Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    Instruments for Use with the STEALTHSTATION System, Medtronic Reusable Instruments for Use with the IPC
    POWEREASE System, Medtronic Navigated Reusable Instruments for Use with the STEALTHSTATION and IPC POWEREASE
    Systems

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

    With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

    When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    The CD HORIZON® SPIRE™ Plate is a posterior, single-level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined), spondylolisthesis, trauma, and/or tumor.

    In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

    The Navigated Disc Preparation Instruments are intended to be used to facilitate a discectomy during spinal surgery. The Navigated Trials are intended to be used to facilitate implant size selection of Medtronic intervertebral body fusion devices during spinal surgery.

    The Navigated Probe is intended to be used during pedicle and disc preparation during spinal fusion procedures.

    Navigated instruments are specifically designed for use with the StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

    IPC® System is indicated for the incision/cutting, removal, drilling and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.

    The IPC® POWEREASE® System is indicated for drilling, tapping and driving screws and working end attachments during spinal surgery, including open and minimally invasive procedures. It is also used in placement or cutting of screws, posts and rods.

    Medtronic Navigated Reusable Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Navigated Reusable Instruments are specifically designed for use with the StealthStation® System, which is indicated for any medical condition in which stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Navigated Reusable Instruments are also compatible with the IPC® POWEREASE™ System.

    Device Description

    The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples, and connecting components, as well as implant components from other Medtronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. The subject CD HORIZON® Spinal System offers cannulated, multi-axial bone screws that have an opened head (MAS) or closed head (CMAS). The subject screws offer additional angulation for anatomies or trajectories that require a sharper angle than what is typically used when placing a rod into the screw head. The subject MAS and CMAS are provided both sterile and non-sterile.

    The subject Medtronic Navigated Reusable probe is specifically designed for use in procedures where the use of stereotactic surgery may be appropriate. The subject probe is used to prepare for placement of the subject bone screws into the bone.

    The subject Medtronic Reusable drivers, torque multiplier adapter, and taps are spine preparation instruments made of high grade stainless steel. The subject drivers, torque multiplier adapter, and taps are compatible with Medtronic's IPC® POWEREASE® System may be connected to the IPC® POWEREASE® handpiece. The subject drivers, and taps can be used manually using existing Medtronic Class I Exempt quick connect handles in place of the IPC® POWEREASE® handpiece.

    The subject Medtronic Navigated Reusable drivers and taps are spine preparation instruments made of high grade stainless steel. These instruments were specifically designed for use in procedures where the use of stereotactic surgery may be appropriate. The subject drivers and taps are also compatible with Medtronic's IPC® POWEREASE™ System when connected to the POWEREASE™ handpiece or may be used manually with existing Medtronic Class I Exempt quick connect handles in place of the IPC® POWEREASE® handpiece or the NavLock™ tracker.

    AI/ML Overview

    The document provided is a 510(k) premarket notification from the FDA for several Medtronic spinal system components and instruments. It focuses on demonstrating substantial equivalence to previously cleared devices rather than providing detailed acceptance criteria and study results for a new AI/Software as a Medical Device (SaMD).

    Therefore, the requested information regarding acceptance criteria, device performance, sample sizes, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance for an AI/SaMD is not present in this document.

    This document pertains to traditional medical devices (spinal implants and surgical instruments), and the evaluation methods described are typical for such devices, involving biocompatibility and mechanical testing against recognized standards (e.g., ASTM F1717, ASTM F1798).

    Here's the information that can be extracted, albeit not in the context of AI/SaMD:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Biocompatibility:The subject devices for biocompatibility are made from widely used medical-grade materials (Cobalt-Chromium-Molybdenum alloys, Titanium-6 Aluminum-4 Vanadium ELI Alloy, Unalloyed Titanium, Stainless Steel, Radel, Silicone, PEEK). For permanent implants (bone screws), they are classified as "Implant Devices - Tissue/bone - C Permanent (>30 days)." For instruments with limited body contact (up to 24 hours), they are classified accordingly.
    Mechanical Testing (Bone Screws):The subject devices met the pre-determined acceptance criteria for all tests, indicating substantial equivalence to predicate Medtronic bone screws.
    Specific Mechanical Tests:Static Compression
    Static Torsion
    Construct Compression Fatigue
    Axial Grip
    Axial Torsion
    Flexion Extension Static
    Flexion Extension Fatigue
    Design Validation (Instruments):Performed and demonstrated that the subject instruments performed as intended.

    Missing Information (as it relates to AI/SaMD):

    • Specific numerical values or thresholds for the acceptance criteria for mechanical testing are not detailed (e.g., "Pass" is stated, but the exact pass/fail criteria and specific measured values are not provided).
    • The document does not describe the types of studies typically conducted for AI/SaMD (e.g., diagnostic accuracy studies).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • This information is not present as the studies described are in vitro mechanical and biocompatibility tests of physical devices, not data-driven performance evaluations of an algorithm. There is no "test set" in the context of data for an AI/SaMD.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This information is not present. Ground truth in the context of AI/SaMD refers to expert-validated labels or outcomes. For traditional medical devices and the tests described, ground truth is established by objective measurements against engineering standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This information is not present. Adjudication is relevant for resolving discrepancies in expert labeling of data, which is not applicable to the types of tests described in this document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This information is not present. MRMC studies are specific to evaluating diagnostic devices, particularly those involving human readers and AI assistance. This document describes physical spinal system components and instruments.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This information is not present. There is no standalone algorithm being evaluated in this submission.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • For the biocompatibility evaluation, the "ground truth" is defined by established international standards (ISO-10993) and the known safe history of the materials used in medical implants.
    • For mechanical testing, the "ground truth" is established by adherence to recognized ASTM standards (e.g., ASTM F1717, ASTM F1798) and pre-determined engineering acceptance criteria.

    8. The sample size for the training set

    • This information is not present. There is no "training set" as this is not an AI/SaMD submission.

    9. How the ground truth for the training set was established

    • This information is not present. As there is no training set for an AI/SaMD, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K153463
    Date Cleared
    2015-12-30

    (29 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    CD HORIZON Spinal System, IPC POWEREASE System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fosion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, tumor, pseudarthrosis, and/or failed previous fusion.

    Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion. With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

    When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat patients diagnosed with the following conditions: spondylolisthesis/ spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    The CD HORIZON® SPIRE™ Plate is a posterior, single-level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/ attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined), spondylolisthesis, trauma, and/or tumor.

    In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

    IPC® System is indicated for the incision/cutting, removal, drilling and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.

    The IPC®POWEREASE® System is indicated for drilling, tapping and working end attachments during spinal surgery, including open and minimally invasive procedures. It is also used in placement or cutting of screws, posts and rods.

    Device Description
    1. The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® plates, staples and connecting components, as well as implant components from other Medtronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. The subject devices include: Taps Screw Drivers The subject CD HORIZON® Spinal System device will be available in similar sizes as the predicate systems.
    2. IPC® POWEREASE® System. The CD HORIZON® reusable instruments (subject taps and screw drivers) are compatible with Medtronic's IPC® POWEREASE® System, and are spine preparation instruments which are manufactured from materials commonly used in orthopedic procedures which meet available national or international standards specifications. The subject taps and drivers may be connected to the POWEREASE® Driver or used manually if desired. These instruments are also compatible with various Medtronic spinal implant systems.
    AI/ML Overview

    This document does not contain information about the acceptance criteria and a study proving a device meets acceptance criteria in the context of diagnostic or screening performance.

    Instead, this document is a 510(k) premarket notification summary for the CD HORIZON® Spinal System and the IPC® POWEREASE® System. It focuses on demonstrating substantial equivalence to predicate devices.

    Here's what the document does provide in relation to product validation:

    1. Acceptance Criteria and Reported Device Performance:

    The document describes performance data related to biocompatibility and mechanical testing for the CD HORIZON® Spinal System devices (taps and drivers) in combination with the IPC® POWEREASE® System.

    • Biocompatibility:

      • Acceptance Criteria (implied): The devices must be safe for body contact for up to 24 hours, in accordance with "FDA's Draft Guidance for Industry and FDA Staff 'Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing'". They are manufactured from medical grade stainless steel (ASTM F899).
      • Reported Device Performance: "Medical grade stainless steel has a long history of safe and effective use in spinal surgery and biocompatibility testing is not required and no testing was conducted." The material choice is the proof of meeting the (implied) acceptance criteria.
    • Mechanical Testing:

      • Acceptance Criteria (implied): The devices must demonstrate substantial equivalence to predicate devices and not represent a "new worst case" as per "Guidance for Industry and FDA Staff -Spinal System 510(k)'s".
      • Reported Device Performance: "It was determined that subject devices do not represent a new worst case. Engineering rationales were used to demonstrate substantial equivalence." No specific quantitative metrics or results from mechanical tests are provided in this summary, only the conclusion that they met the criteria through engineering rationales.

    The other points you requested are not applicable or not explicitly detailed in this 510(k) summary because it's not a study evaluating diagnostic or screening performance of an AI/ML device.

    Here's why these points are not present:

    • 2. Sample sized used for the test set and the data provenance: Not applicable. This is not a study with a 'test set' in the sense of clinical data for diagnostic performance.
    • 3. Number of experts used to establish the ground truth... and qualifications: Not applicable. There's no ground truth establishment for diagnostic purposes here.
    • 4. Adjudication method: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable. This is not a comparative effectiveness study involving human readers or AI assistance.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is an instrument used by a human surgeon.
    • 7. The type of ground truth used: Not applicable.
    • 8. The sample size for the training set: Not applicable. This is not an AI/ML device with a training set.
    • 9. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K143019
    Date Cleared
    2015-04-08

    (170 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    CD HORIZON Spinal System, IPC POWEREASE System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

    Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

    With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

    When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis / spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    The CD HORIZON® SPIRE™ Plate is a posterior, single-level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/ attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined), spondylolisthesis, trauma, and/or tumor.

    In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

    IPC® System is indicated for the incision/cutting, removal, drilling and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.

    The IPC® POWEREASE™ System is indicated for drilling, tapping and working end attachments during spinal surgery, including open and minimally invasive procedures. It is also used in placement or cutting of screws, posts and rods.

    Device Description
    1. The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® plates, staples and connecting components, as well as implant components from other Medtronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

    2. MEDTRONIC REUSABLE INSTRUMENTS COMPATIBLE WITH THE IPC® POWEREASE® System
      The Medtronic Reusable drivers and taps that are compatible with Medtronic's IPC® POWEREASE® System are spine preparation instruments, which are manufactured from materials commonly used in orthopedic procedures which meet available national or international standards specifications. The subject taps and drivers may be connected to the POWEREASE® Driver or used manually if desired. These instruments are also compatible with various Medtronic spinal implant systems.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification for the Medtronic Sofamor Danek CD HORIZON® Spinal System and the IPC® POWEREASE® System. It focuses on demonstrating substantial equivalence to existing predicate devices, rather than providing a detailed study proving the device meets specific acceptance criteria in the manner of an AI/algorithm-driven device.

    Therefore, many of the requested sections (acceptance criteria, device performance, sample sizes for test/training, number of experts, adjudication methods, MRMC studies, standalone performance, and ground truth establishment) are not applicable or not provided in this type of regulatory submission.

    The document primarily addresses the safety and effectiveness of the device by comparing its technological characteristics and intended uses to previously cleared devices.

    Here's an attempt to answer the questions based on the provided text, indicating when information is not applicable or not found.

    1. A table of acceptance criteria and the reported device performance

    This document does not present a table of specific acceptance criteria (e.g., performance metrics like sensitivity, specificity, accuracy) or quantitative device performance metrics in the context of an algorithm's classification or prediction capabilities. This submission focuses on demonstrating substantial equivalence to predicate devices, primarily through comparison of indications for use, technological characteristics, and material safety.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. This submission does not describe a clinical study with a test set of patient data to evaluate algorithmic performance. The "performance data" mentioned refers to material biocompatibility and mechanical testing (not detailed in this summary).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. No ground truth establishment for a test set is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No adjudication method for a test set is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/algorithm device that supports human readers; it is a spinal surgical system (implants and instruments).

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/algorithm device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. No ground truth for an algorithmic performance evaluation is described. The "performance data" refers to biocompatibility testing, where the "ground truth" would be established by standard analytical and biological testing methods for materials.

    8. The sample size for the training set

    Not applicable. This document is not about an AI/algorithm requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. No training set is involved.


    Summary of Relevant Information from the Document:

    While the direct questions about "acceptance criteria" for an algorithm's performance are not applicable, the document does contain information about how the substantial equivalence of the device was demonstrated, which is the regulatory "acceptance criterion" for a 510(k) submission.

    • Device Description:

      • CD HORIZON® Spinal System: Consists of various shapes and sizes of rods, hooks, screws, CROSSLINK® plates, staples, and connecting components, which can be rigidly locked into various configurations.
      • MEDTRONIC REUSABLE INSTRUMENTS COMPATIBLE WITH THE IPC® POWEREASE® System: Spine preparation instruments (drivers and taps) manufactured from materials commonly used in orthopedic procedures. These can be connected to the POWEREASE® Driver or used manually.
    • Predicate Devices (used for comparison to establish substantial equivalence):

      • CD HORIZON® Spinal System (K141605, S.E. 07/14/2014)
      • IPC® POWEREASE® System (K111520, S.E. 10/26/2011)
    • Comparison of Technological Characteristics:

      • The subject Medtronic CD HORIZON® Reusable drivers and taps, compatible with the IPC® POWEREASE® System, have the same fundamental technology and stainless steel materials as the predicate devices.
      • Materials: Medical grade stainless steel (ASTM A564, ASTM F899, ASTM A693, Custom 465, Custom 455, 17-4 Stainless Steel).
    • Performance Data (Type of data provided, not quantitative results):

      • Biocompatibility: The instruments are classified as external communicating devices with limited body contact (up to 24 hours), according to FDA's Draft Guidance for Industry and FDA Staff: Use of International Standard ISO-10993, Biological Evaluation Method Devices Part 1: Evaluation and Testing. (No specific test results are provided in this summary, but the materials meet standards).
    • Conclusion: Based on risk analysis and supporting documentation, the subject CD HORIZON® Spinal System is substantially equivalent to the listed predicates.

    In essence, this document demonstrates that the new device is as safe and effective as previously cleared devices by showing similar design, materials, and intended use, rather than demonstrating improved or specific quantitative performance against defined clinical acceptance criteria through a study with a test set of data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111520
    Date Cleared
    2011-10-26

    (146 days)

    Product Code
    Regulation Number
    882.4310
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    IPC POWEREASE SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IPC® System is indicated for the incision / cutting, removal, drilling, and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.

    The IPC® POWEREASE™ System is indicated for drilling, tapping, and driving screws and working end attachments during spinal surgery, including open and minimally invasive procedures. It is also used in the placement of screws, or cutting of screws, posts, and rods.

    Device Description

    The purpose of this submission is to add a new driver and an accompanying software module to the IPC® . The new system is the IPC® POWEREASE™System. The IPC® POWEREASE™ System provides a powered driver that performs multiple functions and also has nerve stimulation capabilities. The IPC® POWEREASE™ System consists of the Integrated Powered Console (K081475, 10/17/2008), a driver handpiece, and cables to connect the driver, IPC® and NIM-ECLIPSE® System together. The driver is powered by the previously cleared Integrated Powered Console equipped with the appropriate software to operate the driver. The driver is used for drilling, tapping and driving screws in the pedicle, placing screws, and driving working end attachments capable of breaking set screw heads, cutting posts, and cutting rods during spinal surgery. The POWEREASE™ Driver also has nerve stimulation capabilities when connected to the Medtronic NIM -ECLIPSE® System (K061113 - 05/23/2006; K061639 -11/01/2006). An external POWEREASE™NIM Cable that connects the NIM-ECLIPSE® System to the driver enables the driver to have nerve stimulating capability. It should be noted that the IPC® does not provide the stimulation. The stimulation used with the IPC® POWEREASE™ is provided by the Medtronic NIM-ECLIPSE® System.

    The driver is equipped with a quick connect feature which allows for easy and secure attachment of different working end attachments compatible with the IPC® POWEREASE™ Driver to perform the various functions mentioned above.

    AI/ML Overview

    The provided document describes the Medtronic IPC® POWEREASE™ System, a medical device for surgical procedures, and its substantial equivalence to predicate devices. It does not contain information typically found in studies evaluating acceptance criteria and device performance for AI/ML-based medical devices.

    Therefore, I cannot extract the requested information such as:

    • A table of acceptance criteria and reported device performance.
    • Sample sizes used for test sets.
    • Number of experts used to establish ground truth or their qualifications.
    • Adjudication methods.
    • MRMC comparative effectiveness study results.
    • Standalone performance.
    • Type of ground truth used.
    • Sample size for the training set.
    • How ground truth for the training set was established.

    Instead, the document focuses on:

    • Device Description: The IPC® POWEREASE™ System, a powered driver with nerve stimulation capabilities for spinal surgery.
    • Indications for Use: Detailed uses in various surgical procedures, including drilling, tapping, driving screws, and cutting screws/posts/rods in spinal surgery.
    • Predicate Device Identification: A list of previously cleared devices to which the IPC® POWEREASE™ System is deemed substantially equivalent.
    • Comparison to Predicate Devices: A table comparing features like indications, handpiece compatibility, and nerve monitoring capability.
    • Testing: Stating that "Laboratory bench testing conducted on the POWEREASE™ System demonstrates substantially equivalent performance characteristics to the predicate devices currently on the market. Testing was performed according to IEC 60601 Medical Electrical Equipment."
    • Conclusion: The device is substantially equivalent based on bench tests, intended use, and design control activities.

    This document describes a traditional medical device clearance, not an AI/ML diagnostic or prognostic device that would typically involve the types of performance metrics and ground truth establishment you are asking about.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1