Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K970703
    Device Name
    GALILEO HALLEY
    Date Cleared
    1997-05-19

    (82 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Galiteo Halley Electroencephalograph is intended for use in acquiring and storing bioelectric signals produced by brain activity for the diagnosis and prognosis of neurological disease.

    Device Description

    The GALILEO HALLEY option is an intelligent patient box (fig. 1), able to work both in connection with a GALILEO station or as stand-alone. Therefore it includes the amplification, filtering, sampling, A/D conversion and data transmission functions, which in the other GALILEO stations are performed by the umplified patient box and by the GALILEO ACQ board. In the stand-alone mode HALLEY also memorizes the acquired data in GALILEO standard file format; in the other GALILEO stations the operative system of the PC itself accomplishes this operation.

    AI/ML Overview

    The ESAOTE Galileo Halley Biosound Essote (K970703) is an electroencephalograph (EEG) device intended for acquiring and storing bioelectric signals produced by brain activity for the diagnosis and prognosis of neurological disease.

    This submission is a 510(k) premarket notification, which means the device is seeking clearance based on substantial equivalence to a predicate device, rather than requiring extensive clinical trials for a new device. The provided document does not contain a study explicitly designed to "prove" the device meets acceptance criteria in the way a clinical trial would. Instead, the "study" is a comparison to predicate devices, demonstrating that the Galileo Halley has similar technical specifications and intended use, implying comparable safety and effectiveness.

    Here's an breakdown based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the technical specifications of the predicate devices, and the reported device performance indicates that the Galileo Halley meets or exceeds these.

    ItemPredicate Devices (Galileo Vega 24 EEG K923950 & Galileo Pleiade EEG K923894) Acceptance CriteriaGalileo Halley Performance
    Intended useEEG signal acquisitionEEG signal acquisition
    General system approachAmplification and pre-processing of EFG signalAmplification and pre-processing of EEG signal
    User input devicesSerial protocolSerial interface RS232
    User output devicesSerial protocolSerial interface RS232
    Operating channels3216/32 + 4
    A/D conversion12 bit12 bit
    Sampling rate512 s/sec/channel512 s/sec/channel
    Resolution0.5 $\mu$V/bit0.5 $\mu$V/bit
    CMRR>90dB>90dB (specifically >100dB in technical characteristics)
    IMRR>120 dB>120 dB
    Noise
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1