    Date Cleared
    2021-05-14

    (228 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Gx-IVF, Gx-TL, Gx-MOPS PLUS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Gx-IVF™ medium is intended for preparation and handling of gametes, for in vitro fertilisation and intrauterine insemination.

    Gx-TL™ medium is intended for culture of embryos from fertilisation to the blastocyst stage and for embryo transfer.

    Gx-MOPS™ PLUS medium is intended for handling and manipulating oocytes and embryos in ambient atmosphere.

    Device Description

    The subject devices are culture and handling media consisting of physiological salts, energy substrates, amino acids, buffering agents, nutrients supplements, antioxidants, gentamicin and human serum albumin. These devices have different applications in Assisted Reproduction Technology (ART) procedures handled by IVF professionals. The media are aseptically filtered into gamma sterilised PETG bottles with HDPE closure and a tamper evident seal and tested for pH, osmolality, embryo toxicity, endotoxins, and sterility.

    AI/ML Overview

    This document describes the acceptance criteria and the study that proves the device meets those criteria for the Gx-IVF™, Gx-TL™, and Gx-MOPS™ PLUS media used in Assisted Reproduction Technology (ART).

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the subject devices (Gx-IVF™, Gx-TL™, Gx-MOPS™ PLUS) are primarily defined by comparison to their predicate devices and specific performance tests. The tables below summarize the key criteria and reported performance for each device.

    Gx-IVF™

    CharacteristicAcceptance Criteria (Predicate: G-IVF™ PLUS K081116)Reported Device Performance (Gx-IVF™)
    Composition
    AntibioticsGentamicinGentamicin
    ProteinHuman Serum AlbuminHuman Serum Albumin
    Amino acidsYESYES
    GlucoseYESYES
    Physiological saltsYESYES
    BufferSodium bicarbonateSodium bicarbonate
    Product & Performance Specifications
    pH7.30 ± 0.107.30 ± 0.10
    Osmolality (mOsm/kg)262 ± 5265 ± 5
    Bacterial endotoxin80% expanded blastocyst within 96 hours≥ 80% embryos developed to expanded blastocyst at 96 hours
    Sterilization MethodAseptic filtrationAseptic filtration
    Storage ConditionsStore dark at +2 to +8°CStore dark at +2 to +8°C
    BottleNalgene PETG bottle with HDPE closure and tamper evident sealNalgene PETG bottle with HDPE closure and tamper evident seal

    Gx-MOPS™ PLUS

    CharacteristicAcceptance Criteria (Predicate: G-MOPS™ PLUS K081115)Reported Device Performance (Gx-MOPS™ PLUS)
    Composition
    AntibioticsGentamicinGentamicin
    ProteinHuman Serum AlbuminHuman Serum Albumin
    Amino acidsYesYes
    GlucoseYesYes
    Physiological saltsYesYes
    BufferMOPSMOPS
    Product & Performance Specifications
    pH7.27 ± 0.077.27 ± 0.07
    Osmolality (mOsm/kg)261 ± 5265 ± 5
    Bacterial endotoxin
    • Bacterial endotoxins testing per USP
    • pH measurements per USP
    • Osmolality per USP
    • Mouse Embryo Assay (MEA)
    • Shelf-life testing
    • Open/close stability testing
    • Transportation testing (ASTM D4169-16)
    • Biocompatibility studies:
      • Cytotoxicity testing (ISO 10993-5:2009)
      • Tests for Irritation and Skin Sensitization (ISO 10993-10:2010, Guinea Pig Maximization Sensitization Test)
      • Tests for Irritation and Skin Sensitization (ISO 10993-10:2010, Vaginal Mucosal Irritation Study in Rabbits)

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This document describes technical performance and biocompatibility studies for medical devices (reproductive media and supplements). It does not involve human subject data or image analysis that would typically require expert consensus for ground truth establishment. Therefore, information regarding "number of experts" or their qualifications is not applicable in this context. The "ground truth" here is defined by established scientific and regulatory standards (e.g., USP for sterility, ISO 10993 for biocompatibility).

    4. Adjudication Method for the Test Set

    Not applicable for this type of non-clinical, in-vitro performance testing where results are based on objective laboratory measurements against defined standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This document describes the testing of in-vitro fertilization media, not an AI-assisted diagnostic device. Therefore, an MRMC comparative effectiveness study to assess human reader improvement with AI assistance is not relevant or applicable.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. The devices are media for ART, not algorithms requiring standalone performance analysis.

    7. The Type of Ground Truth Used

    The ground truth for these devices is based on established scientific and regulatory standards and objective laboratory methods for determining chemical, physical, and biological properties. This includes:

    • Regulatory Standards: USP , USP , USP , USP , ISO 13408-1:2008, ISO 13408-2:2018, ISO 10993-5:2009, ISO 10993-10:2010.
    • Defined Acceptance Specifications: pH ranges, osmolality ranges, endotoxin limits, percentage of embryo development in MEA, absence of microbial growth, non-cytotoxic, non-sensitizing, non-irritating outcomes.
    • Comparison to Predicate Devices: The performance of the subject devices is demonstrated to be substantially equivalent to that of the legally marketed predicate devices, implying the predicate devices' established safety and effectiveness serve as a benchmark.

    8. The Sample Size for the Training Set

    Not applicable. There is no machine learning or AI algorithm involved that would require a training set. The "training" in this context refers to manufacturing validation and quality control processes.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set for an algorithm is involved.