Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202136
    Device Name
    IDS Cortisol
    Date Cleared
    2021-04-13

    (256 days)

    Product Code
    Regulation Number
    862.1205
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    IDS Cortisol

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS Cortisol assay is an in vitro diagnostic device intended for the quantitative determination of cortisol in human serum and plasma on the IDS system. Results are to be used in conjunction with other clinical and laboratory data to assist clinicians in the diagnosis and treatment of disorders of the adrenal gland.

    Device Description

    The IDS Cortisol assay consists of a reagent cartridge. The reagent cartridge contains multiple reagents:

    • MPE1:Magnetic particles coated rat anti-mouse monoclonal antibody in a phosphate buffer with Proclin as preservative.
    • CONJ: Cortisol coupled with an acridinium ester derivative in = phosphate buffer with Proclin as a preservative.
    • mAb: Mouse anti-cortisol monoclonal antibody in phosphate buffer with Proclin as a preservative .;
    • BUF: HEPES buffer containing Proclin as preservative .
    AI/ML Overview

    This document describes the analytical performance of the IDS Cortisol assay, an in vitro diagnostic device, and demonstrates its substantial equivalence to a predicate device (Roche Elecsys Cortisol II). The acceptance criteria and the study proving the device meets these criteria are detailed below.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this in vitro diagnostic device are primarily based on demonstrating analytical performance that is comparable to, or meets specified standards relative to, established laboratory methods and a predicate device.

    Performance CharacteristicAcceptance Criteria (from context/implied standard)Reported Device Performance (IDS Cortisol)
    PrecisionRepeatability (Within-run): Lower CV%Within-Run / Repeatability (n=80 per sample, 1 lot, 1 system):
    • 0.94 µg/dL: 7.8% CV
    • 1.84 µg/dL: 4.6% CV
    • 5.75 µg/dL: 2.4% CV
    • 13.06 µg/dL: 2.4% CV
    • 19.94 µg/dL: 1.8% CV
    • 44.63 µg/dL: 1.9% CV
      |
      | | Intermediate Precision (Within-System/Total): Lower CV% | Within-System (n=80 per sample, 1 lot, 1 system):
    • 0.94 µg/dL: 16.2% CV
    • 1.84 µg/dL: 10.9% CV
    • 5.75 µg/dL: 5.2% CV
    • 13.06 µg/dL: 3.9% CV
    • 19.94 µg/dL: 5.1% CV
    • 44.63 µg/dL: 4.2% CV
      Total (Combind 3 lots, 3 systems, n=240 per sample):
    • 0.88 µg/dL: 15.3% CV
    • 1.78 µg/dL: 10.1% CV
    • 5.75 µg/dL: 4.5% CV
    • 13.09 µg/dL: 3.3% CV
    • 20.22 µg/dL: 4.8% CV
    • 44.48 µg/dL: 5.0% CV |
      | Linearity/Reportable Range | Data should demonstrate linearity across the claimed measuring range. | Measuring Range: 0.59 to 45.00 µg/dL.
      Regression: Observed = 1.01 * Expected + 0.01 µg/dL; R²: 1.00
      (Accepted based on R² close to 1 and slope ~1, intercept ~0) |
      | Detection Limits (LoB, LoD, LoQ) | Specific quantifiable low limits. | LoB: 0.10 µg/dL
      LoD: 0.24 µg/dL
      LoQ: 0.59 µg/dL |
      | Analytical Specificity (Interference) | Non-significant bias (
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1