510(k) Data Aggregation

    K Number
    K132180
    Manufacturer
    Date Cleared
    2014-06-25

    (345 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    IDEAL LIFE GLUCO-MANAGER BLOOD GLUCOSE MONITORING SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDEAL LIFE GlucoManager™ Blood Glucose Monitoring System, Model GMM0002 is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip or alternative sites (palm, forearm, upper arm, calf, and thigh). The system is intended to be used by a single patient and should not be shared.

    The IDEAL LIFE GlucoManager™ Blood Glucose Monitoring System, Model GMM0002 is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The system is not to be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should only be done during steady state times when blood glucose is not changing rapidly.

    The IDEAL LIFE Blood Glucose Test Strips, Model AGS-1112 are for use with the IDEAL LIFE GlucoManager™ to quantitatively measure glucose in fresh capillary whole blood drawn from the fingertips, palm, forearm, upper arm, calf, and thigh.

    The IDEAL LIFE Control Solution, Model GCS-0104 is intended for use with IDEAL LIFE GlucoManager™ and IDEAL LIFE Blood Glucose Test Strips, Model AGS-1112. The control solution can be used to check that the qlucose meter and test strips are working properly and that the measurement is accurate.

    The IDEAL LIFE GlucoManager™ can wirelessly communicate with a communication gateway such as the IDEAL LIFE Pod™ or the IDEAL LIFE Gateway Application™. The IDEAL LIFE Gateway Application receives data wirelessly from IDEAL LIFE devices to transmit over the Internet from the user's mobile device. The Gateway Application is intended to aid people at home and health care professionals to review and evaluate historical blood glucose results, to support effective health care management.

    The IDEAL LIFE Gateway Application™ makes no interpretation, evaluation, medical iudgment or recommendations for treatment. This device is not intended as a substitute for medical care.

    Device Description

    The IDEAL LIFE Gluco-Manager™ Blood Glucose Monitoring System, Model GMM0002 consists of a blood glucose meter, single use test strips and control solution. The new device IDEAL LIFE Gluco-Manager™, Model GMM0002 is based on an electrochemical biosensor technology (electrochemical) and the principle of capillary action. Capillary action at the end of the test strip draws the blood into the action chamber and the blood glucose result is displayed in 5 seconds. The control solution available is used to test the performance of the device. It uses the same technological characteristics for testing with its predicate device. The IDEAL LIFE Gluco-Manager™ can be used alone to measure your blood glucose. If the user would like to transmit blood glucose information for display on a personal computer, the IDEAL LIFE Gluco-Manager™ can wirelessly communicate with a communication gateway (the optional IDEAL LIFE Pod™) or by using the Gateway Application, a mobile medical application.

    AI/ML Overview

    The IDEAL LIFE Gluco-Manager™ Blood Glucose Monitoring System, Model GMM0002, demonstrates its performance through compliance with ISO 15197: In vitro diagnostic test systems- Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. While the document asserts compliance with this standard, it does not explicitly state specific acceptance criteria values or detailed performance results in the provided text.

    Therefore, I cannot directly complete Table 1 with specific acceptance criteria and reported device performance values from the provided text. The document refers to the ISO standard as the basis for performance.

    Here's what can be inferred and what is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (from ISO 15197)Reported Device Performance (from K132180)
    System AccuracyISO 15197 (2003) Requirements:
    • For glucose concentrations
    Ask a Question

    Ask a specific question about this device