Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K160347
    Date Cleared
    2016-03-03

    (24 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 2MP/3MP LCD Monitor (HMD2G21) is intended to be used in displaying and viewing digital images (excluding digital mammography) for review and analysis by trained medical practitioners.

    The 5MP LCD Monitor (HMD5G21) is intended to be used in displaying digital images for review and analysis by trained medical practitioners, including digital mammography.

    Device Description

    The 2MP/3MP/5MP LCD Monitor (HMD2G21/HMD3G21) is a display system for medical viewing, with high resolution 1600x1200(HMD2G21)/2048 x 1536(HMD3G21)/ 2560 x 2048(HMD5G21), built-in brightness stabilization circuit, front sensor and ambient light sensor, stable brightness and persistent calibration can be guaranteed. The display can support both landscape and portrait mode. The antiglare screen can prevent display from reflection under highlight conditions, make the image and display clearer.

    AI/ML Overview

    The document describes a 510(k) premarket notification for several Hisense LCD monitor models (HMD2G21, HMD3G21, HMD5G21). The submission aims to demonstrate substantial equivalence to legally marketed predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by compliance with IEC 60601-1 and IEC 60601-1-2 standards, and by demonstrating performance comparable to or better than the predicate devices. The study detailed below focuses on comparing the Hisense monitors' specifications to those of their respective predicate devices.

    Here's a consolidated table comparing the proposed Hisense devices (HMD2G21, HMD3G21, HMD5G21) with their identified predicate devices (BARCO Nio 2MP, Nio 3MP, Nio 5MP) based on the provided "General Comparison" tables.

    FeatureAcceptance Criteria (Predicate Device Performance)Hisense HMD2G21 PerformanceHisense HMD3G21 PerformanceHisense HMD5G21 Performance
    Intended UseDisplaying and viewing digital images (excluding digital mammography for 2MP/3MP; including digital mammography for 5MP) for review and analysis by trained medical practitioners.Same as predicateSame as predicateSame as predicate
    Panel Size & Type21.3", TFT LCD display (2MP, 5MP); 21.3", TFT LCD display (3MP)21.3", TFT LCD display21.2", IPS-NEO, antiglare21.3", IPS-NEO, antiglare
    Pixel Pitch0.27 mm (2MP); 0.2115 mm (3MP); 0.165 mm (5MP)0.27 mm0.21075 mm0.165 mm
    Native Resolutions1600x1200 (2MP); 2048x1536 (3MP); 2560x2048 (5MP)1600x12002048x15362560x2048
    Brightness1650 cd/m2 (2MP); 800 cd/m2 (3MP); 700 cd/m2 (5MP)1900 cd/m21700 cd/m21200 cd/m2
    Contrast Ratio850:1 (2MP); 1700:1 (3MP); 800:1 (5MP)1400:11400:11200:1
    Network InterfaceUSB (1 Up, 2 Downstream)USB (1 Up, 2 Downstream)USB (1 Up, 2 Downstream)USB (1 Up, 2 Downstream)
    Active Display Size (HxV)432mm x 324mm (2MP); 433mm x 325mm (3MP); 422mm x 338mm (5MP)432.0mm x 324.0mm431.6mm x 323.7mm422.4mm x 337.9mm
    Operating Temp.0°C ~ 35°C (2MP); 0°C ~ 40°C (3MP, 5MP)0°C ~ 40°C0°C ~ 40°C0°C ~ 40°C
    Transport/Storage Temp.-20°C ~ 60°C-20°C ~ 60°C-20°C ~ 60°C-20°C ~ 60°C
    Operating Humidity8% ~ 80% (non-condensing)20% ~ 80%20% ~ 80%20% ~ 80%
    Transport/Storage Humidity5% ~ 85% (2MP); 5% ~ 95% (3MP, 5MP)10% ~ 90%10% ~ 90%10% ~ 90%
    Power Capacity<70W (2MP); ≤40W (3MP); 61W (5MP)≤60W≤60W<80W
    Input Voltage100240V (2MP, 3MP); 100250V (5MP)DC24V/2.2A (2MP/3MP); DC24V/3A (5MP)DC24V/2.5ADC24V/3A
    UsabilityButton operation, LED indicatorButton operation, LED indicatorButton operation, LED indicatorButton operation, LED indicator
    Mode of OperationContinuous operationContinuous operationContinuous operationContinuous operation
    BiocompatibilityNo direct/indirect patient contact, no evaluation needed.No direct/indirect patient contact, no evaluation needed.No direct/indirect patient contact, no evaluation needed.No direct/indirect patient contact, no evaluation needed.
    SterilityNot neededNot neededNot neededNot needed
    Electrical & Mechanical Safety & Thermal SafetyComplying with IEC 60601-1Complying with IEC 60601-1Complying with IEC 60601-1Complying with IEC 60601-1
    EMC EvaluationComplying with IEC 60601-1-2Complying with IEC 60601-1-2Complying with IEC 60601-1-2Complying with IEC 60601-1-2

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a clinical study in terms of patient data or image sets. It is a technical comparison (bench testing) between the proposed devices and predicate devices. Therefore, there is no "test set" in the context of clinical images or patient data. The "sample size" implicitly refers to the three models of proposed devices (HMD2G21, HMD3G21, HMD5G21) being compared against their respective predicate models (BARCO Nio 2MP, Nio 3MP, Nio 5MP).

    The data provenance is from the manufacturer's (Qingdao Hisense Medical Equipment Co., Ltd.) specifications and the specifications of the predicate devices. This type of comparison is typically based on published technical specifications and internal testing, not retrospective or prospective clinical data. The country of origin for the proposed device manufacturer is China.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is not a clinical study involving experts to establish ground truth for image interpretation. The "ground truth" for this submission are the technical specifications and performance characteristics of the predicate devices and the relevant IEC safety standards.

    4. Adjudication Method for the Test Set

    Not applicable. There was no clinical test set requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This submission is for medical display monitors, not an AI or CAD (Computer-Aided Detection/Diagnosis) software. The study is a technical comparison of display performance, not an evaluation of human reader performance with or without AI.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is not an algorithm or AI product.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" used for this submission is based on:

    • Technical Specifications: Published specifications of the predicate devices (BARCO Nio 2MP, Nio 3MP, Nio 5MP) and the proposed Hisense devices.
    • Regulatory Standards: Compliance with established medical device safety and performance standards, specifically IEC 60601-1 (Electrical, Mechanical & Thermal safety) and IEC 60601-1-2 (Electromagnetic Compatibility - EMC).
    • Intended Use Statements: Formal declarations of intended use, which must be comparable to the predicate.

    8. The Sample Size for the Training Set

    Not applicable. There is no AI or machine learning component, so no training set is relevant.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1