Search Results
Found 1 results
510(k) Data Aggregation
(274 days)
GC FujiCEM 2 (Improved)
- -Cementation of metal-based inlays, onlays, crowns and bridges
- -Cementation of resin inlays, onlays, crowns and bridges
- Cementation of all ceramic inlays -
- -Cementation of high strength (e.g. zirconia based, lithium disilicate) ceramic onlays, crowns and bridges
- Cementation of metal, ceramic and fiber posts -
GC FujiCEM 2 (Improved) is a resin modified glass ionomer luting cement. The device is composed of the two pastes, Paste A and Paste B, packaged in an automix dual barrel syringe.
The provided document is a 510(k) summary for a dental cement (GC FujiCEM 2 (Improved)). It focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing. It does not contain information about software or AI-based devices. Therefore, a table of acceptance criteria and reported device performance for an AI/software device cannot be generated from this text.
The document discusses the physical and chemical properties of the dental cement through performance bench tests and biocompatibility assessments against established ISO standards.
Here's an analysis of the available information regarding the device's evaluation, although it's for a physical dental cement product, not an AI device:
1. Table of Acceptance Criteria and Reported Device Performance (Non-AI/Software):
Property | Acceptance Criteria (Requirement) | Reported Device Performance |
---|---|---|
Biocompatibility | Conforms to ISO 10993-1:2009 standards | Acceptable per biological evaluation results |
Cytotoxicity | Based on criteria of ISO 10993-5 | Pass |
Sensitivity | Based on criteria of ISO 10993-10 | Pass |
Irritation | Based on criteria of ISO 10993-10 | Pass |
Performance Bench Tests (ISO 9917-2: 2017) | Conforms to ISO 9917-2: 2017 and company standard | Suitable for intended use |
Appearance | No visible sign of extraneous matter, free of any gelation | Conforms to requirement |
Working time | Indenter shall make a complete circular indentation | Conforms to requirement (implies appropriate working time) |
Setting time | Indenter shall fail to make a complete circular indentation | Conforms to requirement (implies appropriate setting time) |
Flexural strength | More than 25 MPa | Conforms to requirement (implies > 25 MPa) |
Radiopacity | More than equivalent thickness of aluminum | Conforms to requirement (implies > equivalent thickness of Al) |
2. Sample size used for the test set and the data provenance:
- Not applicable for AI/Software. The document describes non-clinical bench testing. The sample sizes for each specific test (e.g., number of specimens for flexural strength) are not explicitly stated within this summary. Data provenance is implied to be from the manufacturer's own testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable for AI/Software. Ground truth for dental cement properties is established through standardized physical and chemical testing methods, not expert consensus in the way it's used for AI diagnostic outputs.
4. Adjudication method for the test set:
- Not applicable for AI/Software. The performance criteria for the dental cement are objective metrics measured by equipment and standardized procedures, not subjective interpretations requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This document is for a dental cement, not an AI or software device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This is a physical medical device (dental cement), not an algorithm or software. No standalone algorithm performance was assessed.
7. The type of ground truth used:
- For biocompatibility: Ground truth is defined by the criteria and methods outlined in ISO 10993 series standards.
- For performance bench tests: Ground truth is defined by the specifications and methods in ISO 9917-2: 2017 (Dentistry - Water-based cements - Part 2: Resin-modified cements) and the company's internal standards. These are objective, measurable physical and chemical properties.
8. The sample size for the training set:
- Not applicable. This is a physical medical device. There is no training set in the context of AI/machine learning.
9. How the ground truth for the training set was established:
- Not applicable. See point 8.
Ask a specific question about this device
Page 1 of 1