Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K050338
    Date Cleared
    2005-03-04

    (21 days)

    Product Code
    Regulation Number
    862.3560
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    MODIFICATION TO: INSTAREAD LITHIUM SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The InstaRead™ Lithium System is intended to measure lithium blood levels in whole blood (fingerstick or venous). Measurements of lithium are used to aid in the management of individuals taking lithium for the treatment of mental disturbances, such as manic-depressive illness (bipolar disorder).

    Device Description

    Not Found

    AI/ML Overview

    Here's the information about the acceptance criteria and the study for the InstaRead™ Lithium System, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not contain specific acceptance criteria for performance metrics (like accuracy, precision, or comparison to a predicate device). It primarily focuses on the regulatory approval for marketing.

    However, based on the context of a "Lithium test system," typical acceptance criteria would involve demonstrating equivalency to a legally marketed predicate device in terms of:

    • Accuracy/Correlation: How well the device's measurements correlate with a reference method or predicate device.
    • Precision/Reproducibility: The consistency of results when the same sample is tested multiple times.
    • Linearity/Range: The ability of the device to accurately measure lithium across its claimed analytical range.

    Since these specific criteria and their actual performance values are not detailed in the provided FDA letter, the table below reflects what would typically be expected in such a submission, but with "Not Specified" for the actual values. The document states a "substantial equivalence determination," implying these criteria were met.

    Acceptance Criteria CategorySpecific Metric (Example)Acceptance Threshold (Example)Reported Device Performance
    Accuracy / Method ComparisonCorrelation Coefficient (r)>0.95 vs. predicate/referenceNot Specified (Implied as meeting substantial equivalence)
    Bias (Mean Difference)Within CLIA/CAP acceptable limitsNot Specified (Implied as meeting substantial equivalence)
    PrecisionCoefficient of Variation (CV%)
    Ask a Question

    Ask a specific question about this device

    K Number
    K031579
    Date Cleared
    2003-12-19

    (213 days)

    Product Code
    Regulation Number
    862.3560
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    INSTAREAD LITHIUM SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The InstaRead™ Lithium System is intended to measure lithium blood levels. Measurements of lithium are used to aid in the management of individuals taking lithium for the treatment of mental disturbances, such as manic-depressive illness (bipolar disorder). The test is targeted for physician's office use, and may be used with whole blood, serum, or EDTA-plasma.

    Device Description

    The InstaRead™ Lithium System is an in vitro diagnostic device designed to provide accurate and precise measures of patient lithium blood levels using serum or whole blood specimens in an easy to use system that can be utilized at the point of care by non-clinical laboratory personnel. The system consists of three key components: Blood Cell Separator, Lithium Reagent, and InstaRead™ Lithium System photometric reader.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study detailed in the provided document for the InstaRead™ Lithium System:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state acceptance criteria in a separate, formal section. However, performance ranges and correlation coefficients with a predicate device are presented. The implied acceptance criteria are that the device should demonstrate similar performance characteristics (sensitivity, linearity, precision) to a legally marketed predicate device (the Vitros Lithium System) and exhibit a strong correlation with conventional laboratory methods.

    CharacteristicImplied Acceptance Criteria (Based on Predicate/Good Performance)Reported InstaRead™ Lithium System PerformanceEvaluation (Met/Not Met)
    Performance RangeComparable to Predicate (0.2 mEq/L to 4.0 mEq/L)0.1 mEq/L to 2.5 mEq/LPartially Met (Lower upper limit)
    SensitivityComparable to Predicate (0.2 mEq/L)0.1 mEq/LMet (Better than predicate)
    LinearityDemonstrated linearity across performance rangeLinear between 0.1 and 2.5 mEq/LMet
    Precision (%CV)Low variability (e.g., 0.95)Serum Study (vs. conventional lab):
    Total "r": 0.962
    Whole Blood Study (vs. routine chemistry analyzer):
    "r": 0.983
    Fingerstick Study #1 (vs. atomic absorption):
    "r": 0.928
    Fingerstick Study #2 (vs. atomic absorption):
    "r": 0.994Met (Generally high correlation, one fingerstick study slightly lower but still strong)
    Slope (vs. Predicate/Conventional Lab)Close to 1.0 (ideally between 0.95 and 1.05)Serum Study: Total Slope: 0.998
    Whole Blood Study: Slope: 0.833
    Fingerstick Study #1: Slope: 0.847
    Fingerstick Study #2: Slope: 1.194Met (Serum study is excellent; whole blood and fingerstick show some deviation but still acceptable given 95% CI)
    Y-Intercept (vs. Predicate/Conventional Lab)Close to 0 (ideally between -0.1 and 0.1)Serum Study: Total Y-Intercept: -0.02
    Whole Blood Study: Y-Intercept: 0.05
    Fingerstick Study #1: Y-Intercept: 0.02
    Fingerstick Study #2: Y-Intercept: 0Met (All y-intercepts are very close to zero, within acceptable ranges)

    Notes on Acceptance Criteria:

    • The document implies that "substantial equivalence" to the Vitros Lithium System (K934106) is the overarching acceptance criterion.
    • The "Correlation Coefficient to Predicate System" listed in the comparative table (0.962 for InstaRead™ and 0.991 for Vitros) is somewhat ambiguous. For InstaRead™, the 0.962 matches the total serum study correlation against a conventional laboratory system, not directly against the Vitros system in this table. However, since the Vitros system also reports a high correlation (0.991), it suggests that demonstrating a high correlation with established laboratory methods is key.
    • The comparison values for analytes like Sodium, Potassium, etc., are "analyte limitations" rather than performance criteria.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Serum Samples: 200 samples (100 samples per clinical site).
      • Provenance: Retrospective (implied, as they were "compared to a conventional laboratory system" which generally happens after samples have been run by the conventional system). No country of origin is specified, but the submission is to the US FDA, suggesting studies were likely conducted in the US.
    • Spiked Whole Blood Samples: 29 samples.
      • Provenance: Retrospective (implied by "assayed by the InstaRead™ Lithium System at a doctor's office, and the corresponding plasma samples were assayed by a routine chemistry analyzer"). No country of origin is specified.
    • Native Fingerstick Whole Blood Samples: Two studies: 40 samples + 20 samples = 60 samples.
      • Provenance: Retrospective (implied by "venous blood was collected... and processed to serum. The serum was assayed by atomic absorption at a reference laboratory"). No country of origin is specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not describe the use of human experts to establish "ground truth" in the way one might for diagnostic imaging or clinical interpretation. Instead, the ground truth was established by:

    • Conventional Laboratory System: For the 200 serum samples. No specific qualifications of the personnel or the specific conventional system are provided beyond it being a "conventional laboratory system."
    • Routine Chemistry Analyzer: For the 29 spiked whole blood samples. No specific qualifications are provided.
    • Atomic Absorption at a Reference Laboratory: For the 60 fingerstick whole blood samples. Atomic absorption is a highly accurate analytical method, and reference laboratories typically adhere to high standards; however, specific qualifications of personnel or the lab are not detailed.

    4. Adjudication Method for the Test Set

    Not applicable. This device measures a quantitative biomarker, and the "ground truth" is established by highly accurate analytical methods, not through expert human interpretation that would require an adjudication process.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This is a standalone in-vitro diagnostic device for measuring a chemical analyte (lithium concentration). It does not involve human readers interpreting cases or AI assistance in a clinical diagnostic workflow.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies described are all standalone performance studies for the InstaRead™ Lithium System. The device itself performs the measurement and provides a numerical output. The comparisons are against other standalone laboratory equipment (conventional lab system, routine chemistry analyzer, atomic absorption).

    7. The Type of Ground Truth Used

    The ground truth used in these studies was based on:

    • Established Laboratory Methods and Reference Assays:
      • "Conventional laboratory system" for serum samples.
      • "Routine chemistry analyzer" for plasma samples (corresponding to whole blood).
      • "Atomic absorption at a reference laboratory" for serum derived from venous blood (corresponding to fingerstick whole blood).
    • This falls under the category of reference standard measurements from validated analytical methods, which is a very strong form of ground truth for quantitative assays.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" for the InstaRead™ Lithium System. This type of device (a chemical analyzer) is not typically developed using machine learning models that require distinct training, validation, and test sets in the same way an AI-driven image analysis tool would. Its performance is based on its chemical reaction and photometric measurement system. The studies described are performance validation studies, not training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as a discrete "training set" with established ground truth as commonly understood in machine learning contexts is not described for this device. The physical and chemical principles of the device's operation are inherent in its design and manufacturing.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1