device (K192833).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not available in the provided document. The document refers to previous clearances (K193244 and K192833) for performance data, but does not detail how ground truth was established for those studies.
4. Adjudication Method for the Test Set
This information is not available in the provided document.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
An MRMC study was not mentioned or implied. The clinical data cited (EXTRACT-PE) appears to focus on device efficacy and safety outcomes rather than reader performance with or without AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device (Indigo Aspiration System Lightning Aspiration Tubing) is a physical medical device, not an algorithm or AI system. Therefore, standalone performance in the context of AI is not relevant here.
7. The Type of Ground Truth Used
For the clinical study (EXTRACT-PE) related to the reference device, the document states it "demonstrated that the Indigo Aspiration System showed substantially equivalent safety and effectiveness outcomes for acute PE. The primary efficacy and safety endpoints were met." This implies that the ground truth was based on clinical outcomes data related to the efficacy of thrombus removal and patient safety in acute pulmonary embolism settings. Specific methods for establishing definitive "ground truth" for patient outcomes (e.g., imaging confirmation, pathology reports, adjudicated clinical events) are not detailed.
8. The Sample Size for the Training Set
This is not applicable. The device is a physical medical device, not an AI model that requires a training set.
9. How the Ground Truth for the Training Set was Established
This is not applicable. The device is a physical medical device, not an AI model.
Ask a specific question about this device
Page 1 of 1