Search Results
Found 1 results
510(k) Data Aggregation
(27 days)
HEALIX ADVANCE KNOTLESS BR ANCHOR (6.5MM)
The Healix Advance Knotless Anchors are indicated for use in the following procedures for reattachment of soft tissue to bone:
Shoulder
- Rotator Cuff
- Biceps Tenodesis
The proposed Healix Advance Knotless Anchor is a one piece implantable cannulated, threaded anchor designed to secure soft tissue to bone. The anchor is provided loaded on a disposable inserter driver device. The proposed anchors will be offered in a 6.5 mm size. The proposed 6.5 mm Healix Advance Knotless BR Anchor is manufactured from "biocryl rapide" material.
The provided document does not describe the acceptance criteria or a study that proves the device meets specific performance criteria in the context of an AI/ML device.
Instead, this document is a 510(k) summary for a medical device called the "HEALIX ADVANCE KNOTLESS BR ANCHOR (6.5mm)". It focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving the performance of an AI/ML algorithm against predefined acceptance criteria.
The device in question is a physical bone anchor for soft tissue reattachment, not an AI/ML diagnostic or treatment planning system. Therefore, most of the requested information (like sample sizes for test/training sets, experts for ground truth, MRMC studies, standalone performance, etc.) is not applicable to this type of medical device submission.
Here's the relevant information that can be extracted, framed within the context of a traditional medical device submission for substantial equivalence:
1. Table of Acceptance Criteria and Reported Device Performance
For this type of device, "acceptance criteria" are tied to demonstrating that the new device (6.5mm anchor) performs similarly from a safety and effectiveness perspective to the predicate devices (4.75mm and 5.5mm anchors). The "reported device performance" are the results of non-clinical bench testing.
Acceptance Criteria (Implied for Substantial Equivalence) | Reported Device Performance |
---|---|
Device performance is comparable to predicate devices in terms of mechanical properties | Product Design Verification and Design Validation activities, such as: |
- Insertion Torque tests
- Torque to Failure tests
- Anchor Pullout tests
Results demonstrated that the proposed device is substantially equivalent to the predicate devices regarding performance and safety. (Specific numerical values or thresholds are not provided in this summary but would be in the full submission). |
| No new technological characteristics are introduced that would raise new safety/effectiveness concerns. | The proposed 6.5mm anchor has the same design principles and is manufactured from the same "biocryl rapide" material as the predicate 4.75mm and 5.5mm anchors. The only change is size. |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: Not explicitly stated as this is non-clinical bench testing. The "test set" would refer to the number of anchors or test specimens used for the mechanical tests (Insertion Torque, Torque to Failure, Anchor Pullout).
- Data Provenance: Not applicable in the context of clinical data for AI/ML. The "data" comes from laboratory bench testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. Ground truth in this context is determined by validated engineering test methods and measurements, not expert clinical interpretation.
4. Adjudication method for the test set
- Not applicable. This is not a clinical study involving human readers.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is not an AI/ML system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is not an AI/ML system.
7. The type of ground truth used
- For the non-clinical testing, the "ground truth" would be established by validated engineering standards and measurement equipment for mechanical properties (e.g., torque, force).
8. The sample size for the training set
- Not applicable. This device is not an AI/ML system.
9. How the ground truth for the training set was established
- Not applicable. This device is not an AI/ML system.
Ask a specific question about this device
Page 1 of 1