Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K050169
    Date Cleared
    2005-04-05

    (69 days)

    Product Code
    Regulation Number
    870.1875
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HL 200 ELECTRONIC STETHOSCOPE is intended for medical diagnostic purposes only. It can be used for the amplification heart, lung and other body sounds with the use of selective frequency and can be used on any patient undergoing a physical assessment.

    Device Description

    HL 200 ELECTRONIC STETHOSCOPE is a non-sterile, reusable Electronic Stethoscope . HL 200 ELECTRONIC STETHOSCOPE, equipped with built-in micro-computer chip, includes state-of-the-art amplification and filtering systems. It is ideal for picking up difficult-to-hear heart and other body sounds. The features of HL 200 ELECTRONIC STETHOSCOPE is as follows

    • Providing three-mode choices for adequate requirements: heart sound mode, lung sound mode and combination mode
    • Eight volumes levels provide up to 18X amplification of acoustic stethoscopes
    • Electronic design with fully eliminates sound loss and resonance effects associated with traditional acoustic stethoscopes. With amplified signal, you could even hear the slight sounds much clearer than the traditional one.
    • Ergonomically designed with comfortable handholding, subtle with material and touch
    • Four-segments display shows the battery conditions could inform you to know the battery using state
    • Auto shut-off after 3 minutes for extending battery life
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the HL 200 ELECTRONIC STETHOSCOPE, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device PerformanceStudy Type
    Performance SpecificationConformance to applicable standardsConforms to IEC 60601-1 and IEC 60601-1-2 & related requirements.Bench Testing
    SafetyConformance to applicable standardsConforms to IEC 60601-1 and IEC 60601-1-2 & related requirements.Bench Testing
    EMC RequirementsConformance to applicable standardsConforms to IEC 60601-1 and IEC 60601-1-2 & related requirements.Bench Testing
    Technological CharacteristicsNo new questions of safety or effectiveness compared to predicate.Differences in technological characteristics do not raise any new questions of safety or effectiveness.Bench Testing, Comparison to Predicate

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify a sample size for a test set. This submission relies on substantial equivalence to a predicate device, supported by bench testing. The data provenance is also not explicitly stated as country of origin or retrospective/prospective. The description suggests testing was likely performed by the manufacturer.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of experts to establish a ground truth for a test set in the context of clinical performance or diagnostic accuracy. The evaluation is focused on technical and electrical safety standards.

    4. Adjudication Method for the Test Set

    Since there is no mention of a test set involving clinical data or expert review for diagnostic accuracy, there is no adjudication method described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done or at least not reported in this 510(k) summary. The submission focuses on substantial equivalence based on technical standards and comparisons to a predicate's characteristics.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This device is an electronic stethoscope, which is a hardware device for sound amplification. It is not an "algorithm only" device in the sense of AI. Its performance is inherent in its electrical and acoustic properties, which are evaluated by bench testing, not as a standalone algorithm's diagnostic accuracy. So, technically, the "standalone" performance is the bench testing, but not in the AI sense.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this submission is adherence to electrical safety and EMC (electromagnetic compatibility) standards (IEC 60601-1 and IEC 60601-1-2), and the functional characteristics of the device as compared to a predicate electronic stethoscope. This is determined through bench testing against these established technical specifications, rather than clinical ground truth like pathology or expert consensus on diagnoses.

    8. The Sample Size for the Training Set

    The concept of a "training set" is typically associated with machine learning or artificial intelligence models. As this is an electronic stethoscope (a hardware device), there is no training set in that context. The device's design and manufacturing processes are likely informed by engineering principles and possibly prior product development, but not by a formal "training set" like an AI model.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no "training set" in the AI sense for this device, the question of how its ground truth was established is not applicable. The "ground truth" in its development would relate to engineering specifications, component performance, and adherence to design requirements, verified through internal testing and quality control processes.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1