Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K972054
    Manufacturer
    Date Cleared
    1997-08-14

    (73 days)

    Product Code
    Regulation Number
    876.5320
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indication for Use - Stimulation: The electrode is indicated for use in the treatment of urinary incontinence. The electrode is reusable if reuse instructions are followed. Indication for Use - Sensing: The electrode monitors and allows assessment of the EMG activity of the pelvic floor muscles. The electrode is reusable if reuse instructions are followed.

    Device Description

    The Reusable Vaginal and Rectal Electrodes are identical to the predicate devices except for the addition of "multi-patient use" and reuse instructions. There were no design changes required or made to allow for this change.

    AI/ML Overview

    This submission pertains to the Reusable Vaginal and Rectal Electrodes manufactured by Empi, Inc. The core change from predicate devices is the addition of "multi-patient use" and reuse instructions, indicating that the device itself is not new, but rather its usage paradigm has been expanded.

    Therefore, the acceptance criteria and study information provided in the original document focus on validating the reuse aspect rather than the inherent performance of the electrode itself.

    Here's a breakdown of the requested information based on the provided text:

    1. Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative "acceptance criteria" in terms of performance metrics (like sensitivity, specificity, accuracy). Instead, the acceptance is based on validating the sterilization process and reuse instructions to ensure the device remains safe and effective for multi-patient use.

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Sterilization EfficacyValidate the sterilization process to ensure electrodes are safe for multi-patient use."Non-clinical tests were performed to validate the sterilization process and instructions."
    Reuse InstructionsValidate that the reuse instructions are adequate to maintain the device's safety and effectiveness."Non-clinical tests were performed to validate... the reuse instructions." Device is "reusable if reuse instructions are followed."
    Design Integrity for ReuseNo design changes required or made to allow for multi-patient use."There were no design changes required or made to allow for this change [multi-patient use and reuse instructions]."

    2. Sample Size for Test Set and Data Provenance

    The document does not provide details on the sample size used for any specific tests. It only states that "Non-clinical tests were performed."

    • Sample Size for Test Set: Not specified.
    • Data Provenance: The tests were "Non-clinical," implying laboratory or bench testing rather than human subject data. Country of origin not specified, but the submission is to the U.S. FDA, suggesting testing likely occurred in the U.S. or under U.S. regulatory standards. The data is prospective for demonstrating the new reuse claim.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This section is not applicable. For non-clinical tests validating sterilization and reuse instructions, "experts" in the sense of medical professionals establishing clinical ground truth are not relevant. The ground truth would be established by scientific methods for sterility, material integrity, and functionality after sterilization cycles, typically by microbiologists, materials scientists, or engineers. The document does not specify these roles.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation of medical data by multiple experts. The described tests are non-clinical validations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This type of study is designed to assess the impact of an AI algorithm on human reader performance, which is not relevant to validating the reuse of an electrode.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No, a standalone performance study in the context of an AI algorithm was not done. This device is an electrode, not an AI algorithm. The core functionality of the device (stimulation/sensing) was already established by predicate devices. The study focused on the "standalone" aspect of its reusability through non-clinical validation.

    7. Type of Ground Truth Used

    The ground truth used was based on scientific validation of sterilization processes and materials science principles to ensure safety and functionality after multiple uses and reprocessing. This would involve:

    • Microbiological testing: To confirm sterility (e.g., Sterility Assurance Level, SAL).
    • Material integrity testing: To ensure the electrode material and electrical properties are maintained after reprocessing cycles.
    • Engineering validation: To ensure the device remains functional and safe for electrical stimulation/sensing.

    8. Sample Size for the Training Set

    Not applicable. This device is a physical electrode, not an AI algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1