Gammex PI Hybrid Surgical Gloves Tested for Use with Chemotherapy Drugs
Gammex PI Hybrid Surgical Gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Gammex Pl Hybrid Surgical Gloves are sterile and disposable devices. Gloves are made of synthetic rubber blend of polyisoprene and polychoroprene. A polyurethane polymer coating is applied to the inner surface of the glove to make donning eay. Ansell has performed independent laboratory testing on the already cleared device to support a labeling modification for the subject device that the gloves are safe and effective for handling chemotherapy agents.
The provided text describes the acceptance criteria and performance of the Gammex PI Hybrid Surgical Gloves, specifically regarding their resistance to chemotherapy drug permeation.
Here's the breakdown of the information requested:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for chemotherapy drug permeation are implicitly defined by the reported "Minimum Breakthrough Detection Time" for each drug when tested according to ASTM D6978-05. A longer breakthrough time indicates better performance. For most drugs, the performance (>240 minutes) suggests that the gloves resist permeation for at least 4 hours, which is likely the desired acceptance criterion for practical use. The specific acceptance criteria for most physical/biological aspects are meeting the referenced ASTM and ISO standards.
Acceptance Criteria (Standard/Requirement) | Reported Device Performance (Gammex PI Hybrid Surgical Gloves) |
---|---|
Chemotherapy Permeation (ASTM D6978-05) | |
Bleomycin sulfate (15.0 mg/ml) | >240 minutes |
Busulfan (6.0 mg/ml) | >240 minutes |
Carboplatin (10.0 mg/ml) | >240 minutes |
Carmustine (3.3 mg/ml) | 23.6 minutes (Individual samples: 23.8, 26.5, 23.6) |
Cisplatin (1.0 mg/ml) | >240 minutes |
Cyclophosphamide (20.0 mg/ml) | >240 minutes |
Cytarabine (100.0 mg/ml) | >240 minutes |
Dacarbazine (10 mg/ml) | >240 minutes |
Daunorubicin (5.0 mg/ml) | >240 minutes |
Docetaxel (10.0 mg/ml) | >240 minutes |
Doxorubicin Hydrochloride (2.0 mg/ml) | >240 minutes |
Epirubicin (2.0 mg/ml) | >240 minutes |
Etoposide (20.0 mg/ml) | >240 minutes |
Fludarabine (25.0 mg/ml) | >240 minutes |
Fluorouracil (50.0 mg/ml) | >240 minutes |
Gemcitabine (Gemzar) (38.0 mg/ml) | >240 minutes |
Idarubicin (1.0 mg/ml) | >240 minutes |
Ifosfamide (50.0 mg/ml) | >240 minutes |
Irinotecan (20.0 mg/ml) | >240 minutes |
Mechlorethamine HCl (1.0 mg/ml) | >240 minutes |
Melphalan (5.0 mg/ml) | >240 minutes |
Methotrexate (25.0 mg/ml) | >240 minutes |
Mitomycin C (0.5 mg/ml) | >240 minutes |
Mitoxantrone (2.0 mg/ml) | >240 minutes |
Oxaliplatin (2.0 mg/ml) | >240 minutes |
Paclitaxel (Taxol) (6.0 mg/ml) | >240 minutes |
Rituximab (10.0 mg/ml) | >240 minutes |
ThioTEPA (10.0 mg/ml) | 36.1 minutes (Individual samples: 36.1, 37.0, 36.7) |
Vincristine Sulfate (1.0 mg/ml) | >240 minutes |
Dimensions (ASTM D3577-09) | Meets ASTM D3577-09 requirements for length, width, and thickness. |
Length: Minimum 265mm | Average 303mm |
Palm Width (e.g., size 7: 89±6mm) | Average values provided for sizes 5.5 to 9 (e.g., size 7: 91mm). |
Finger Thickness: Minimum 0.10mm | 0.22mm |
Palm Thickness: Minimum 0.10mm | 0.2mm |
Cuff Thickness: Minimum 0.10mm | 0.15mm |
Physical Properties (ASTM D3577-09) | Meets ASTM D3577-09 requirements for tensile strength and elongation at break before and after accelerated aging. |
Freedom from holes (ASTM D3577-09 & ASTM D5151-06) | Meets AQL 1.5 requirements. |
ASTM D3577-09 Inspection level/AQL: GI/AQL 0.65; Lot Size: 35,001 to 150,000 pieces; Sample size: 100 pieces; Accept number: 0 | Actual found: 0 |
Powder-Free (ASTM D3577-09 & ASTM D6124-06) | Meets applicable requirement for Powder Free; ≤ 2 mg per glove. Averaged residual powder content for the glove during process validation is 0.48mg per glove. |
Sterility (ANSI/AAMI/ISO 11137-1:2006) | Meets requirement of 10-6 SAL. |
Biocompatibility: | |
ISO in vitro cytotoxicity (ISO 10993-5:2009) | Device extract was found to be cytotoxic. (However, ISO 10993-11 systemic toxicity test showed no systemic response in the animal model, which is typically the follow-up for cytotoxic findings). |
ISO Skin Irritation Study (ISO 10993-10:2010) | Not an irritant. |
ISO Maximization Sensitization Study (ISO 10993-10:2010) | Not a sensitizer. |
ISO Systemic Toxicity Study (ISO 10993-11:2006) | No mortality or evidence of systemic toxicity. |
2. Sample Size Used for the Test Set and Data Provenance
- Chemotherapy Permeation: For the permation testing, the table in section {5} (and {2}) indicates that the "Minimum Breakthrough Detection Time" is based on 3 samples (Sample 1, 2, 3), specifically shown for Carmustine and ThioTEPA. For other drugs, only the aggregate ">240" is given, implying consistent results across samples.
- Freedom from holes: For the specific ASTM D3577-09 test mentioned in section {6}, the sample size for this attribute was 100 pieces for a lot size of 35,001 to 150,000 pieces.
- Powder-Free: The average residual powder content (0.48mg per glove) was determined "during process validation," but the specific sample size for this validation is not detailed.
- Other tests (Dimensions, Physical Properties, Biocompatibility, Sterility): The sample sizes for these tests are not explicitly stated in the provided documents, but they would have been conducted according to the requirements of the respective ASTM/ISO standards.
- Data Provenance: The studies were conducted by Ansell Healthcare Products LLC, presumably at an independent laboratory, to support a labeling modification. The data is retrospective in the sense that it was performed on an already cleared device (K151694) to add a new claim (chemotherapy drug handling), rather than being part of the initial device development for the base glove. The country of origin of the data is not specified, but the applicant is based in the USA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable as the tests performed are objective, laboratory-based chemical and physical property tests, not interpretive studies requiring human expert assessment (like image analysis or clinical diagnosis). The "ground truth" is established by direct measurement of permeation time or other physical/chemical properties according to standardized methods.
4. Adjudication Method for the Test Set
This information is not applicable. As these are laboratory measurements, there is no adjudication method in the sense of reconciling different expert opinions. The results are quantitative measurements against a predefined standard. For instance, breakthrough time is measured directly using analytical techniques.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. The document explicitly states: "A clinical study was not conducted on the subject or predicate devices." The evaluation focuses on the physical and chemical resistance properties of the gloves, not on human reader performance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No, a standalone algorithm performance study was not done. This device (surgical gloves) is a physical product and does not involve any artificial intelligence algorithms or software. The evaluation is solely on the physical and chemical properties of the gloves themselves.
7. The Type of Ground Truth Used
The ground truth used for these studies is based on:
- Direct Physical and Chemical Measurements: For chemotherapy permeation, dimensions, physical properties, freedom from holes, powder residue, and sterility.
- Standardized Biological Assay Results: For biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity).
These are objective, quantitative results obtained through testing methods specified in internationally recognized standards (ASTM, ISO).
8. The Sample Size for the Training Set
This information is not applicable. The device is a physical product (surgical gloves) and does not involve any machine learning algorithms that require a "training set."
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the same reason as point 8. No training set or associated ground truth was required for the evaluation of this physical medical device.