Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K131446
    Date Cleared
    2013-10-28

    (161 days)

    Product Code
    Regulation Number
    870.1300
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    INTEGRA JARIT HEPARIN CANNULAS, INJECTORS, AND NEEDLES

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Integra® Jarit® Heparin Cannulas, Injectors, and Needles are used in conjunction with various syringe sizes (not supplied by Integra®) for flushing, irrigation and solution injection into a vessel or cavity during vascular, bypass or other cardiovascular surgical procedures.

    Device Description

    Integra® Jarit® Heparin Cannulas, Injectors, and Needles are manual, stainless steel catheter cannulas provided in various sizes, shapes, and lengths for use during cardiovascular surgical procedures such as coronary artery bypass surgery. The cannulas are intended to be connected to a syringe filled with heparinized saline to confirm there is no leakage as well as ensure patency of harvested and/or grafted veins. The syringes are not being supplied with the cannulas nor are the syringes offered by Integra; therefore they are not part of this submission. These reusable devices are packaged non-sterile and are steam sterilizable.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Integra® Jarit® Heparin Cannulas, Injectors, and Needles:

    Context: The provided text is a 510(k) Summary for a medical device seeking substantial equivalence to predicate devices, not an AI/ML medical device. Therefore, many of the typical acceptance criteria and study components associated with AI devices (like ground truth, expert consensus, MRMC studies, training/test sets, etc.) are not applicable here. This document focuses on demonstrating physical and material properties, and reusability.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Standard/Test)Reported Device Performance
    Cleaning ValidationManual Cleaning Validation (Protein Analyses) per AAMI TIR30:2003Pass
    Mechanical Cleaning Validation (Protein Analyses) per AAMI TIR30:2003Pass
    Sterilization ValidationPre-Vacuum (wrapped) Steam Sterilization Validation per ANSI/AAMI ST79:2010 & A1:2010 and ANSI/AAMI/ISO 14937:2009 at 270°F (132°C) with an Exposure Time of 4 minutes and Drying Time of 30 minutesPass
    Material/Durability TestingRepeated Autoclave Testing, Boiling Water Testing, Copper Sulfate Corrosion Testing and Thermal Testing per ISO 13402:1995 and ASTM F 1089-02Pass
    Mechanical StrengthLaser Weld Joint Mechanical Strength TestingPass
    Brazed Joint Mechanical Strength TestingPass
    BiocompatibilityUse of recognized biocompatible materials (304 and 303 Stainless Steel) per ASTM F899-11 and ISO 7153-1, with a long history of safe and effective use.Not performed (deemed unnecessary based on material properties and history)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The tests conducted are typically on a sample of the devices designed to be representative of production, but the specific number is not disclosed in this summary.
    • Data Provenance: The tests are controlled bench tests performed on the physical devices manufactured by Integra York PA, Inc. The data is prospective for the purpose of this submission; it's generated specifically to demonstrate compliance for the 510(k). Country of origin for data generation is not specified beyond "Integra York PA, Inc." (USA).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This question is not directly applicable to this type of device submission. The "ground truth" for these tests is established by recognized industry standards (e.g., AAMI, ASTM, ISO) and the scientific principles behind them (e.g., measuring protein residue, assessing material integrity, verifying sterilization efficacy). These standards themselves are developed by committees of experts, but specific individual experts are not cited for establishing the ground truth for this particular test of this device.


    4. Adjudication Method for the Test Set

    Not applicable. This is not a study requiring human interpretation or consensus. The tests performed yield objective, measurable results (e.g., "Pass" or "Fail" based on whether predefined thresholds from the standards are met).


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This device is a manual surgical instrument, not an AI/ML-powered diagnostic or assistive tool for human readers.


    6. If a Standalone (i.e., Algorithm-Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. This is a physical, manual surgical instrument, not an algorithm.


    7. The Type of Ground Truth Used

    The "ground truth" used is defined by adherence to established national and international performance standards and test methods (e.g., ANSI/AAMI ST79, ASTM TIR 30, ISO 13402, ASTM F1089-02, ISO 7153-1). These standards specify acceptable levels for cleanliness, sterilization, corrosion resistance, and mechanical strength. In the case of biocompatibility, the ground truth is based on the recognized biocompatibility profiles of the specified stainless steel materials as per relevant material standards (ASTM F899-11 and ISO 7153-1) and their long history of safe use in medical devices.


    8. The Sample Size for the Training Set

    Not applicable. This device does not involve a "training set" as it is not an AI/ML algorithm.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1