Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K120012
    Date Cleared
    2012-09-11

    (252 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    INTEGRA JARIT VIDEO ASSISTED THORACIC SURGERY INSTRUMENTS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Integra™ Jarit® Video Assisted Thoracic Surgery (VATS) Instruments are manually operated instruments designed to perform specific functions such as cutting, grasping, clamping, dissecting, probing, draining, aspirating, suturing or ligating during open, mini-open, or thoracoscopic surgical procedures.

    Device Description

    Integra™ Jarit® Video Assisted Thoracic Surgery (VATS) Instruments are a family of manual, stainless steel instruments consisting of clamps, forceps, needle holders, scissors, knot tiers/pushers, and suction instruments. These reusable devices are packaged non-sterile and are steam sterilizable.

    AI/ML Overview

    The Integra™ Jarit® Video Assisted Thoracic Surgery (VATS) Instruments are manual, stainless steel surgical instruments. The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than a standalone clinical study on AI performance. Therefore, most of the requested information regarding AI study details (like sample sizes for test/training sets, expert qualifications, MRMC studies, etc.) is not applicable to this submission.

    Here's a breakdown of the acceptance criteria and performance data as presented in the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Testing PerformedAcceptance Criteria (Implicit)Reported Device Performance
    Manual Cleaning Validation (Protein Analyses) per AAMI TIR30:2003.Conformance to AAMI TIR30:2003 guidelines for cleaning, specifically for protein removal. (Details of specific protein levels are not provided.)Pass
    Mechanical Cleaning Validation (Protein Analyses) per AAMI TIR30:2003.Conformance to AAMI TIR30:2003 guidelines for cleaning, specifically for protein removal. (Details of specific protein levels are not provided.)Pass
    Pre-Vacuum (wrapped) Steam Sterilization Validation per ANSI/AAMI ST79:2010 & A1:2010 and ANSI/AAMI/ISO 14937:2009 at 270°F (132°C) with an Exposure Time of 4 minutes.Sterilization to achieve sterility assurance level (SAL). (Specific SAL not detailed, but implied by standard conformance.)Pass
    Repeated Autoclave Testing, Boiling Water Testing, Copper Sulfate Corrosion Testing and Thermal Testing per ISO 13402 and ASTM F 1089-02.Maintenance of integrity and functionality after exposure to specified conditions, including resistance to corrosion and thermal exposure. (Specific pass/fail thresholds not detailed.)Pass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable as this document describes the validation of physical surgical instruments, not an AI device or a study involving human-in-the-loop performance with an AI. The "test set" here refers to the physical instruments tested. The report indicates the testing was performed, but details on the number of instruments or specific units tested are not provided. The testing was conducted by Integra York PA, Inc., likely within the US, but explicit provenance is not stated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable. Ground truth in this context would refer to the objective standards set by the cited performance standards (e.g., a specific level of protein removal, a defined sterility assurance level, or physical properties of materials). The assessment of "pass" or "fail" is based on these objective criteria as measured in laboratory settings, not on expert interpretation of images or data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. Adjudication methods are relevant for studies involving human interpretation or subjective assessments. The performance tests here are objective measurements against defined standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This submission is for manual surgical instruments, not an AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This submission is for manual surgical instruments, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for these tests is based on established industry and international standards for medical device performance, material science, and sterilization. These include:

    • ANSI/AAMI ST79:2010 & A1:2010 (Sterilization)
    • ANSI/AAMI/ISO 14937:2009 (Sterilization)
    • ASTM TIR 30:2003 (Cleaning)
    • ASTM F 1089-02 (Corrosion)
    • ISO 13402:1995 (Resistance against autoclaving, corrosion, thermal exposure)
    • ISO 7153-1:2001 (Metallic Materials - Stainless Steel)
    • ASTM F899-11 (Wrought Stainless Steels for Surgical Instruments)

    The "ground truth" is defined by the requirements and methodologies within these standards.

    8. The sample size for the training set

    This is not applicable. There is no "training set" as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    This is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1