510(k) Data Aggregation

    K Number
    K200041
    Manufacturer
    Date Cleared
    2020-03-10

    (62 days)

    Product Code
    Regulation Number
    882.4560
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    FlexPointer 1.5 Single Use, FlexTube 3 Single Use

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The devices FlexPointer 1.5 Single Use and FlexTube 3 Single Use are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The devices are indicated for use with the Fiagon Navigation system using electromagnetic navigation.

    The devices are indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.

    Example procedures include, but are not limited to:

    ENT Procedures;
    Transphenoidal access procedures.
    Intranasal procedures.
    Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies.
    ENT related anterior skull base procedures.

    Device Description

    The Fiagon Navigation - FlexPointer 1.5 Single Use and FlexTube 3 Single Use are disposable instruments intended to be used with the Fiagon Navigation system. The FlexPointer 1.5 Single Use is an electromagnetically navigated pointing device (malleable, sensor within the tip). The FlexTube 3 Single Use is an electromagnetically navigated suction device (malleable, sensor within the tip).

    Each device incorporates a sensor device, which is tracked by the navigation system within the low-energy magnetic field of a field generator (part of the navigation system). The navigation software (part of the navigation system) displays the position of the instruments in preoperative scans (e.g., CT, MRI, fluoroscopy).

    AI/ML Overview

    The provided text describes a 510(k) submission for the Fiagon Navigation - FlexPointer 1.5 Single Use and FlexTube 3 Single Use devices. The primary purpose of the submission is to gain clearance for modified versions of previously cleared instruments, specifically a change in material and a change from reprocessed to sterile, single-use, disposable devices. The FDA's review determines substantial equivalence to predicate devices.

    Here's an analysis of the acceptance criteria and study proving compliance, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria. However, it states the performance benchmark for accuracy.

    Acceptance CriterionReported Device Performance
    Mean accuracy of the device on an anatomical phantom (mean target registration error on fiducial markers)