Search Results
Found 1 results
510(k) Data Aggregation
(160 days)
HEALICOIL SUTURE ANCHOR WITH REGENESORB
The Smith & Nephew HEALICOIL Absorbable Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications:
Shoulder:
Bankart lesion repairs
Slap lesion repairs
Capsular shift or capsulolabral reconstructions
Acromioclavicular separation repairs
Deltoid repairs
Rotator cuff tear repairs
Biceps tenodesis
Knee:
Extra-capsular repairs:
Medial collateral ligament
Lateral collateral ligament
Posterior oblique ligament
Patellar realignment and tendon repairs:
Vastus medialis obliquous advancement
Iliotibial band tenodesis.
Foot & Ankle:
Hallux valgus repairs
Medial or lateral instability repairs/reconstructions
Achilles tendon repairs/reconstruction
Midfoot reconstructions
Metatarsal ligament/tendon repairs/reconstructions
Elbow:
Ulnar or radial collateral ligament reconstructions
Lateral epicondylitis repair
Biceps tendon reattachment
The HEALICOIL Absorbable Suture Anchor is a bioabsorbable suture anchor provided with up to three strands of non-absorbable sutures, pre-loaded on an insertion device and is available in two sizes, 4.75mm and 5.5mm.
Here's a summary of the acceptance criteria and study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The provided text for the K123393 submission for the HEALICOIL Absorbable Suture Anchor does not explicitly state numerical acceptance criteria with pass/fail thresholds. Instead, it uses the concept of "substantial equivalence" to predicate devices.
Acceptance Criteria (Stated as Substantial Equivalence Goal) | Reported Device Performance |
---|---|
Insertion Strength substantially equivalent to predicate devices | Demonstrated substantially equivalent to predicate devices |
Pull-out Strength substantially equivalent to predicate devices | Demonstrated substantially equivalent to predicate devices |
In vitro degradation substantially equivalent to predicate devices | Demonstrated substantially equivalent to predicate devices |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample size used for the test set in the performance studies. It mentions that "performance testing was conducted," but no details on the number of samples for insertion strength, pull-out strength, or in vitro degradation are provided.
The data provenance is not explicitly stated (e.g., country of origin). The studies appear to be prospective in nature, given they are performance tests specifically conducted for this submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts
This information is not applicable to the type of device and study described. This device is a physical medical implant, and its performance is evaluated through biomechanical testing (insertion strength, pull-out strength, degradation), not through subjective interpretation by human experts to establish ground truth.
4. Adjudication Method for the Test Set
This is not applicable as the studies described are biomechanical performance tests, not studies requiring expert adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic or imaging devices where human readers interpret data, often with and without AI assistance, to assess the AI's impact on human performance. This submission is for a physical orthopedic implant.
6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study
No, a standalone algorithm-only study was not done because this device is a physical orthopedic implant, not an algorithm or AI system.
7. Type of Ground Truth Used
The "ground truth" for the performance studies would be the objective measurements obtained from the biomechanical tests themselves. For example:
- Insertion Strength: The force required to insert the anchor, measured by a load cell.
- Pull-out Strength: The force required to pull the anchor out, measured by a load cell.
- In vitro Degradation: Measurements of material properties or mass loss over time in a controlled environment.
This is objective, quantifiable data from physical experiments, not expert consensus, pathology, or outcomes data in the typical sense of AI/diagnostic device ground truth.
8. Sample Size for the Training Set
This information is not applicable. This device is a physical medical implant, not an AI/machine learning model that requires a training set of data.
9. How the Ground Truth for the Training Set was Established
This information is not applicable, as there is no training set for this device.
Ask a specific question about this device
Page 1 of 1