Search Results
Found 12 results
510(k) Data Aggregation
(106 days)
HydroCoil Embolic System (HES)
The HydroCoil Embolic System (HES) are intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The HES are also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The HydroFrame coils in the HydroCoil Embolic System (HES) consist of implant coil made of platinum alloy with inner hydrogel core. The coils are designed in 3D spherical structure in various loop sizes and lengths. The coil is attached to V-Trak™ or V-Trak™ Advanced Delivery Pusher via polyolefin elastomer filament. The Delivery Pusher is a variable stiffness stainless steel hypotube with platinum and stainless steel coils at the distal end. The proximal end of the Delivery Pusher is inserted into a hand held battery powered V-Grip™ Detachment Controller. When the Detachment Controller is activated, the flow of electrical current heats the polyolefin elastomer filament, resulting in detachment of the implant segment.
This document describes the HydroCoil Embolic System (HES), specifically the HydroFrame® 10 for neurovascular and peripheral vascular embolization.
Here's an analysis of the acceptance criteria and the study performed, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implied by the nature of the "Result" column, which uniformly states "All test samples passed testing", "Non-toxic", "No sensitizer response", "Non-irritant", "Non-hemolytic", "No adverse effect on coagulation time", "Non-pyrogenic", and "Negative response for mutagenicity". The specific quantitative thresholds for "passing" are not explicitly stated in this summary but would be detailed in the full test reports (e.g., PDH-HFRM-ATP for visual inspection).
Test Category | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Bench Testing | Meets established performance requirements for the test | All test samples passed testing |
Visual Inspection | Coils inspect per device drawing (PDH-HFRM-ATP) | All test samples passed testing |
Dimensional Measurement | Coil's secondary wire diameter inspectable using a microscope | All test samples passed testing |
Simulated Use | Performs as expected in an in-vitro cerebrovascular benchtop model | All test samples passed testing |
Reposition Time | Gel performs as expected when device is repositioned | All test samples passed testing |
Advancement/Retraction Force | Maximum force for advancement/retraction meets specifications | All test samples passed testing |
Expanded Gel Diameter | Expanded diameter of hydrogel (post hydration) meets specifications | All test samples passed testing |
Spring Constant | Spring constant force (maximum force to break monofilament) meets specifications | All test samples passed testing |
Weld Tensile | Coil/coupler weld tensile strength meets specifications | All test samples passed testing |
Biocompatibility (HydroFrame 10 Implant & Delivery Pusher) | ||
Cytotoxicity (MEM Elution Test) | Non-toxic | Non-toxic |
Cytotoxicity (ISO Cell Culture Agar Overlay) | Non-toxic | Non-toxic |
Sensitization (Guinea Pig Max.) | No sensitizer response | No sensitizer response |
Irritation (Intracutaneous React.) | Non-irritant | Non-irritant |
Hemocompatibility (Hemolysis) | Non-hemolytic | Non-hemolytic |
Hemocompatibility (Prothrombin Time) | No adverse effect on coagulation time | No adverse effect on coagulation time |
Systemic Toxicity (IV injection) | Non-toxic | Non-toxic |
Systemic Toxicity (Rabbit Pyrogen Test) | Non-pyrogenic | Non-pyrogenic |
Genetic Toxicology (Ames Test) | Negative response for mutagenicity | Negative response for mutagenicity |
Implantation (7-day muscle) | Non-irritant | Non-irritant |
Implantation (13-week intramuscular) | Non-irritant | Non-irritant |
Implantation (26-week intramuscular) | Non-irritant | Non-irritant |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document repeatedly states "All test samples passed testing" but does not specify the precise number of samples used for each test. This information would typically be found in the detailed test protocols or reports referenced (e.g., PDH-HFRM-ATP).
- Data Provenance: The tests performed are pre-clinical bench and biocompatibility laboratory tests. The country of origin of the data is not explicitly stated, but the company, MicroVention, Inc., is based in Tustin, California, USA, suggesting the testing was likely conducted in the USA or by labs compliant with US regulatory standards. Since these are in-vitro and animal tests, they are inherently prospective as they are specifically conducted for this submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
This document describes technical and biological tests, not studies requiring expert human interpretation of medical images or conditions to establish ground truth. Therefore, the concept of "experts used to establish ground truth" in the clinical sense (e.g., radiologists, pathologists) does not apply here. The "ground truth" for these tests is established by the predefined pass/fail criteria and objective quantitative measurements of the device's physical and biological performance against established standards (e.g., ISO 10993 series for biocompatibility).
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods like 2+1 or 3+1 refer to consensus processes among multiple human reviewers for clinical endpoints or image interpretation. The tests described are objective performance and biocompatibility assays with predefined pass/fail criteria.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was mentioned or performed. The document focuses on demonstrating substantial equivalence through technical and biocompatibility performance data, not through human reader performance with or without AI assistance.
6. Standalone (i.e. algorithm only without human-in-the-loop performance) Study
Not applicable. The HydroCoil Embolic System is a physical medical device (coils for embolization), not an algorithm or AI software. Therefore, an "algorithm only" performance study is not relevant. The performance evaluated is the device's physical and biological function.
7. Type of Ground Truth Used
The "ground truth" for the tests described is based on:
- Quantitative Bench Test Specifications: Predefined engineering and performance specifications for physical characteristics like dimensions, forces, and material properties.
- Biocompatibility Standards: Established international standards (e.g., ISO 10993 series) for evaluating the biological response to medical devices. These standards define the acceptable biological reactions (e.g., non-toxic, non-irritant).
8. Sample Size for the Training Set
Not applicable. This is not an AI/ML device that requires a training set. The device is a physical medical device.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no training set for this type of medical device.
Ask a specific question about this device
(37 days)
MicroPlex Coil System (MCS), Hydrocoil Embolic System (HES)
The MicroPlex Coil System (MCS) and HydroCoil Embolic System (HES) are intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The MCS and HES are also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The MCS devices consist of an implantable coil made of platinum alloy (MCS) or a platinum allot with an inner hydrogel core (HES). The coil is attached to a V-Trak delivery pusher via a polymer filament. The proximal end of the pusher is inserted into a hand held battery powered V-Grip Detachment Controller (sold separately). The implant segment detaches upon activation of the Detachment Controller.
The provided document contains information regarding the MicroPlex Coil System (MCS) and HydroCoil Embolic System (HES). However, it focuses primarily on the regulatory submission and technical comparison to a predicate device, rather than a clinical study evaluating the device's performance against specific acceptance criteria.
Therefore, many of the requested details about acceptance criteria, study design, sample sizes, expert involvement, and ground truth establishment are not explicitly available in the provided text.
Here's an attempt to extract the information that is present and highlight what is missing:
1. Table of Acceptance Criteria and Reported Device Performance
The document provides "Bench Test Summary" tables, which outline tests performed and their results. These represent acceptance criteria for the bench tests to demonstrate manufacturing quality and function. It does not provide acceptance criteria for clinical efficacy or safety, which would typically be part of a larger clinical study.
Test | Acceptance Criteria (Implied by "met acceptance criteria") | Reported Device Performance |
---|---|---|
Visual Inspection | Product drawings | All test samples met the acceptance criteria and passed testing. |
Pusher Resistance | Implied standard for digital multi-meter measurement (specific values not provided) | All test samples met the acceptance criteria and passed testing. |
Simulated Use | Successful: • Introduction • Tracking inside microcatheter • Deployment • Microcatheter Movement • Detachment | All test samples met the acceptance criteria and passed testing. |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: Not explicitly stated for bench tests.
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The tests are "bench tests," implying laboratory-based testing rather than patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of Experts: Not specified.
- Qualifications of Experts: Not specified. (This would be more relevant for clinical ground truth, which is not described here).
4. Adjudication method for the test set
- Adjudication Method: Not specified. Bench tests typically involve direct measurement or observation against predefined specifications, often without peer adjudication in the way clinical studies do.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No, this document describes a regulatory submission for medical devices (coils) for neurovascular embolization, not an AI-powered diagnostic device. Therefore, no MRMC study or AI-related metrics are applicable or mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone Performance: Not applicable, as this is a medical device (coils) and not an algorithm.
7. The type of ground truth used
- Ground Truth Type: For the bench tests, the "ground truth" seems to be defined by engineering specifications, product drawings, and functional requirements for the device. For example, for "Visual Inspection," the ground truth is "product drawings."
8. The sample size for the training set
- Training Set Sample Size: Not applicable. This document refers to physical device testing rather than machine learning algorithm training.
9. How the ground truth for the training set was established
- Ground Truth Establishment for Training Set: Not applicable.
Summary of Missing Information:
The provided document is a 510(k) summary for a neurovascular embolization device. It details the device's description, indications for use, comparison to a predicate device, and bench testing results. It does not contain information about clinical trials, AI performance, or studies involving human readers or expert-established ground truth in a clinical context. The "acceptance criteria" and "performance" discussed are limited to manufacturing and functional bench tests.
Ask a specific question about this device
(91 days)
MICROPLEX COIL SYSTEM (MCS), HYDROCOIL EMBOLIC SYSTEM (HES)
Intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. Also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
Treatment of cerebral aneurysms can be performed either by surgical clipping or endovascular techniques (e.g., embolization coils). Endovascular methods to embolize intracranial aneurysms with implantable platinum embolization coils were developed in the 1970s. Embolization coils including the MicroVention MCS and HES products are now widely used to treat all aneurysms.
The MCS consists of an implantable coil made of bare platinum alloy (Platinum/Tungsten), and the HES consists of an implantable coil made of the same platinum alloy with a hydrogel inner core. The coil is attached to a delivery pusher via a polyolefin elastomer material. The delivery pusher contains radiopaque positioning markers at the distal end. The proximal end is inserted into a hand held battery powered V-Grip™ Detachment Controller. When the Detachment Controller is activated, the flow of electrical current heats the polyolefin elastomer filament, resulting in detachment of the implant segment. The V-Grip is packaged and sold separately.
The modified MCS and HES devices are substantially equivalent to the cleared predicate devices with regard to intended use, principal of operation, materials, manufacturing processes, packaging configuration, and sterilization method.
The provided Special 510(k) K132952 describes the MicroPlex Coil System (MCS) and HydroCoil Embolic System (HES). The submission primarily focuses on the substantial equivalence of these devices to previously cleared predicate devices, with a minor modification related to the V-Trak Advanced delivery pusher. As such, the study summarized is a design verification and validation bench test, along with biocompatibility testing, rather than a clinical study involving human subjects or algorithm performance.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Test / Test Description | Acceptance Criteria | Reported Device Performance |
---|---|---|
Visual Inspection (and measurements using device drawing) | Per product drawing | Pass (Met established criteria) |
Simulated Use (assess deployment, repositioning, and detachment in an aneurysm) | All performance ratings shall be ≥ 3 | Pass (Met established criteria) |
Pusher Resistance (measure resistance through a microcatheter) | 36.7-53.0 Ω | Pass (Met established criteria) |
Detachment Zone Tensile (measure breakforce at detachment zone) | ≥ 0.08 lbf | Pass (Met established criteria) |
Biocompatibility Tests (for HES Coil Implant, MCS Coil Implant, and Delivery Pusher) | Requirement (ISO standards) | Passed (for all listed tests, e.g., MEM Elution, Sensitization, Hemolysis, etc.) |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes for the individual bench tests. It refers to "design verification and validation bench tests," which typically involve a specified number of units per test to ensure statistical significance or manufacturing quality. However, the exact number of coils or components tested is not provided.
The data provenance is pre-clinical bench testing conducted in a lab environment. There is no information regarding country of origin for this testing, but it can be inferred to be from the manufacturer, MicroVention, Inc., located in Tustin, California, U.S.A. This is retrospective in the sense that it's performed on manufactured devices for regulatory submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
This information is not applicable as the reported tests are physical and functional bench tests, not clinical studies requiring expert interpretation of medical images or patient outcomes to establish ground truth. The "acceptance criteria" are based on engineering specifications and established industry standards (e.g., ISO for biocompatibility).
4. Adjudication Method for the Test Set
This is not applicable as there were no subjective expert assessments that would require adjudication. The tests were objective measurements against defined acceptance criteria.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A MRMC comparative effectiveness study was not conducted and is not described in this submission. This type of study would typically be performed for devices involving human interpretation (e.g., imaging devices with AI assistance), which is not the case for this neurovascular embolization device.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
A standalone performance study was not conducted and is not described. This submission is for a physical medical device (coils and delivery system), not a software algorithm or AI.
7. Type of Ground Truth Used
The "ground truth" for the bench tests was based on:
- Engineering Specifications: For visual inspection and measurements ("Per product drawing").
- Performance Ratings: Based on simulated use scenarios, likely defined by internal engineering or clinical experts within the company to assess deployment, repositioning, and detachment.
- Physical Properties/Standards: For pusher resistance and detachment zone tensile strength, these refer to measurable physical properties with specific numerical ranges.
- International Standards (ISO 10993 series): For biocompatibility testing, compliance with these established standards serves as the ground truth for biological safety.
8. Sample Size for the Training Set
This is not applicable. This submission is for a physical medical device, not a machine learning model, so there is no "training set."
9. How the Ground Truth for the Training Set Was Established
This is not applicable as there is no training set for a machine learning model.
Ask a specific question about this device
(30 days)
HYDROCOIL EMBOLIC SYSTEM (HES)
The HES is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The HES consists of implantable coil made of platinum alloy with inner hydrogel core. The helical-shaped implantable coil is available in various outer dimensions and lengths. The coil is attached to a V-Trak delivery pusher. The proximal end of the pusher is inserted into a hand held battery powered V-Grip Detachment Controller (sold separately). The implant segment detaches upon activation of the Detachment Controller. The modified HES is substantially equivalent to the cleared predicate devices with regard to intended use, principal of operation, materials, manufacturing processes, packaging configuration, and sterilization method.
The provided text describes a Special 510(k) submission for a modification to the HydroCoil Embolic System (HES), specifically to the HydroFill component. This is a submission demonstrating substantial equivalence to a predicate device, not a de novo clearance or a novel device submission. Therefore, the "acceptance criteria" discussed are primarily related to bench testing demonstrating that the modified device performs similarly to the predicate device, rather than clinical efficacy endpoints. The study described is a series of bench tests.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
For this Special 510(k), the "acceptance criteria" are implied by demonstrating that the modified device met the same criteria as the predicate device in a series of bench tests. The reported performance is that all tests met these implied criteria.
Test | Acceptance Criteria (Implied) | Reported Device Performance (Subject Device) |
---|---|---|
Visual Inspection | Match predicate device's visual characteristics | Met same criteria as predicate |
Dimensional Measurement | Match predicate device's dimensional specifications | Met same criteria as predicate |
Simulated Use | Perform similarly to predicate in: | Met same criteria as predicate |
• Introduction | • Introduction | • Introduction |
• Tracking | • Tracking | • Tracking |
• Reposition / Deployment | • Reposition / Deployment | • Reposition / Deployment |
• Detachment | • Detachment | • Detachment |
• Overall Performance | • Overall Performance | • Overall Performance |
Spring Constant | Match predicate device's spring constant | Met same criteria as predicate |
Gel Expansion | Match predicate device's gel expansion properties | Met same criteria as predicate |
Weld Tensile | Match predicate device's weld tensile strength | Met same criteria as predicate |
Advancement / Retraction Force | Match predicate device's forces | Met same criteria as predicate |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not specify the exact sample sizes (number of devices or tests performed) for each bench test. It only states that the modified device "Met same criteria as predicate." The data provenance is internal to MicroVention, Inc., and is derived from bench testing (presumably in a laboratory setting). It is not clinical data, so terms like "country of origin" or "retrospective/prospective" are not applicable in a clinical sense.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
Not applicable. For bench tests demonstrating substantial equivalence, ground truth is established by engineering specifications, material properties, and performance benchmarks against the predicate device, not by expert clinical consensus.
4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set
Not applicable. Adjudication methods are typically used in clinical studies involving interpretation of medical images or outcomes, where human expert consensus is required to establish ground truth or resolve discrepancies. Bench tests rely on objective measurement and predefined engineering criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a submission for a modification to a medical device (embolic coils), not an AI-powered diagnostic or therapeutic system. An MRMC study is not relevant to this type of device and submission.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an algorithm or AI device.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the bench tests effectively consisted of the established performance and material specifications of the predicate device. The goal was to demonstrate that the modified device performed equivalently to these known specifications. This is an engineering/performance-based "ground truth."
8. The Sample Size for the Training Set
Not applicable. This is not an AI/machine learning device that requires a training set. The data presented is from verification and validation bench testing of the physical device.
9. How the Ground Truth for the Training Set was Established
Not applicable. As noted above, there is no training set for this type of device.
Ask a specific question about this device
(24 days)
HYDROCOIL EMBOLIC SYSTEM (HES)
The HES is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The HydroSoft Plus HES consists of implantable coil made of platinum alloy with inner hydrogel core. The helical-shaped implantable coil is available in various outer dimensions and lengths. The coil is attached to a V-Trak delivery pusher. The proximal end of the pusher is inserted into a hand held battery powered V-Grip Detachment Controller (sold separately). The implant segment detaches upon activation of the Detachment Controller.
With the exception of the coil dimensions (see comparison table below), the HydroSoft Plus line extension is substantially equivalent to the cleared predicate devices with regard to intended use, principal of operation, materials, manufacturing processes, packaging configuration and sterilization method.
This document describes the 510(k) summary for MicroVention, Inc.'s HydroCoil Embolic System (HES) - HydroSoft Plus Line Extension. This submission is for a medical device (neurovascular embolization device), not an AI/ML powered device, so many of the requested categories are not applicable.
Here's an analysis of the provided text, focusing on the acceptance criteria and the study (bench testing) performed:
1. Table of Acceptance Criteria and the Reported Device Performance:
Acceptance Criteria (Test) | Reported Device Performance |
---|---|
Visual Inspection | Met same criteria as predicate |
Dimensional Measurement | Met same criteria as predicate |
Simulated Use | Met same criteria as predicate |
- Introduction | Met same criteria as predicate |
- Tracking | Met same criteria as predicate |
- Reposition / Deployment | Met same criteria as predicate |
- Detachment | Met same criteria as predicate |
- Overall Performance | Met same criteria as predicate |
Spring Constant | Met same criteria as predicate |
Gel Expansion | Met same criteria as predicate |
Weld Tensile | Met same criteria as predicate |
Advancement / Retraction Force | Met same criteria as predicate |
Predicate/Subject Device Comparison:
This table highlights the specific device characteristics that define the acceptance criteria for a "substantially equivalent" device:
Feature | Predicate Device (K112226) | Subject Device (HydroSoft Plus Line Extension) | Acceptance Criteria Met? |
---|---|---|---|
Coil shape | Helical | Same | Yes |
Implant diameter | 2 mm - 10 mm | 5 mm - 24 mm | Difference |
Implant restrained length | 2 cm - 10 cm | 15 cm - 50 cm | Difference |
Deliver pusher length | 185 cm | Same | Yes |
Main coil wire material | Platinum/Tungsten (92/8 %) alloy | Same | Yes |
Coupler material | Platinum (90%) / iridium (10%) | Same | Yes |
Adhesive material | Dymax 1128-AM-VT | Same | Yes |
Implant to pusher material | Polyolefin Elastomer | Same | Yes |
Stretch resistant filar material | Polyolefin Elastomer | Same | Yes |
Gel material | Hydrophilic Copolymer | Same | Yes |
MRI compatibility | Yes | Yes | Yes |
Method of supply | Sterile, single use | Same | Yes |
Packaging configuration | Dispenser, pouch, carton | Same | Yes |
Note on Differences: The "Implant diameter" and "Implant restrained length" are explicitly different between the predicate and subject devices. The core of this 510(k) submission is to demonstrate that despite these dimensional differences (which represent the "line extension"), the new device is still substantially equivalent in terms of safety and effectiveness, performing to "the same criteria as predicate" in the bench tests.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: The document does not specify the exact sample sizes for each bench test conducted (e.g., how many coils were subjected to "Weld Tensile" testing). It only states that the tests were performed and "Met same criteria as predicate."
- Data Provenance: The tests are "bench tests," meaning they were conducted in a laboratory setting. There is no mention of country of origin of data or whether it was retrospective or prospective, as these terms typically apply to clinical studies involving human or animal subjects, which is not the case for this type of submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This is a medical device 510(k) submission based on bench testing of physical properties and performance, not diagnostic imaging or AI assessment that requires expert ground truth labeling. The "ground truth" for these tests is defined by established engineering and material science standards and the performance of the predicate device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods are used in clinical trials or studies where subjective assessments or discrepancies in expert interpretations require resolution. Bench tests involve objective measurements against predefined specifications.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This submission is for a physical medical device (coil embolization system), not an AI-powered diagnostic or assistive tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an algorithm or AI device. The testing described is performance testing of a physical product.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the bench tests is the performance specifications and established criteria of the predicate device (K112226), along with general engineering and material science standards applicable to such devices. The new HydroSoft Plus line extension is considered to meet the acceptance criteria if its performance in these tests is "the same" as or equivalent to the predicate device.
8. The sample size for the training set
Not applicable. There is no "training set" as this is not an AI/ML device.
9. How the ground truth for the training set was established
Not applicable. There is no "training set" for physical device bench testing.
Ask a specific question about this device
(57 days)
HYDROCOIL EMBOLIC SYSTEM (HES)
The HES is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The HES consists of implantable coil made of platinum alloy with inner hydrogel core. The helical-shaped implantable coil is available in various outer dimensions and lengths. The coil is attached to a V-Trak delivery pusher. The proximal end of the pusher is inserted into a hand held battery powered V-Grip Detachment Controller (sold separately). The implant segment detaches upon activation of the Detachment Controller. The RD11-006 HES is a line extension which is substantially the same as the predicate devices (K070656, K080666, K091641). The implantable coil portion of both the predicate and the subject device is constructed from platinum alloy wire. The RD11-006 implantable coil is constructed of an oval-shaped platinum alloy wire, whereas the predicate devices are constructed of a round platinum alloy wire. All other materials and design features are the same for both the predicate and subject devices.
This document describes an FDA 510(k) summary for the HydroCoil Embolic System (HES) and asserts its substantial equivalence to a predicate device. It specifically focuses on a line extension with an oval-shaped platinum alloy wire, while the predicate devices used a round wire. There is no information in the provided text about a study (clinical or otherwise) that uses human readers, ground truth, or statistical analysis in the context of device performance. The entire submission is a declaration of equivalence based on bench testing.
Here's an analysis based on the provided text:
1. Table of Acceptance Criteria and the Reported Device Performance:
The document indicates that the acceptance criteria for the subject device are the "same criteria as predicate" for all bench tests. The reported device performance is that it "Met same criteria as predicate."
Acceptance Criteria (Bench Test) | Reported Device Performance |
---|---|
Visual Inspection | Met same criteria as predicate |
Dimensional Measurement | Met same criteria as predicate |
Simulated Use (Introduction, Tracking, Reposition/Deployment, Detachment, Overall Performance) | Met same criteria as predicate |
Spring Constant | Met same criteria as predicate |
Gel Expansion | Met same criteria as predicate |
Weld Tensile | Met same criteria as predicate |
Advancement / Retraction Force | Met same criteria as predicate |
2. Sample size used for the test set and the data provenance:
- Sample Size for Test Set: Not explicitly stated. The "Bench Test Summary" implies testing of device features, but the number of units tested is not provided.
- Data Provenance: The data is from bench testing, presumably conducted by the manufacturer, MicroVention, Inc., in the USA (Tustin, California). It is not retrospective or prospective in the clinical sense, but rather a series of laboratory experiments.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This is not applicable as the provided document details a 510(k) submission for a medical device (embolic coils) where the primary evidence presented is bench testing for substantial equivalence to a predicate device. There is no mention of a performance study involving human readers or experts to establish ground truth in the context of diagnostic interpretation.
4. Adjudication method for the test set:
This is not applicable for the reasons stated above. The evaluation is based on objective measurements from bench tests, not expert adjudication of diagnostic images or outcomes.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable. The device is an embolic coil, not an AI-powered diagnostic tool. No MRMC study was performed or is relevant for this type of device based on the information provided.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
This is not applicable. The device is a physical medical implant, not an algorithm.
7. The type of ground truth used:
The "ground truth" for this submission is related to the physical and mechanical properties of the device. The "acceptance criteria" for the bench tests effectively serve as the ground truth against which the subject device's performance is measured. This includes:
- Meeting dimensional specifications.
- Achieving established thresholds for mechanical properties (e.g., spring constant, weld tensile strength, advancement/retraction force).
- Performing as expected during simulated use (introduction, tracking, deployment, detachment).
- Exhibiting expected gel expansion characteristics.
These are objective, measurable characteristics that confirm functional equivalence to the predicate device.
8. The sample size for the training set:
This is not applicable. There is no "training set" in the context of an embolic coil device submission for bench testing. The device is not learning from data.
9. How the ground truth for the training set was established:
This is not applicable for the reasons stated above.
Ask a specific question about this device
(98 days)
HYDROCOIL EMBOLIC SYSTEM (HES)
The HydroCoil® Embolic System (HES) is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.
The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The HydroCoil® Embolic System (HES) is based on electrical coil detachment. The HES consist of an implantable coil attached to a delivery system called a V-TRAK™ Delivery Pusher. The Delivery Pusher is a variable stiffness, stainless steel and tapered mandrel. Two silver electrical leads run along the outside of the mandrel from the proximal to the distal end. Platinum and stainless steel wires are wound around the distal end of the mandrel to form the electrical heater and provide kink-resistance. Two outer layers of PET tubing cover the distal end of the pusher assembly. A layer of polyimide tubing covers the proximal end. The proximal end of the HES coils incorporates a platinum coupler for attachment to the Delivery Pusher. A polyolefin elastomer filament is attached to the proximal end of the coil. This filament runs through the inner lumen of the coil coupler and is attached to the distal end of the Delivery Pusher.
The Delivery Pusher is powered by a hand-held, battery-powered V-GRIP™ Detachment Controller designed specifically for the HES. A gold-plated stainless steel connector at the proximal end of the Delivery Pusher is used to connect it to the Detachment Controller, The Detachment Controller is provided separately. The coil is delivered to treatment site on the Delivery Pusher through standard neuro-interventional micro-catheters. A removable introducer sheath on the outside of the Delivery Pusher assists in the placement of the HES into the micro-catheter. Once the coil is deployed, the proximal end of the Delivery Pusher is connected to the Detachment Controller. When the Detachment Controller is turned on, the flow of current melts the polyolefin elastomer filament at the coupler/pusher junction resulting in coil detachment.
The provided text describes a 510(k) premarket notification for the HydroCoil® Embolic System (HES) and states that performance testing has demonstrated its equivalence to predicate devices. However, the document does not contain specific details about acceptance criteria, the study design, sample sizes, ground truth establishment, or expert involvement as requested in the prompt.
Therefore, I cannot populate the table or answer most of the questions directly from the provided text.
Here's a summary of what can be extracted and what cannot:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Not specified in document | "Performance testing has demonstrated that the HydroCoil® Embolic System (HES) with the HES-HC-HS (10) coils is equivalent in performance to the predicate devices." |
2. Sample size used for the test set and the data provenance
- Sample size for test set: Not specified in the provided document.
- Data provenance (e.g., country of origin of the data, retrospective or prospective): Not specified in the provided document.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not specified in the provided document.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not specified in the provided document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not specified in the provided document. This device is an embolization coil system, not an AI-assisted diagnostic tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable as this is a medical device (embolization coil system), not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not specified in the provided document. The performance testing aimed to show equivalence to predicate devices, likely based on engineering performance metrics (e.g., detachment reliability, coil stability, mechanical properties) rather than diagnostic accuracy against a "ground truth" as typically understood in AI/imaging studies.
8. The sample size for the training set
- Not applicable as this is a medical device, not an AI/ML algorithm requiring a training set.
9. How the ground truth for the training set was established
- Not applicable as this is a medical device, not an AI/ML algorithm.
Additional Information from the text:
- Device Name: HydroCoil® Embolic System (HES)
- Intended Use: Endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae; also for vascular occlusion of blood vessels within the neurovascular system and peripheral vasculature.
- Regulatory Status: 510(k) cleared, determined to be "substantially equivalent" to predicate devices.
- Nature of Study: Performance testing was conducted to demonstrate equivalence to predicate devices, focusing on operating principle, method of applications for use, design, materials, packaging, and sterilization.
Ask a specific question about this device
(74 days)
MICROPLEX COIL SYSTEM (MCS) AND HYDROCOIL EMBOLIC SYSTEM (HES)
The MicroPlex® Coil System (MCS) and HydroCoil® Embolic System (HES) are intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.
The MCS and HES are also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other neurovascular abnormality, and for arterial and venous embolizations in the peripheral vasculature.
The MicroPlex® Coil System (MCS) and HydroCoil® Embolic System (HES) consist of an implantable coil attached to a delivery system called a Delivery Pusher. The Delivery Pusher is powered by a hand-held, battery-powered Detachment Controller designed specifically for the MCS and HES. The Detachment Controller is provided separately.
The coil is delivered to treatment site on the Delivery Pusher through standard neurointerventional micro-catheters. A removable introducer sheath on the outside of the Delivery Pusher assists in the placement of the MCS and HES into the micro-catheter. During deployment, the proximal end of the Delivery Pusher is connected to the Detachment Controller. When the controller is turned on, the coil detaches.
The provided documentation describes a 510(k) premarket notification for the MicroPlex® Coil System (MCS) and HydroCoil® Embolic System (HES). This type of submission relies on demonstrating substantial equivalence to legally marketed predicate devices, rather than establishing acceptance criteria and conducting a primary study to prove that the device meets those criteria.
Therefore, much of the requested information regarding specific acceptance criteria, detailed study designs, and performance metrics in the typical sense of a de novo or PMA submission is not explicitly available in this document. The document emphasizes equivalence through design, materials, and existing performance.
Here's an analysis based on the information provided, highlighting what is and isn't available:
There are no specific acceptance criteria for performance metrics (such as sensitivity, specificity, accuracy) provided in this document as it is a 510(k) for substantial equivalence. The "performance data" referred to is primarily comparative testing against predicate devices to demonstrate equivalent performance, rather than a standalone study establishing new performance benchmarks.
Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Not explicitly stated as quantifiable metrics. The implicit acceptance criterion is demonstration of substantial equivalence to predicate devices. | "Performance testing has demonstrated that the MicroPlex® Coil System (MCS) and HydroCoil® Embolic System (HES) with a modified detachment system are equivalent in performance to the predicate devices." |
Study Information (Based on Substantial Equivalence Approach)
-
Sample size used for the test set and the data provenance:
- Not Applicable (N/A) / Not provided in this document. No "test set" in the sense of a clinical or analytical performance study is detailed. The performance data mentioned relates to comparative testing for substantial equivalence, likely bench or animal testing, not a clinical trial with a defined test set of patients or medical images.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- N/A / Not provided. Ground truth establishment by experts is typically done for diagnostic or prognostic devices, often involving human interpretation of medical data. This document describes a medical device for embolization, where performance is assessed through engineering and possibly animal studies, not expert ground truth on a test set.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- N/A / Not provided. As no clinical "test set" requiring expert interpretation and adjudication is described, this information is not relevant to this submission type.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC study was not done. This device is an embolization coil system, not an AI-assisted diagnostic or imaging system. Therefore, MRMC studies are not applicable.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- N/A / Not applicable. This is a physical medical device, not a software algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- N/A / Not explicitly stated for performance testing. For a mechanical device, ground truth often relates to engineering specifications, material properties, detachment mechanisms, and possibly animal studies demonstrating successful embolization. The document only vaguely mentions "performance testing."
-
The sample size for the training set:
- N/A / Not applicable. There is no "training set" in the context of an AI/ML algorithm or a device that learns from data.
-
How the ground truth for the training set was established:
- N/A / Not applicable. As there is no training set, this information is not relevant.
In summary: The provided document is a 510(k) submission for substantial equivalence. It does not detail specific acceptance criteria or a comprehensive study with the types of metrics and methodologies commonly found in submissions for AI/ML devices or novel diagnostic tools. The "performance data" mentioned refers to comparative testing that established equivalence to predicate devices in design, materials, manufacturing, and intended use.
Ask a specific question about this device
(28 days)
MODIFICATION TO: HYDROCOIL EMBOLIC SYSTEM (HES)
The HydroCoil® Embolic System (HES) is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.
The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The HydroCoil® Embolic System (HES) consists of an implantable coil attached to a fluid injection delivery system called a delivery pusher. The delivery pusher is a variable stiffness stainless steel tube with several outer layers of PET tubing. A luer hub at the proximal end of the pusher is used for system de-airing and coil detachment. The HES coils are platinum helical coils with an outer layer of a hydrophilic polymer. The polymer material is a cross-linked copolymer of acrylamide and acrylic acid. The proximal end of the coil incorporates a coupler for attachment to the delivery pusher. PET tubing is heatshrunk over the coupler/pusher junction in order to attach the coil to the delivery pusher. The coil is delivered to the treatment site on the delivery pusher through standard neurointerventional micro-catheters. An introducer sheath on the outside of the delivery pusher assists in the placement of the HES into the micro-catheter. A 1.0-cc syringe is used for system de-airing and a 0.25-cc syringe is used for coil detachment.
Below is a summary of the acceptance criteria and the study that proves the device meets the acceptance criteria, based on the provided text.
Acceptance Criteria and Device Performance
The provided document describes the HydroCoil® Embolic System (HES), an embolization coil. The core of its acceptance for market entry (510(k) clearance) is based on its substantial equivalence to previously cleared predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria Category | Specific Criteria | Reported Device Performance (HydroCoil® Embolic System - HES) |
---|---|---|
Substantial Equivalence | Demonstrates equivalence in operating principle, method of application, indications for use, design, materials, packaging, and sterilization to predicate devices. | The modified HydroCoil® Embolic System (HES) is substantially equivalent to the HydroCoil® Embolic System (HES) cleared on October 22, 2003 (K032590) and December 30, 2003 (K033836), and to the Matrix™ Stretch Resistant Detachable Coils cleared on May 14, 2003 (K031168) in these characteristics. |
Performance | Performance testing demonstrates equivalence to predicate devices. | "Performance testing has demonstrated that the modified HydroCoil® Embolic System (HES) is equivalent in performance to the predicate devices." (Specific performance metrics are not detailed in the provided text, but the overall conclusion of equivalence is stated.) |
Indications for Use | Must align with the indications of the predicate device or be deemed safe and effective for the stated uses. | The HES is intended for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities (e.g., arteriovenous malformations, arteriovenous fistulae), and for vascular occlusion within the neurovascular system and arterial/venous embolizations in the peripheral vasculature. These indications are accepted by the FDA based on substantial equivalence. |
2. Sample size used for the test set and the data provenance:
The provided text does not specify a "test set" in the context of a clinical study with real patient data. The evaluation is based on "performance testing" demonstrating equivalence to predicate devices. This typically refers to
- Bench testing: In-vitro mechanical, material, and functional tests.
- Animal studies: Pre-clinical evaluations of safety and performance in an animal model, if required.
- Comparison to existing clinical data from predicate devices: This is implied by the substantial equivalence claim, rather than new clinical data for this specific modified device.
Therefore, details like country of origin or whether it was retrospective/prospective are not applicable as a separate clinical "test set" for this specific device's clearance is not described. The substantial equivalence relies on the established safety and effectiveness of the predicate devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
This information is not provided as the clearance is based on substantial equivalence and "performance testing," not a clinical study requiring expert ground truth assessment of patient cases.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
This information is not provided as a clinical test set requiring adjudication is not described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable. The device (HydroCoil® Embolic System) is a medical implant (embolization coil), not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study comparing human reader performance with and without AI assistance was not conducted.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
This is not applicable. The device is an embolization coil, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For the purpose of this 510(k) clearance, the "ground truth" for the device's acceptance is based on:
- Established safety and effectiveness of predicate devices: The primary ground truth is that the predicate devices (HydroCoil® Embolic System and Matrix™ Stretch Resistant Detachable Coils) have already been deemed safe and effective and are legally marketed.
- Performance testing (bench and potentially animal studies): These tests likely compared the new device's physical properties, mechanical performance, and perhaps immediate biological interactions to specifications or established performance of the predicate devices.
8. The sample size for the training set:
This information is not provided and is not applicable in the context of this device's regulatory submission, as it is not an AI/machine learning product requiring a training set.
9. How the ground truth for the training set was established:
This information is not provided and is not applicable for the reasons mentioned above.
Ask a specific question about this device
(20 days)
HYDROCOIL EMBOLIC SYSTEM (HES)
The HydroCoil® Embolic System (HES) is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.
The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The HydroCoil® Embolic System (HES) consists of an implantable coil attached to a fluid injection delivery system called a delivery pusher. The delivery pusher is a variable stiffness stainless steel tube with several outer layers of PET tubing. A luer hub at the proximal end of the pusher is used for system dc-airing and coil detachment. The HES prominer platinum helical coils with an outer layer of a hydrophilic polymer. The polymer material is a cross-linked copolymer of acrylamide and acrylic acid. The proximal end of the coil incorporates a coupler for attachment to the delivery pusher. PET tubing is heatshrunk over the coupler/pusher junetion in order to attach the coil to the delivery pusher. The coil is delivered to the treatment site on the delivery pusher through standard neurointerventional micro-catheters. An introducer sheath on the outside of the delivery pusher assists in the placement of the HES into the micro-catheter. A 1.0-cc syringe is used for system de-airing and a 0.25-cc syringe is used for coil detachment.
The provided document is a 510(k) premarket notification for the HydroCoil® Embolic System (HES). This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving that a device meets specific acceptance criteria based on detailed performance studies with clinical outcomes in the way a PMA (Premarket Approval) submission would.
Therefore, the document does not contain the information requested in your prompt regarding acceptance criteria, a specific study proving it, sample sizes for test/training sets, expert consensus details, or MRMC studies.
Instead, it states the following:
1. Table of Acceptance Criteria and Reported Device Performance:
- This information is not present in the document. The document states that "Performance testing has demonstrated that the HydroCoil® Embolic System (HES) with the HES-HC-S (10) coils is equivalent in performance to the predicate devices." However, it does not provide specific acceptance criteria or the numerical results of this performance testing.
2. Sample size used for the test set and the data provenance:
- This information is not present in the document.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not present in the document.
4. Adjudication method for the test set:
- This information is not present in the document.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:
- This information is not present in the document. The device in question is an embolization coil, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- This information is not present in the document. The device is an embolization coil, not an algorithm.
7. The type of ground truth used:
- This information is not present in the document.
8. The sample size for the training set:
- This information is not present in the document.
9. How the ground truth for the training set was established:
- This information is not present in the document.
In summary, the provided document focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics, indications for use, design, materials, packaging, and sterilization. It does not provide the detailed performance study data, acceptance criteria, or ground truth establishment methods typically found in clinical trials or AI algorithm validation studies.
Ask a specific question about this device
Page 1 of 2