510(k) Data Aggregation

    K Number
    K232792
    Manufacturer
    Date Cleared
    2024-04-05

    (207 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    icotec Interbody Cage System (icotec Cervical Cage, icotec PLIF Lumbar Cage, icotec ETurn™ TLIF Lumbar

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    icotec Cervical Cage: The icotec Cervical Cage is intended for spinal fusion procedures at one or more levels (C2-T1) in skeletally mature patients with degenerative cervical disc disease (defined as discogenic back pain with degeneration of the disc and/or myelopathy with herniated disc and/or osteophyte formation on posterior endplates confirmed by patient history and radiographic studies), instability, trauma including fractures, deformity defined as kyphosis, or scoliosis, cervical spondylotic myelopathy, spinal stenosis, and failed previous fusion at one or more levels of the cervical spine with accompanying radicular symptoms. Implants are to be packed with autogenous bone. This device is intended to be used with a supplemental internal fixation system appropriate for use in the cervical spine. This device is intended to be used in patients who have had six weeks of non-operative treatment. When used with the icotec Anterior Cervical Plate System, the icotec Cervical Cage is intended to stabilize the cervical spinal column as an adjunct to fusion in patients with spinal infection (e.g., spondylodiscits, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression.

    icotec PLIF Cage: The icotec PLIF Cage is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of nonoperative treatment. The device is intended for use level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use without integrated fixation and must be used in conjunction with posterior supplemental fixation (e.g., pedicle screws). The device system is intended to be used with autograft. When used with the VADER® Pedicle System, the icotec PLIF Cage is intended to stabilize the lumbar spinal column as an adjunct to fusion in patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression.

    icotec ETurn® TLIF Cage: The icotec ETurn® TLIF Cage is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in sketally mature patients who have had six months of nonoperative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by radios. The device system is designed for use without integrated fixation and must be used in conjunction with posterior supplemental fixation (e.g., pedicle screws). The device system is intended to be used with autograft. When used with the VADER® Pedicle System, the icotec ETurn TLIF Cage is intended to stabilize the lumbar spinal column as an adjunct to fusion in patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression.

    Device Description

    The icotec Interbody Cage System (i.e., icotec Cervical Cage, icotec PLIF Cage, and icotec ETurn® TLIF Cage) is an interbody cage system designed to restore the intervertebral height and to facilitate intervertebral body fusion of the spine. The devices are manufactured from high strength carbon fiber reinforced polyetheretherketone (Carbon/PEEK, BlackArmor®) and incorporate a rough titanium coating. The devices are intended to be used with supplemental spinal fixation. There is one type of cervical cage and two lumbar cages (i.e., PLIF and ETurn® TLIF) included in this portfolio. Each cage is provided sterile and is available in an assortment of heights, footprints, and/or lordosis angles to accommodate patient anatomy.

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria or a study proving that a device meets those criteria. The document is a 510(k) premarket notification clearance letter from the FDA for the icotec Interbody Cage System, which primarily focuses on device description, indications for use, and a statement of substantial equivalence to predicate devices. It mentions "Performance Testing" but only states that clinical data supports substantial equivalence and explicitly notes that the data did not support additional benefit for infection-related indications due to the specific material or technological characteristics. No specific acceptance criteria or details of a study (like sample size, ground truth, expert qualifications, etc.) are provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172480
    Manufacturer
    Date Cleared
    2018-05-17

    (274 days)

    Product Code
    Regulation Number
    888.3080
    Why did this record match?
    Device Name :

    icotec Interbody Cage System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    icotec Cervical Cage: The icotec Cervical Cage is intended for spinal fusion procedures at one level (C2-T1) in skeltally mature patients with degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be packed with autogenous bone. This device is intended to be used with a supplemental internal fixation system appropriate for use in the cervical spine. This device is intended to be used in patients who have had six weeks of non-operative treatment.

    icotec PLIF Cage: The icotec PLIF Cage is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of nonoperative treatment. The device is intended for use level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use without integrated fixation and must be used in conjunction with posterior supplemental fixation (e.g. pedicle screws). The device system is intended to be used with autograft.

    icotec ETurn™ TLIF Cage: The icotec ETurn™ TLIF Cage is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of nonoperative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use without integrated fixation and must be used in conjunction with posterior supplemental fixation (e.g. pedicle screws). The device system is intended to be used with autograft.

    Device Description

    The icotec Interbody Cage System (i.e., icotec Cervical Cage, icotec PLIF Cage, and icotec ETurn TLIF Cage) is an interbody cage system designed to restore the intervertebral height and to facilitate intervertebral body fusion of the spine. The devices are manufactured from high strength carbon fiber reinforced polyetheretherketone (Carbon/PEEK, BlackArmor®) and incorporate a porous titanium coating. The devices are intended to be used with supplemental spinal fixation. There is one type of cervical cage and two lumbar cages (1.e., PLIF and ETurn TLIF) included in this portfolio. Each cage is provided sterile and is available in an assortment of heights, footprints, and lordosis angles to accommodate patient anatomy.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device (icotec Interbody Cage System), not a study describing acceptance criteria for an AI/ML powered device. Therefore, the provided text does not contain the information requested in your prompt regarding acceptance criteria, study details, ground truth, or expert involvement for an AI medical device.

    The document discusses the substantial equivalence of the icotec Interbody Cage System to legally marketed predicate devices based on indications, design, function, and performance, supported by preclinical testing according to ASTM standards. It does not involve any artificial intelligence or machine learning components.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161608
    Manufacturer
    Date Cleared
    2016-09-30

    (112 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    InterForm Cervical Interbody Cage System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The InterForm Cervical Interbody Cage System is intended for spinal fusion procedures at one level (C3-C7) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should have had at least six weeks of non-operative treatment. The device is intended to be implanted via an open, anterior approach and used with autogenous bone graft to facilitate fusion. Additionally, the InterForm Cervical Interbody Cage System is intended for use with supplemental spinal fixation systems cleared for use in the cervical spine. Patients with previous non-fusion spinal surgery at the involved level may be treated with the device.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text from the FDA 510(k) K161608 document does not contain information about the acceptance criteria or a study that proves the device meets those criteria. The document is primarily an FDA clearance letter and an "Indications for Use" statement for the InterForm Cervical Interbody Cage System.

    It states that the device is substantially equivalent to legally marketed predicate devices, which is the basis for its clearance, rather than a detailed report of a new study to establish acceptance criteria for the device's performance.

    Therefore, I cannot provide the requested information, including:

    1. A table of acceptance criteria and reported device performance
    2. Sample size and data provenance for a test set
    3. Number and qualifications of experts for ground truth establishment
    4. Adjudication method
    5. MRMC comparative effectiveness study details
    6. Standalone algorithm performance
    7. Type of ground truth used
    8. Sample size for the training set
    9. Ground truth establishment for the training set
    Ask a Question

    Ask a specific question about this device

    K Number
    K143126
    Date Cleared
    2015-02-06

    (98 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Renovis S141 Lumbar Interbody Cage System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Renovis S141 Lumbar Interbody Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade 1 spondylolisthesis at the involved level(s). Renovis S141 System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six months of non-operative treatment. The Renovis S141 System must be used with supplemental fixation cleared by FDA for use in the lumbar spine.

    Device Description

    The Renovis S141 Lumbar Interbody Cage System consist of 3 designs and each design is available in either PEEK or titanium alloy. The PLIF implants are available as a lordotic form, while the OLIF implants are provided with no lordosis. The implants are available in various heights and lengths to accommodate patients' anatomy. The implants are provided sterile, and a set of instrument for implantation is provided to facilitate implantation. Two radiographic markers made of tantalum per ASTM F650 are included in the PLIF and OLIF PEEK implants, and three radiographic markers made of tantalum per ASTM F560 are included in the TLIF PEEK implant to allow radiographic visualization.

    AI/ML Overview

    This document is a 510(k) summary for the Renovis S141 Lumbar Interbody Cage System. It primarily focuses on demonstrating substantial equivalence to predicate devices based on indications for use, materials, dimensions, function, and mechanical testing. It is not a study that proves a device meets acceptance criteria in the sense of a clinical trial or an AI/software performance study.

    Therefore, many of the requested items related to AI/software performance, ground truth, expert adjudication, and clinical study details are not applicable or cannot be extracted from this type of regulatory submission. This document describes mechanical performance testing against established standards, not a clinical study with human subjects, AI algorithms, or expert readers.

    Here's a breakdown of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Testing Standards:The Renovis S141 Lumbar Interbody Cage System was tested according to:
    ASTM F2077- Static and dynamic axial compression
    ASTM F2267- Subsidence
    - Expulsion
    Outcome:Test results demonstrate that the Renovis S141 Lumbar Interbody Cage System is substantially equivalent to the predicate devices.

    Note: The document states the device was tested to these standards and found substantially equivalent, but it does not provide specific numerical acceptance criteria or performance metrics for each test (e.g., "subsidence must be less than X mm, and observed subsidence was Y mm"). It only states the overall conclusion of substantial equivalence.

    Regarding the other requested information:

    • 2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Not Applicable / Not Provided. This document describes mechanical testing of physical devices, not a study involving human patient data or software/AI. The "test set" here refers to the physical samples of the interbody cage system tested mechanically. The sample size for these mechanical tests is not specified, nor is data provenance in the context of clinical/patient data.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not Applicable. Ground truth in the context of expert review applies to diagnostic devices or AI, not mechanical orthopedic implants. The "ground truth" for mechanical testing is defined by the physical properties and performance characteristics measured against engineering standards.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not Applicable. Adjudication is relevant for expert review or clinical event assessment, not for mechanical testing of physical devices.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not Applicable. This is a mechanical device, not an AI or diagnostic imaging device. Therefore, no MRMC study, human readers, or AI assistance is involved.
    • 6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

      • Not Applicable. This is a physical medical device (lumbar interbody cage system), not an algorithm or software. No standalone algorithm performance was assessed.
    • 7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

      • Not Applicable. For mechanical testing, the "ground truth" is defined by the established engineering standards (ASTM F2077 and ASTM F2267) and the measured physical performance (axial compression, subsidence, expulsion).
    • 8. The sample size for the training set

      • Not Applicable. This document does not describe a machine learning algorithm, so there is no "training set."
    • 9. How the ground truth for the training set was established

      • Not Applicable. As there is no training set for a machine learning algorithm, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K133967
    Manufacturer
    Date Cleared
    2014-01-31

    (38 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    INTERBODY CAGE SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Lumbar System Indications: The Aurora Spine Interbody Fusion System devices are indicated for use as intervertebral body fusion devices in skeletally mature patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Patients should have six months of non-operative treatment prior to surgery. These implants are used to facilitate fusion in the lumbar spine and are placed via either a posterior, transforaminal, lateral or anterior approach using autogenous bone. When used as interbody fusion devices these implants are intended for use with supplemental fixation systems cleared for use in the thoracolumbar spine.

    Cervical System Indications: The Aurora Spine Interbody Fusion System is indicated for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one level of the cervical spine with accompanying radicular symptoms. Patients should have six weeks of non-operative treatment prior to surgery. Cervical implants are used to facilitate fusion in the cervical spine (C2-T1) and are placed via an anterior approach using autogenous bone. When used as an interbody fusion device, supplemental fixation must be used.

    Device Description

    The Aurora Spine Interbody Fusion System, manufactured from PEEK-Optima®, consist of implants available in various foot prints, heights and lordotic configurations with an open architecture to accept packing of autograft materials. The exterior of the device has "teeth" or other generally sharp engagement members on the superior and inferior surfaces to help prevent the device from migrating once it is surgically positioned. The device comes in a PEEK or PEEK with a plasma-sprayed commercially pure titanium coating on the superior and inferior surfaces.

    AI/ML Overview

    The Aurora Spine Interbody Fusion System is a medical device, and its acceptance criteria are based on mechanical testing to demonstrate substantial equivalence to a predicate device. There is no clinical study involving human patients or AI reported in this document.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test DescriptionAcceptance Criteria (Standard)Reported Device Performance (as stated)
    Static CompressionASTM F2077"All data indicates the device is substantial equivalence to the predicate systems"
    Dynamic CompressionASTM F2077"All data indicates the device is substantial equivalence to the predicate systems"
    TorsionASTM F2077"All data indicates the device is substantial equivalence to the predicate systems"
    Compression-ShearASTM F2077"All data indicates the device is substantial equivalence to the predicate systems"
    Subsidence TestingASTM F2267"All data indicates the device is substantial equivalence to the predicate systems"
    Expulsion Testing(Not specified)"All data indicates the device is substantial equivalence to the predicate systems"
    Static Shear (Coating)ASTM F1044"All data indicates the device is substantial equivalence to the predicate systems"
    Static Tension (Coating)ASTM F1147"All data indicates the device is substantial equivalence to the predicate systems"
    Abrasion (Coating)ASTM F1978"All data indicates the device is substantial equivalence to the predicate systems"

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not applicable. The document describes non-clinical mechanical testing, not a test set of patient data.
    • Data Provenance: Not applicable. The data is from mechanical testing performed on the device components, not from patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device's acceptance is based on mechanical engineering standards, not expert clinical interpretation of data.

    4. Adjudication method for the test set:

    • Not applicable. As described, no human adjudication was involved in the mechanical testing for substantial equivalence.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. This document explicitly states, "Clinical data and conclusions were not needed for this device." There is no mention of AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is a physical medical device (interbody fusion system), not an algorithm or AI.

    7. The type of ground truth used:

    • The "ground truth" for this device's acceptance is adherence to established ASTM (American Society for Testing and Materials) standards for mechanical properties and the demonstration of substantial equivalence to a legally marketed predicate device (Tyber Medical Interbody System, K130573) based on these mechanical tests.

    8. The sample size for the training set:

    • Not applicable. There is no training set mentioned or required for this type of device submission, as it involves mechanical testing and not machine learning or AI.

    9. How the ground truth for the training set was established:

    • Not applicable. As there is no training set, there is no corresponding ground truth establishment process.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131082
    Manufacturer
    Date Cleared
    2013-08-26

    (130 days)

    Product Code
    Regulation Number
    888.3080
    Why did this record match?
    Device Name :

    INTERFORM INTERBODY CAGE SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as an intervertebral body fusion device, the InterForm Lumbar Interbody Cage System is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft to facilitate fusion. Additionally, the InterForm Lumbar Interbody Cage System is intended for use with supplemental spinal fixation systems cleared for use in the lumbar spine. Patients with previous non-fusion spinal surgery at involved level may be treated with the device.

    Device Description

    The InterForm PLIF Interbody Fusion Implants consists of cage footprint widths of 9, 11, 13mm , lengths of 20, 22, 24, 26, 30mm, ranging in height from 7mm to 14mm with lordosis of 5°, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. These implants are manufactured from implant grade polyetheretherketone (PEEK). The hollow geometry of the implants allows them to be packed with autogenous bone graft in lumbar interbody fusion procedures. The InterForm PLIF Interbody Fusion devices are intended to be implant via an open or minimally invasive, posterior approach and used with supplemental fixation and autogenous bone graft.

    The InterForm TLIF Interbody Fusion Implants consists of cage footprints widths of 9, 11, 13mm , lengths of 22, 24, 26, 28, 30, 32mm, ranging in height from 7mm to 14mm with lordosis of 5°, which can be inserted between two lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. These implants are manufactured from implant grade polyetheretherketone (PEEK). The hollow geometry of the implants allows them to be packed with autogenous bone graft in lumbar interbody fusion procedures. The InterForm TLIF Interbody Fusion devices are intended to be implant via an open or minimally invasive, transforaminal approach and used with supplemental fixation and autogenous bone graft.

    The InterForm OTLIF Interbody Fusion Implants consists of cage footprint widths of 9, 11, 13mm , lengths of 20, 22, 24, 26, 30, 32, 34mm, ranging in height from 7mm to 14mm with lordosis of 0°, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. These implants are manufactured from implant grade polyetheretherketone (PEEK). The hollow geometry of the implants allows them to be packed with autogenous bone graft in lumbar interbody fusion procedures. The InterForm OTLIF Interbody Fusion devices are intended to be implant via an open or minimally invasive, posterior or transforaminal approach and used with supplemental fixation and autogenous bone graft.

    The InterForm LLIF Interbody Fusion Implants consists of cage footprint widths of 18, 20, 22, 24mm, lengths of 40, 45, 50, 55, 60mm, ranging in height from 8mm to 18mm with lordosis of 0° or 7°, which can be inserted between two lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. These implants are manufactured from implant grade polyetheretherketone (PEEK). The hollow geometry of the implants allows them to be packed with autogenous bone graft in lumbar interbody fusion procedures. The InterForm LLIF Interbody Fusion devices are intended to be implant via a lateral approach and used with supplemental fixation and autogenous bone graft.

    The InterForm ALIF Interbody Fusion Implants consists of cage footprints of 30x24mm, and 39x30mm (DxW), ranging in height from 10mm to 20mm with lordosis of 0°, 7° or 12° which can be inserted between two lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. These implants are manufactured from implant grade polyetheretherketone (PEEK). The hollow geometry of the implants allows them to be packed with autogenous bone graft in lumbar interbody fusion procedures. The InterForm ALIF Interbody Fusion devices are intended to be implant via an open or minimally invasive, anterior approach and used with supplemental fixation and autogenous bone graft.

    All implants are packaged non-sterile to be sterilized at the hospital.

    Materials:

    PEEK Optima LT1 conforming to ASTM F2026. Unalloyed tantalum conforming to ASTM F560.

    Function:

    Maintain adequate disc space until fusion occurs.

    AI/ML Overview

    The InterForm Interbody Cage System is intended for spinal fusion procedures.
    Here's a breakdown of the acceptance criteria and the study done:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Implied by equivalence to predicate devices)Reported Device Performance
    Static Compression (per ASTM F2077)Comparable to predicate devicesIndicates equivalence to predicate devices
    Dynamic Compression (per ASTM F2077)Comparable to predicate devicesIndicates equivalence to predicate devices
    Subsidence (per ASTM F2267)Comparable to predicate devicesIndicates equivalence to predicate devices
    ExpulsionComparable to predicate devicesIndicates equivalence to predicate devices

    Explanation: The documentation explicitly states that the device is considered "substantially equivalent to the predicate devices in terms of intended use, design, materials used, mechanical safety and performances." The acceptance criteria, while not explicitly numerical, are therefore met by demonstrating performance comparable to the already legally marketed predicate devices. The study concludes that the results indicate the InterForm Interbody Cage System is equivalent.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the document. The studies listed are non-clinical tests (mechanical testing), not studies involving human or animal subjects that would generate a "test set" of data in the typical sense of evaluating AI or clinical outcomes. The provenance of the samples (e.g., specific batches of PEEK and tantalum used in testing) is not detailed, nor is it relevant given the mechanical nature of the tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the studies are non-clinical mechanical tests, not clinical evaluations requiring expert interpretation or ground truth establishment.

    4. Adjudication Method for the Test Set

    This information is not applicable as the studies are non-clinical mechanical tests.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical studies were performed." This device is a physical medical implant, not an AI or imaging device that would typically involve human reader studies.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable as the device is a physical medical implant and not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical tests would be the established mechanical properties and performance standards outlined in the ASTM F2077 and ASTM F2267 standards, and the performance characteristics of the predicate devices. For mechanical tests, the ground truth is objectively measured physical properties and behaviors of the material and device under specific conditions.

    8. The Sample Size for the Training Set

    This information is not applicable as the device is a physical medical implant and would not have a "training set" in the context of machine learning or AI. For mechanical testing, the "samples" would be the manufactured devices or materials subjected to the tests. The exact number of units tested for each non-clinical test (static compression, dynamic compression, subsidence, expulsion) is not specified, but it would typically be a statistically significant number of physical samples to ensure reliability and reproducibility of the results according to the ASTM standards.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable for the same reasons as #8. The "ground truth" for material and device properties in mechanical testing is established through standardized testing procedures and validated measurement techniques, as defined by organizations like ASTM.