Fluidshield* 2 Fog-Free Surgical Mask (62113); Fluidshield* 2 Fog-Free Surgical Mask with Wraparound
The Fluidshield* 2 Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. The Fluidshield* 2 Surgical Mask is a single use, disposable device, provided non-sterile.
The Fluidshield* 2 Procedure Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. The Fluidshield* 2 Procedure Mask is a single use, disposable device, provided non-sterile.
The subject device(s) are ASTM Level 2 Surgical and Procedure Mask(s). Fluidshield* 2 Surgical Mask(s), Fluidshield* 2 Procedure Mask(s) family of products are a four-layer mask, constructed of well-known non-woven materials used in facial protection. The Surgical Masks are provided with ties while the Procedure Masks are provided with earloops and a malleable nosepiece.
The Fluidshield* 2 Surgical Mask(s), Fluidshield* 2 Procedure Mask(s) family will be provided in designs with and without a protective visor. The Fluidshield* 2 Surgical Mask(s), Fluidshield* 2 Procedure Mask(s) family are single use, disposable devices, provided non-sterile.
The provided document is a 510(k) Premarket Notification for medical devices, specifically surgical and procedure masks. It describes non-clinical performance testing to demonstrate substantial equivalence to a predicate device, rather than a study proving the device meets acceptance criteria derived from clinical performance. Therefore, many of the requested items (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, training set sample size, ground truth for training set) are not applicable or cannot be extracted from this document as it pertains to a different type of evaluation.
However, I can extract information related to the acceptance criteria and reported device performance from the non-clinical testing section.
1. Table of acceptance criteria and the reported device performance:
Purpose | Test | Acceptance Criteria | Reported Device Performance (Result) |
---|---|---|---|
Level 2 - Face Mask Performance | ASTM F2100 | ASTM Level 2 | Pass |
Bacterial Filtration Efficiency | ASTM F2101 | >98% | Pass |
Particulate Filtration Efficiency | ASTM F2299 | >98% | Pass |
Differential Pressure | EN 14683 |