Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K140838
    Date Cleared
    2014-12-05

    (247 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    IV ADMINISTRATION SET WITH HAND PUMP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IV Administration Sets with Hand Pump are intravenous administration sets intended for delivery of fluids from a container into a patient's vascular system. When the hand pump component is activated to deliver only crystalloid and colloid resuscitative fluids. These devices may patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy.

    Device Description

    The IV Administration Sets with Hand Pump are single use, disposable, intravenous administration sets with a pressure pump and air guard filter used to deliver fluids from a container into a patient's vascular system rapidly through the use of the pressure pump and/or gravity flow. These sets may be comprised of various generic components which are broadly used throughout the industry such as a bag spike, drip chamber, slide clamp, roller clamp, luer access device, stopcocks, manifolds, tubing, and luer connections. Each of these needleless components have already been cleared through the 510(k) process either individually or as part of a B. Braun IV pump 510(k) which included sets. The unique components to these sets include the manual hand pump and a 15um filter integrated within the drip chamber to reduce the potential for bubble formation within the set during use. The addition of the hand pressure pump provides the capability for delivering IV fluids more rapidly by compressing the pump by hand.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the B. Braun IV Administration Set with Hand Pump and outlines the testing conducted to demonstrate its substantial equivalence to predicate devices. However, the document does not present specific acceptance criteria or reported device performance in a quantitative manner that can be organized into a table. The performance testing section broadly states that functional performance testing was completed to demonstrate that the sets perform as intended and that results of testing demonstrate that the proposed device performs similarly to the predicate device and can be used safely and effectively according to its intended use. This is a qualitative conclusion rather than a presentation of specific metrics against predefined thresholds.

    Based on the provided text, here is an attempt to address your request, acknowledging the limitations of the input:


    1. Table of Acceptance Criteria and Reported Device Performance

    As the document does not specify quantitative acceptance criteria or precise performance metrics, the table below reflects the qualitative statements provided.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred/Qualitative)Reported Device Performance (Inferred/Qualitative)
    Functional PerformanceDevice performs as intended for fluid delivery, including gravity flow and increased flow with hand pump."Results of testing demonstrate that the proposed device performs similarly to the predicate device and can be use safety and effectively according to its intended use."
    Material BiocompatibilityMaterials are safe for clinical application."Biocompatibility test results verify that the IV Administration Sets with Hand Pump materials of construction are safe for their clinical application."
    Adherence to StandardsConformance to relevant ISO standards (e.g., fluid path integrity, luer connections, sterilization).The device was evaluated against ISO 594-1, ISO 594-2, ISO 8536-4, ISO 10993-1, and ISO 11135-1. (Implies conformance was met to achieve substantial equivalence).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the performance or biocompatibility testing. It only mentions that "Functional performance testing was completed" and "The materials of construction...were tested."

    The data provenance is not specified in terms of country of origin or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This information is not provided in the document. The testing described is primarily engineering/laboratory-based functional and biocompatibility testing, not clinical studies involving expert interpretation of results to establish a "ground truth" in the way one might for diagnostic devices.

    4. Adjudication Method for the Test Set

    This information is not provided. Given the nature of the testing (functional and biocompatibility), an adjudication method in the context of expert consensus on clinical findings is not applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating substantial equivalence through functional performance and biocompatibility testing against predicate devices and relevant industry standards, not on human reader performance with or without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    This question is not applicable as the device is an IV administration set with a hand pump, not an AI or algorithm-driven device.

    7. The Type of Ground Truth Used

    For the functional performance testing, the "ground truth" would be the expected physical and mechanical performance of an IV administration set as defined by engineering specifications and industry standards (e.g., flow rates, luer connection integrity, pressure resistance). For biocompatibility, the "ground truth" is that the materials do not elicit adverse biological responses, as determined by ISO 10993-1 standard tests.

    8. The Sample Size for the Training Set

    This question is not applicable as the device is not an AI/machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable.


    Ask a Question

    Ask a specific question about this device

    Page 1 of 1