Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K981794
    Manufacturer
    Date Cleared
    1998-07-29

    (69 days)

    Product Code
    Regulation Number
    862.1145
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HiChem Calcium Reagent is intended for the quantitative determination of total calcium in serum, plasma and urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

    Device Description

    The HiChem Calcium Reagent determines calcium by its reaction with arsenazo III to form a blue complex. The resulting increase in absorbance at 650 nm is proportional to the calcium concentration in the sample. The HiChern Calcium Reagent is an adaptation of the arsenazo III calcium method, first reported by Michaylova and Illikova and is intended to be used either as a manual procedure or on clinical analyzers which can automate the required manipulations.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the HiChem Calcium Reagent, based on the provided text:

    Acceptance Criteria and Device Performance

    The provided text describes two main scenarios for the HiChem Calcium Reagent: the manual procedure and its use on the Beckman® SYNCHRON CX® Systems (secondary reagent application). The acceptance criteria are largely implied by the reported performance, as the study aims to demonstrate substantial equivalence to an existing predicate device.

    Table of Acceptance Criteria and Reported Device Performance

    Performance MeasureAcceptance Criteria (Implied)HiChem Calcium Reagent (Manual Procedure) Reported PerformanceHiChem Calcium Reagent (SYNCHRON CX® Systems) Reported Performance
    LinearityDemonstrate linearity over a specific range.Linear between 0.1 to 16 mg/dL. Regression: (HiChem) = -0.1 + 0.980 x (Standard Conc.), r² = 1.000, sy.x = 0.14 mg/dL.Linear from at least 2.0 mg/dL to 15.0 mg/dL. Regression: (HiChem) = 0.0 + 0.967 x (Standard Conc.), r² = 1.000, sy.x = 0.08 mg/dL.
    PrecisionMeet specified SD for various control levels.Serum Control 1 (6.69 mg/dL): within-run SD 0.21 mg/dL, total SD 0.24 mg/dL Serum Control 2 (12.68 mg/dL): within-run SD 0.29 mg/dL, total SD 0.30 mg/dL Urine Pool 1 (2.41 mg/dL): within-run SD 0.11 mg/dL, total SD 0.11 mg/dL Urine Pool 2 (11.10 mg/dL): within-run SD 0.13 mg/dL, total SD 0.16 mg/dLSerum Control 1 (6.89 mg/dL): within-run SD 0.056 mg/dL, total SD 0.064 mg/dL Serum Control 2 (10.03 mg/dL): within-run SD 0.085 mg/dL, total SD 0.089 mg/dL Serum Control 3 (13.21 mg/dL): within-run SD 0.088 mg/dL, total SD 0.090 mg/dL Urine Pool 1 (3.24 mg/dL): within-run SD 0.048 mg/dL, total SD 0.059 mg/dL Urine Pool 2 (11.64 mg/dL): within-run SD 0.075 mg/dL, total SD 0.095 mg/dL
    Method Comparison (Serum/Plasma)Substantial equivalence to Beckman® SYNCHRON® Calcium Reagent.n=92. Regression: (HiChem) = -0.3 + 1.020 × (BMD® Results), r² = 0.887, s(y.x) = 0.21 mg/dL.n=153. Regression: (HiChem) = -0.3 + 1.005 x (Beckman® Results), r² = 0.924, s(y.x) = 0.20 mg/dL.
    Method Comparison (Urine)Substantial equivalence to Beckman® SYNCHRON® Calcium Reagent.Regression: (HiChem) = 0.2 + 1.022 × (BMD® Results), r² = 0.998, s(y.x) = 0.15 mg/dL.Regression: (HiChem) = 0.1 + 1.031 x (Beckman® Results), r² = 0.996, s(y.x) = 0.16 mg/dL.
    Chemical Additives (Interference)Biases less than a specified threshold (e.g., < 0.1-0.2 mg/dL).Biases less than 0.1 mg/dL for sodium heparin, lithium heparin, lithium iodoacetate.Biases less than 0.2 mg/dL for sodium heparin, lithium heparin, lithium iodoacetate.
    SensitivityAchieves claimed sensitivity (e.g., 0.2 mg/dL).Claimed: 0.2 mg/dL. Observed sensitivity limit: 0.15 mg/dL (based on 3 SD of 30 replicate study).Claimed: 2.0 mg/dL. Observed sensitivity limit: < 0.15 mg/dL (based on 3 SD of 30 replicate within-run precision study).
    Calibration Stability (SYNCHRON CX®)Imprecision < 0.26 mg/dL or 2.6% (Beckman's claim).Not applicable to manual procedure.14 days, imprecision < 0.26 mg/dL or 2.6%.
    Onboard Reagent Stability (SYNCHRON CX®)Imprecision < 0.26 mg/dL or 2.6% (Beckman's claim).Not applicable to manual procedure.30 days, imprecision < 0.26 mg/dL or 2.6%.

    Study Details

    The studies described are primarily designed to demonstrate substantial equivalence to the predicate device, the Beckman® Systems Calcium Reagent. This means the new device performs similarly enough to the legally marketed device that it is considered safe and effective for the same intended use.

    1. Sample Size used for the test set and the data provenance:

      • Manual Procedure:
        • Linearity: 23 data points (df) for linearity standards spanning 0-18 mg/dL.
        • Precision: 30 replicates (n) for each of 4 control specimens (2 serum, 2 urine).
        • Method Comparison: 92 mixed serum and plasma specimens.
        • Sensitivity: 30 replicates for a diluted serum control.
      • SYNCHRON CX® Systems Application:
        • Linearity: 29 data points (df) for six linearity standards spanning "at least 2.0 mg/dL to 15.0 mg/dL."
        • Precision: 60 replicates (n) for each of 5 control specimens (3 serum, 2 urine).
        • Method Comparison: 153 mixed serum and plasma specimens.
        • Sensitivity: 30 replicates for a diluted serum control.
      • Data Provenance: Not explicitly stated but implied to be laboratory-generated data for performance characteristics and potentially patient samples for method comparisons. No specific country of origin is mentioned, nor is it explicitly stated as retrospective or prospective, though the nature of laboratory validation studies often implies prospective data collection for the specific tests performed.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This type of study for a laboratory reagent does not typically involve human experts establishing ground truth in the same way as, for example, an imaging AI study. The "ground truth" for these tests comes from:
        • Assigned values of linearity standards: These are precisely manufactured solutions with known calcium concentrations.
        • Assigned values of control sera/urine: These are commercially available reference materials with established target values.
        • Results from the predicate device (Beckman® SYNCHRON® Calcium Reagent): For method comparison studies, the predicate device's results serve as the reference standard against which the new device is compared.
      • Therefore, no information on "number of experts" or their "qualifications" is provided or relevant in this context.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This is a quantitative reagent study, not an imaging or qualitative diagnostic study requiring expert adjudication of results. The results are numerical measurements.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is a study of a laboratory reagent's performance in quantitative measurement, not an AI-assisted diagnostic tool involving human readers.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Yes, the performance data presented (linearity, precision, method comparison, sensitivity, stability) for both the manual procedure and the automated SYNCHRON CX® Systems application represent the standalone performance of the HiChem Calcium Reagent. While a human initiates the tests and interprets the results, the data itself reflects the reagent's analytical capability. There is no "human-in-the-loop" interaction in the sense of an AI interpreting images for a human.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Reference Method/Comparator: The primary "ground truth" for comparison is the Beckman® SYNCHRON® Calcium Reagent, which is the legally marketed predicate device. The study aims to show that the HiChem reagent produces results substantially equivalent to this established method.
      • Known Reference Values: For linearity and precision studies, the ground truth is based on known, pre-assigned concentrations in linearity standards and commercial control materials. These values are derived through rigorous analytical methods by the manufacturers of these standards/controls.
    7. The sample size for the training set:

      • Not applicable. This is a chemical reagent, not a machine learning algorithm that requires a "training set" in the computational sense. The "development" or "formulation" of the reagent would involve extensive R&D and optimization, but it's not parallel to training a neural network on a dataset.
    8. How the ground truth for the training set was established:

      • Not applicable for the reason stated above.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1